Stryker FL28 / FL36

Hi-Lo Actuator Failure or Slow Movement

On this page

Asset Type

Hospital Bed

Manufacturer

Stryker

Model

FL28 / FL36

What This Guide Helps With

Identifies causes of hi-lo actuator failure or slow movement, including power, battery, cabling, and obstruction issues before assuming actuator motor failure.

Step-by-Step Troubleshooting

Ensure Patient Safety

Check Power Connection

Inspect Battery Backup

Examine Hand/Foot Controls

Inspect Cables and Connections

Check for Obstructions

Test Bed Functions

If the Problem Persists

External causes have been ruled out. The actuator or internal control system may be faulty.

Clinical Use Tip

Never troubleshoot while the patient is in bed. Always move the patient to a safe, stable bed before attempting repairs.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Hi-lo actuator slow to respond or not functioning."

Cause

What was observed during troubleshooting.

Example:
"Battery low and actuator cable partially disconnected."

Resolution

What action was taken.

Example:
"Recharged battery, secured cable connections; actuator movement restored."

Helpful Details to Include (If Known)

Final Thought

Patient safety and logical, stepwise troubleshooting are critical. Always rule out power, controls, and mechanical issues before assuming actuator failure. Proper documentation ensures accurate tracking and compliance.

That is successful troubleshooting.