Stryker ProCuity

Head, Knee, or Foot Actuator Not Moving

On this page

Asset Type

Hospital Bed

Manufacturer

Stryker

Model

ProCuity

What This Guide Helps With

Troubleshooting one bed section that will not move due to power, lockouts, controls, obstructions, connections, positioning, or a service-level motion fault.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Continuity of Care

Do not troubleshoot unreliable bed movement while a patient depends on the affected section for positioning, therapy, airway management, or fall prevention.

Notify clinical staff and determine whether the patient must be transferred to another verified bed. Support the mattress platform and patient as clinically appropriate if the section is stuck in an unsafe position. Do not place hands, tools, cables, linens, or equipment near moving linkages.

Expected outcome: The patient is protected, and troubleshooting can proceed without creating a pinch, fall, or positioning hazard.

If safe patient positioning cannot be maintained, remove the bed from clinical use and stop troubleshooting.

2. Confirm the Exact Motion Failure

Identify which section is affected:

Test the function from each available authorized control location, such as the caregiver controls and patient controls, while the bed is unoccupied or safely supported. Observe whether the section is completely inactive, moves intermittently, moves in only one direction, or produces sound without movement.

Expected outcome: The failure is narrowed to a specific section, direction, and control location.

If the section operates normally from another control location, troubleshoot the nonresponsive control rather than the actuator.

3. Verify AC Power and Basic Bed Operation

Confirm that the bed power cord is fully connected to a functioning approved receptacle. Inspect the plug and cord for damage, overheating, contamination, or loose strain relief.

Verify that the bed powers normally and test another unrelated powered function.

Expected outcome: The bed has stable power and other powered functions operate normally.

If restoring power corrects the motion problem, perform final functional verification and troubleshooting can stop.

4. Check Motion Lockouts and Control Status

Inspect caregiver controls for an active lockout affecting the head, knee, foot, or patient controls. Verify that the intended function is enabled using normal authorized controls.

Do not enter restricted service menus or change unrelated configuration settings.

Expected outcome: The affected motion is enabled and no lockout indicator remains active.

If clearing an unintended lockout restores normal motion, verify operation from applicable controls and troubleshooting can stop.

5. Inspect for Mechanical Obstruction or Excess Loading

Inspect beneath and around the mattress platform for:

Do not force the section or manually pry the linkage.

Expected outcome: The affected section has a clear travel path and is not externally restrained.

If removing an obstruction restores full movement, inspect for damage, complete functional testing, and troubleshooting can stop.

6. Compare Control Sources

Operate the affected section from the caregiver control, patient control, and any approved pendant fitted to the bed.

Determine whether:

Expected outcome: The failure is isolated to the motion system or to a specific external control source.

If the actuator works consistently from another control, inspect the failed control, pendant, or external cable and stop actuator-focused troubleshooting.

7. Inspect Accessible Actuator Wiring and Connections

Disconnect AC power and follow facility procedures before inspecting accessible wiring.

Visually inspect accessible actuator cables, connectors, and harness routing for:

Reseat only clearly accessible connectors intended for routine service access. Do not open the actuator or controller enclosure.

Expected outcome: External actuator wiring is secure, undamaged, and correctly routed.

If securing an accessible connection restores operation, perform repeated movement testing and troubleshooting can stop.

8. Test Under Safe, Unloaded Conditions

With the bed unoccupied and clear of obstructions, operate the affected section through a safe range of travel.

Listen for clicking, grinding, repeated stopping, abnormal motor noise, or movement that is uneven or jerky. Stop immediately if the frame binds or the actuator appears mechanically unstable.

Expected outcome: The section moves smoothly in both directions without abnormal sound, delay, or binding.

Normal unloaded movement followed by failure under ordinary loading requires service-level evaluation and should not be considered resolved.

9. Compare With a Known-Good Control or Accessory When Applicable

When the bed uses a removable approved control or pendant, test with a compatible known-good accessory according to facility procedures.

Do not interchange actuators, control modules, or internal components as an external troubleshooting step.

Expected outcome: A known-good external control either restores motion or confirms that the fault remains within the bed.

If the known-good control restores reliable operation, replace the defective accessory, verify all functions, and troubleshooting can stop.

10. Complete Final Functional Verification

After any correction, test the head, knee, and foot sections from applicable caregiver and patient controls. Verify both directions, lockout operation, smooth movement, cable clearance, and safe stopping.

Inspect the frame and mattress platform for interference after movement.

Expected outcome: All intended bed sections move reliably, lockouts operate correctly, and no abnormal condition remains.

The bed may return to service only after the affected function passes complete functional verification.

If the Problem Persists

Common external causes such as power loss, lockouts, control selection, obstructions, accessories, and accessible connections have been ruled out.

The remaining cause may involve an actuator, position feedback device, motion-control circuit, internal harness, mechanical linkage, controller configuration, or another service-level fault.

The bed should be:

Do not continue cycling a stalled or noisy actuator. Do not open sealed actuator assemblies or perform board-level repair. Following service, complete electrical safety testing when applicable and full bed-function verification before return to service.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Transfer the patient before troubleshooting a section that cannot maintain safe positioning or may move unexpectedly.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the head section would not raise from either the patient or caregiver controls."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found an unintended head-motion lockout active on the caregiver control panel."

Resolution

What action was taken.

Example:
"Cleared the lockout and verified smooth head, knee, and foot movement in both directions from all applicable controls."

Helpful Details to Include (If Known)

Final Thought

Protect the patient first, verify controls and external causes before assuming actuator failure, stop when movement is unsafe, escalate appropriately, and document the complaint, cause, correction, and final verification clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide