Stryker LIFEPAK 35

EtCO₂ Sampling Failure

On this page

Asset Type

Defibrillator

Manufacturer

Stryker

Model

LIFEPAK 35

What This Guide Helps With

Troubleshooting absent, intermittent, or unreliable EtCO₂ sampling caused by sampling lines, airway adapters, moisture, blockage, connections, or monitor faults.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot an unresolved EtCO₂ sampling problem while the LIFEPAK 35 is being relied upon for active patient monitoring.

If continuous capnography is clinically required, move the patient to another verified monitoring method or known-good device before continuing.

Expected: Patient care is not dependent on unreliable EtCO₂ monitoring.

Why it matters: The LIFEPAK 35 is intended to provide capnography among its monitoring functions, so an unresolved sampling failure can affect respiratory monitoring reliability.

2. Confirm the Reported Failure

Power on the LIFEPAK 35 and determine the exact behavior:

Record any exact on-screen message before clearing or restarting the device.

Expected: The failure is clearly identified and reproducible.

If normal EtCO₂ operation returns and remains stable, document the findings and stop.

3. Verify Capnography Is Active and Displayed

Confirm the device is configured to display the CO₂ parameter and capnogram as expected.

Check that:

Expected: EtCO₂ monitoring is enabled and the display is configured to show the parameter.

Why it matters: A display or configuration issue can resemble sampling failure even when the hardware path is functional.

4. Inspect the Sampling Line

Examine the complete EtCO₂ sampling line for:

Straighten the tubing and reseat all accessible connections.

Expected: The sampling path is open, dry, intact, and securely connected.

If the issue resolves, replace any questionable disposable component, verify stable operation, and stop.

5. Check for Moisture or Secretions

Inspect the sampling line and patient-side connection for:

Do not attempt to restore a contaminated disposable sampling line by forcing air through it.

Replace the affected disposable sampling component with a compatible new one.

Expected: A clean, unobstructed sampling path is established.

If normal waveform and EtCO₂ values return, the contaminated sampling component was the likely cause; stop.

6. Inspect the Airway Adapter or Cannula

For intubated monitoring, inspect the airway adapter and sampling connection for:

For non-intubated monitoring, inspect the cannula for:

Expected: The patient-side sampling accessory is correctly positioned and unobstructed.

7. Verify the Sampling Line Is Fully Seated at the LIFEPAK 35

Disconnect and reconnect the sampling line at the monitor-side CO₂ connection according to normal accessory handling practices.

Inspect externally for:

Do not insert tools or probes into the device sampling inlet.

Expected: The sampling line seats securely without looseness or visible damage.

8. Substitute a Known-Good Compatible Sampling Set

Replace the complete sampling line and applicable patient-side adapter or cannula with a known-good compatible set.

Do not change multiple device settings at the same time.

Expected: The LIFEPAK 35 produces a stable capnogram and EtCO₂ value with the replacement sampling set.

If the issue resolves, the original sampling accessory was faulty or obstructed; stop.

Stryker provides dedicated LIFEPAK 35 capnography training and accessory resources, supporting verification of the intended capnography setup and compatible accessory path.

9. Determine Whether the Failure Follows the Accessory or the Monitor

Using controlled bench evaluation:

Expected:

Why it matters: This separates an external disposable-path failure from a likely monitor-side fault.

10. Power Cycle the Device Under Controlled Conditions

After recording any active message or failure behavior:

Expected: Capnography initializes normally and remains stable.

If the failure returns, do not repeatedly power cycle the unit as a substitute for repair.

11. Perform a Controlled Capnography Functional Evaluation

Use the facility’s approved test equipment and current Clinical Engineering procedure to evaluate capnography performance.

Verify, as applicable:

Stryker specifically lists a LIFEPAK 35 Performance evaluation – Capnography resource, making a controlled capnography functional check appropriate before return to service.

Expected: The unit passes the complete approved capnography evaluation.

If the unit does not pass, stop troubleshooting and escalate.

If the Problem Persists

If a known-good compatible sampling set, airway accessory, connection check, configuration review, controlled restart, and approved capnography evaluation do not restore reliable EtCO₂ operation, common external causes have been ruled out.

The issue is likely within the monitor-side capnography path or associated internal electronics.

The device should be:

Do not continue into internal pneumatic-path, module, PCB, or component-level repair unless performed under authorized service procedures with appropriate training and test equipment.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Do not troubleshoot unreliable EtCO₂ sampling on an active patient. Transfer respiratory monitoring to another verified method first, then evaluate the LIFEPAK 35 under controlled conditions.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported the LIFEPAK 35 displayed no reliable EtCO₂ value or capnogram during patient monitoring."

Cause

What was observed during troubleshooting.

Example:
"Inspection found the EtCO₂ sampling line obstructed by visible moisture, preventing reliable gas sampling."

Resolution

What action was taken.

Example:
"Removed the contaminated sampling set, installed a compatible replacement, and verified stable capnogram and EtCO₂ response during functional testing."

Helpful Details to Include (If Known)

Final Thought

EtCO₂ sampling failures are often caused by external problems such as obstruction, moisture, damaged tubing, poor connections, or defective disposables. Patient safety comes first, followed by logical isolation of the sampling path. When known-good accessories and controlled functional testing do not restore reliable operation, remove the LIFEPAK 35 from service and escalate appropriately. Clear CCR documentation preserves both the technical findings and the final safety decision.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide