Stryker 8800 Series

Trendelenburg or Lateral Tilt Not Functioning

On this page

Asset Type

Surgical Table

Manufacturer

Stryker

Model

8800 Series

What This Guide Helps With

Identifies causes of non-functioning Trendelenburg or lateral tilt, including control, actuator, safety lock, and load issues before assuming internal failure.

Step-by-Step Troubleshooting

Ensure Patient Safety

Check Power Source

Inspect Hand Control / Pendant

Examine Foot Control (If Equipped)

Check Safety Locks / Brakes

Observe Table Load and Balance

Listen for Motor Response

Inspect Wiring and Actuators (External Checks Only)

Test with Minimal Load

If the Problem Persists

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Trendelenburg and lateral tilt do not respond to hand or foot controls."

Cause

What was observed during troubleshooting.

Example:
"Tilt motors did not respond; external inspection revealed no loose connections or obstructions."

Resolution

What action was taken.

Example:
"Table removed from service and labeled Out of Service for repair."

Helpful Details to Include (If Known)

Final Thought

Patient safety must guide every troubleshooting step. Systematic, logical checks allow you to identify external issues without risking patient or equipment. Proper documentation ensures clear communication and effective repair workflow.

That is successful troubleshooting.