Stryker LIFEPAK 35

NIBP Measurement Failure

On this page

Asset Type

Defibrillator

Manufacturer

Stryker

Model

LIFEPAK 35

What This Guide Helps With

Troubleshooting failed, incomplete, inconsistent, or unavailable noninvasive blood pressure measurements caused by cuff, hose, connection, setup, motion, or device faults.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot an unreliable NIBP function while the LIFEPAK 35 is being relied upon for active patient monitoring.

Move the patient to another verified blood pressure monitoring method or known-good monitor before extended troubleshooting.

Expected: Patient care continues without dependence on an unresolved NIBP measurement problem.

Why it matters: NIBP results support clinical assessment, and repeated failed or questionable measurements should not delay obtaining a reliable blood pressure by another appropriate method.

2. Verify the Reported Failure

Confirm exactly what occurs when an NIBP measurement is initiated.

Check whether the device:

Record the exact displayed message or behavior when available.

Expected: The failure pattern is clearly identified before components are exchanged.

If the problem is not reproducible and repeated controlled measurements complete normally, continue functional verification before returning the device to service.

3. Inspect the NIBP Cuff

Remove the cuff from clinical use for inspection.

Check for:

Replace a visibly damaged or questionable cuff with a known-good compatible cuff.

Expected: The cuff is intact, correctly sized, and capable of maintaining pneumatic pressure.

If replacing the cuff resolves the issue, verify repeat measurements and stop troubleshooting.

4. Verify Cuff Size and Application

Confirm the cuff size is appropriate for the test subject or approved simulator setup.

For controlled testing on a person, verify:

Expected: The cuff inflates and completes a measurement without excessive reinflation or premature termination.

Why it matters: Incorrect sizing, poor placement, and movement can prevent reliable oscillometric measurements even when the monitor is functioning normally.

If correct application resolves the issue, complete repeat testing and stop.

5. Inspect the NIBP Hose

Examine the complete hose from the cuff to the LIFEPAK 35.

Check for:

Straighten any kinked tubing and replace a damaged hose.

Expected: The pneumatic path is open and free of visible leakage or restriction.

If the issue resolves with a known-good hose, document the failed accessory and stop.

6. Reseat All Pneumatic Connections

Disconnect and reconnect:

Confirm each connection is fully seated and mechanically secure.

Do not force a connector that does not fit correctly.

Expected: Connections remain secure during inflation and do not leak or separate.

If reseating the connections restores reliable measurements, repeat testing and stop.

7. Substitute Known-Good Compatible Accessories

Using approved compatible accessories, test with:

Change one component at a time when practical so the failed item can be identified.

Expected: A repeatable accessory-related failure is isolated.

Interpretation:

8. Check for Pneumatic Leak Behavior

Initiate a controlled NIBP measurement and observe inflation behavior without disassembling the device.

Look for:

Do not repeatedly cycle a visibly damaged pneumatic system.

Expected: Inflation is smooth, the cuff maintains pressure as required during the measurement cycle, and controlled deflation occurs normally.

If external leakage is found, replace the affected cuff or hose and retest.

9. Rule Out Motion and Test-Condition Problems

Repeat the measurement under stable conditions.

For a controlled human-subject check, ensure the subject is:

For bench evaluation, use the organization’s approved NIBP analyzer and established test procedure.

Expected: The measurement completes consistently under controlled conditions.

If the failure occurs only with movement or poor application but not during controlled testing, the monitor itself may be functioning normally.

10. Review NIBP Settings and Operating State

Verify that the intended NIBP measurement mode and associated settings are appropriate for the test.

Check for:

Expected: The NIBP function is configured appropriately and accepts a normal measurement request.

If correcting the setup resolves the issue, repeat testing and stop.

11. Perform a Controlled Restart

Only after the device is removed from patient use:

Expected: The device starts normally and NIBP measurements complete consistently.

If the problem resolves, perform repeated functional verification before return to service.

12. Perform Clinical Engineering Functional Verification

Using approved test equipment and your facility’s established inspection procedure:

Do not rely on a single successful reading after an intermittent failure.

Expected: NIBP measurements complete repeatedly and meet the organization’s applicable performance requirements.

If the device fails controlled testing with known-good accessories, stop external troubleshooting and escalate.

If the Problem Persists

If the LIFEPAK 35 continues to fail NIBP measurements after:

then common external causes have been reasonably ruled out.

The issue may involve the internal NIBP pneumatic or monitoring system and requires qualified repair evaluation.

The device should be:

Do not continue repeated measurement attempts or begin unnecessary internal disassembly without the appropriate authorized service process.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Do not troubleshoot unreliable NIBP measurements on an active patient. Transfer blood pressure monitoring to another verified method or device first, then evaluate the LIFEPAK 35 under controlled conditions.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the LIFEPAK 35 repeatedly inflated the NIBP cuff but failed to display a blood pressure measurement."

Cause

What was observed during troubleshooting.

Example:
"Inspection found the NIBP hose leaking near the cuff-side connection, and the failure did not recur with a known-good compatible hose."

Resolution

What action was taken.

Example:
"Replaced the damaged NIBP hose, completed repeated controlled NIBP measurements, verified acceptable operation, and returned the defibrillator to service."

Helpful Details to Include (If Known)

Final Thought

Reliable NIBP troubleshooting starts with patient safety and the external pneumatic path: cuff, sizing, application, hose, connectors, motion, and settings. Controlled substitution and repeatable verification help distinguish accessory problems from internal failure. When known-good accessories and controlled testing still produce failure, removing the device from service and escalating for qualified repair is the correct action. Clear CCR documentation preserves what was reported, what was found, and why the final disposition was appropriate.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide