On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots weak, inconsistent, unexpected, or apparently inaccurate fluid-management performance caused by setup, tubing, accessories, settings, or equipment faults.
Step-by-Step Troubleshooting
1. Protect the Patient and Stop Unreliable Fluid Delivery
If pressure, flow, irrigation, suction-related behavior, or another controlled output appears unreliable during a procedure, notify the clinical team and transition to an approved alternate method or verified system.
Do not continue troubleshooting while the patient depends on questionable output.
Expected outcome: Patient care continues independently of the suspect CrossFlow output.
2. Define the Exact Output Problem
Determine what staff mean by low, unstable, or inaccurate. Record:
- Function affected
- Displayed setting
- Observed behavior
- Whether output is absent, weak, excessive, fluctuating, or delayed
- When the problem begins
- Whether it occurs with all accessories or only one setup
Expected outcome: The reported problem is specific enough to reproduce and test objectively.
3. Verify the Clinical Setup
Inspect the complete external fluid path and relevant accessories. Look for:
- Kinked tubing
- Pinched tubing
- Closed clamps
- Loose connections
- Incorrect tubing routing
- Improperly installed components
- Empty or insufficient fluid source
- Obstructed components
- Incorrect accessory placement
Expected outcome: The external setup is open, secure, properly routed, and ready for normal operation. If correcting the setup restores normal output, proceed to final verification.
4. Confirm Compatible Accessories and Consumables
Verify tubing sets, connectors, and other applicable consumables are approved and compatible with the CrossFlow configuration.
Do not assume a physically fitting component will provide correct performance.
Expected outcome: Correct compatible components are being used.
5. Verify Controls and Settings
Compare the displayed settings with the intended clinical setup or approved test setup. Confirm the selected mode and user-adjustable parameters are appropriate.
Do not change protected calibration or service settings.
Expected outcome: The system is configured for the intended function. If an incorrect user setting caused the problem, correct it and verify output.
6. Check Fluid Source and External Conditions
Confirm the relevant external fluid supply is available and positioned appropriately for the setup. Inspect for collapsed containers, trapped tubing, air where inappropriate, or other external restrictions.
Expected outcome: External supply conditions allow normal CrossFlow operation.
7. Substitute Known-Good External Components
When appropriate, replace suspect tubing, accessories, connectors, or consumables with known-good approved components.
Change one variable at a time when possible.
Expected outcome: If normal performance returns after substitution, the removed external component is the likely cause.
8. Perform an Approved Performance Check
Use applicable manufacturer procedures and approved test equipment to compare actual performance with the device's specified requirements.
Do not estimate accuracy by visual impression alone and do not invent acceptance limits.
Expected outcome: Performance meets applicable manufacturer criteria. If it does, the system can proceed to final verification.
9. Verify Stable Operation
Operate the system through the relevant range or workflow permitted by the approved test procedure. Observe for fluctuating output, unexpected alarms, or recurrence.
Expected outcome: Output remains stable and meets applicable requirements throughout testing. Troubleshooting can stop after successful verification.
10. Escalate Persistent Performance Problems
If tubing, fluid source, accessories, settings, positioning, and known-good substitutions are satisfactory but performance remains outside applicable requirements, stop external troubleshooting.
Expected outcome: The CrossFlow is removed from service and sent for qualified evaluation.
If the Problem Persists
External fluid-path, supply, accessory, and configuration causes have been ruled out. Remaining possibilities may involve internal pumping, sensing, pressure or flow measurement, control electronics, calibration, or other service-level systems.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate Stryker documentation and approved test equipment
- Repaired, adjusted, or calibrated only by qualified personnel
Return to service only after applicable performance criteria and complete functional testing are satisfied. Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
When evaluating an output complaint, reproduce the complete fluid path whenever possible; the console alone may not reveal tubing or accessory restrictions.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Staff reported irrigation output from the CrossFlow was weak and inconsistent during setup."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the installed tubing was partially kinked where it passed through the procedural setup."
Resolution
What action was taken.
Example:
"The tubing was repositioned, stable output was restored, and system performance was verified using an approved nonpatient setup."
Helpful Details to Include (If Known)
- Exact function affected
- Reported output behavior
- User settings observed
- Tubing and fluid-path condition
- Fluid source condition
- Accessory and consumable identifiers
- Known-good substitutions
- Alarm behavior
- Performance-test result
- Final device status
Final Thought
Fluid-management output depends on the entire external setup as well as the console. Eliminate tubing, supply, accessory, positioning, and setting problems before suspecting internal failure, then verify performance objectively.
That is successful troubleshooting.