Stryker InTouch Hospital Bed

Control Panel Not Responding

On this page

Asset Type

Hospital Bed

Manufacturer

Stryker

Model

InTouch

What This Guide Helps With

This guide addresses situations where the Stryker InTouch bed’s control panel fails to respond to user inputs. Symptoms may include:

Focus is on external and easily verifiable causes before assuming internal electronics failure.

Step-by-Step Troubleshooting

Verify Bed Power

Check Main Disconnect / Battery Switch

Inspect Cables and Connectors

Check for Physical Damage

Test Other Control Inputs

Cycle Power Again if Needed

If the Problem Persists

External causes have been ruled out. The control panel or main electronics may be internally faulty.

Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Do not attempt adjustments or troubleshooting while a patient is on the bed. Move the patient to another bed before further evaluation. Only operate controls when safe.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Control panel unresponsive; bed will not adjust height or position.”

Cause

What was observed during troubleshooting.

Example:
Bed has power, but panel does not respond to any inputs. External connections and other controls verified functional.

Resolution

What action was taken.

Example:
Bed removed from service, labeled Out of Service, sent for repair/bench evaluation.

Helpful Details to Include

Final Thought

Proper troubleshooting prioritizes patient safety, confirms external causes first, and avoids unnecessary internal repair attempts. Documentation ensures a clear history for repair and risk mitigation.

That is successful troubleshooting.