Stryker InTouch Hospital Bed

Motor Actuator Failure

On this page

Asset Type

Hospital Bed

Manufacturer

Stryker

Model

InTouch

What This Guide Helps With

This guide addresses situations where the Stryker InTouch bed’s motor actuators fail to operate. Symptoms may include:

Focus is on external checks and verification before considering internal repair.

Step-by-Step Troubleshooting

Verify Bed Power

Why: Actuators require full voltage; low or absent power can mimic motor failure.

Check Control Panel Functionality

Why: If the panel does not send commands, actuators will not move regardless of their condition.

Inspect Bed Lock and Brake Sensors

Why: Some actuators will not operate if the bed is “locked” or if sensors detect unsafe conditions.

Listen for Motor Noise

Why: No noise may indicate loss of power to actuator; unusual noise may indicate mechanical obstruction.

Check Cables and Connections to Actuators

Why: Loose or damaged wiring is a common cause of actuator failure.

Swap or Test Actuator (if spare available)

Why: Confirms whether the actuator itself is faulty or the issue lies elsewhere.

If the Problem Persists

If external power, panel function, brakes, and wiring have been verified, the actuator is likely internally failed.

Knowing when to escalate prevents further patient risk or damage.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
Bed will not adjust positions; motor sounds but no movement.

Cause

What was observed during troubleshooting.

Example:
Actuator did not respond to control panel commands; external power and connections verified.

Resolution

What action was taken.

Example:
Bed removed from service; actuator labeled and sent for bench evaluation.

Helpful Details to Include

Final Thought

Patient safety and methodical, logical troubleshooting are paramount. External checks eliminate simple causes, while knowing when to escalate prevents risk and equipment damage. Document all observations clearly for service or repair follow-up.

That is successful troubleshooting.