Stryker Prime

Trendelenburg or hi-lo function failure

On this page

Asset Type

Stretcher

Manufacturer

Stryker

Model

Prime

What This Guide Helps With

Stretcher not raising, lowering, or tilting due to power, controls, lockouts, or connection issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify Function Attempt

Check Battery Charge Level

Confirm Brake Status

Inspect Control Panels and Cables

Check for Obstructions or Mechanical Binding

Listen for Actuator Response

Power Cycle the Stretcher

Verify Charger Connection

If the Problem Persists

If all external checks have been completed and the stretcher still does not raise, lower, or tilt, the issue is likely internal (actuator, control board, or wiring failure).

Remove the stretcher from service, label it Out of Service, and send for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot lift or tilt functions with a patient on the stretcher. Always transfer the patient first to prevent fall or injury risk.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Stretcher would not raise or go into Trendelenburg during patient transport."

Cause

What was observed during troubleshooting.

Example:
"Battery found fully depleted preventing actuator operation."

Resolution

What action was taken.

Example:
"Charged stretcher and verified full hi-lo and Trendelenburg functionality."

Helpful Details to Include (If Known)

Final Thought

Always approach movement failures by ruling out simple causes like power, controls, and obstructions before assuming actuator failure. Patient safety comes first, and clear documentation ensures continuity of care and efficient repair decisions.

That is successful troubleshooting.