Stryker FL28 / FL36

Power Supply or Battery Backup Failure

On this page

Asset Type

Hospital Bed

Manufacturer

Stryker

Model

FL28 / FL36

What This Guide Helps With

Identifies causes of bed power loss or battery backup failure, including outlet, cabling, battery, and fuse issues before assuming internal electronics failure.

Step-by-Step Troubleshooting

Ensure Patient Safety

Confirm the patient is safely transferred or the bed is unoccupied. Never work on power systems with a patient in the bed.

Check Main Power Source

Expected outcome: Bed powers on when plugged into a confirmed live outlet.

Inspect Power Cord and Connections

Expected outcome: No visible damage; cord firmly connected.

Verify On-Board Fuses or Circuit Breakers

Expected outcome: Bed powers on or shows normal status indicators.

Check Battery Backup

Expected outcome: Bed operates on battery when unplugged; indicator shows charge.

Swap or Test Battery

Expected outcome: Bed powers on with a known good battery.

Observe Bed Behavior

If the Problem Persists

External power sources, cords, fuses, and batteries have been ruled out. The bed should be:

Stopping here prevents unsafe troubleshooting and ensures patient safety.

Clinical Use Tip

Never attempt troubleshooting while a patient remains in bed. Move patient to another bed first. Confirm power loss isn’t intermittent during patient care.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Bed loses power intermittently, fails to operate on battery."

Cause

What was observed during troubleshooting.

Example:
"Battery was discharged and power cord had a loose connection."

Resolution

What action was taken.

Example:
"Secured power cord and replaced battery; bed powers on normally."

Helpful Details to Include (If Known)

Final Thought

Patient safety is the top priority. Systematic, logical troubleshooting ensures external causes are ruled out before escalating. Proper documentation and timely escalation preserve safety and minimize downtime.

That is successful troubleshooting.