Stryker ProCuity

Nurse Call or Room Interface Communication Failure

On this page

Asset Type

Hospital Bed

Manufacturer

Stryker

Model

ProCuity

What This Guide Helps With

Troubleshooting failed nurse call activation or room-interface communication caused by connections, cables, room hardware, positioning, pairing, power, or configuration issues.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot an unreliable nurse call connection while the bed is the patient’s only method of requesting assistance.

Expected outcome: The patient has a reliable alternate method to request assistance during troubleshooting.

2. Confirm the Exact Communication Failure

Determine whether:

Activate nurse call from each available patient and caregiver control while clinical staff confirm whether the correct room and call location appear at the nurse station.

Expected outcome: The failure is isolated to a specific control, the entire bed, the room connection, or the wireless interface.

If all call locations operate correctly, the problem is resolved. Stop troubleshooting.

3. Verify Bed Power and Basic Operation

Confirm that:

A widespread control or power problem may appear to be an isolated nurse call failure.

Expected outcome: The bed is powered normally and capable of processing control inputs.

If restoring power resolves the failure, retest nurse call and stop.

4. Determine Whether the Bed Uses Wired or Cable-Free Nurse Call

Inspect the installation and confirm the intended communication method.

Wired systems use a Stryker-compatible interface cable connected to the ProCuity’s DB-37 connector at the head end and to the room’s patient station, headwall, or docking connection.

Cable-free systems communicate through a mapped Secure Connect device and require the bed to establish a connection with that room device.

Do not assume that wireless bed connectivity automatically means nurse call is configured for cable-free operation.

Expected outcome: The correct communication path is identified before cables, room hardware, or wireless pairing are tested.

5. Inspect the Wired Interface Cable and Connections

For a wired installation:

Stryker specifies use of a Stryker-supplied interface cable and identifies the DB-37 connection as the wired nurse call input.

Expected outcome: Both ends are secure, correctly connected, and free from visible damage.

If reseating the cable restores nurse call operation, secure the routing, perform repeated tests, and stop.

6. Test With a Known-Good Compatible Interface Cable

Replace the installed interface cable with a known-good, facility-approved Stryker-compatible cable.

Retest nurse call from:

Expected outcome: A defective cable is either confirmed or ruled out.

If the known-good cable resolves the issue, replace the failed cable, verify correct room identification, and stop.

7. Compare the Bed in a Known-Good Room Connection

When operationally practical, connect the affected bed to a known-good compatible room interface. Alternatively, connect a known-good ProCuity bed to the original room connection.

Interpret the results:

Expected outcome: The failure is isolated to either the bed or the hospital room infrastructure.

8. Check Individual Nurse Call Controls

Test every available nurse call button separately.

Look for:

Do not open the siderail or control panel during external troubleshooting.

Expected outcome: A localized control failure is distinguished from a complete room-interface failure.

If one control fails but the others reliably activate the correct call, document the failed control and remove the bed from service for repair.

9. Verify Cable-Free Secure Connect Conditions

For a cable-free nurse call installation:

Stryker identifies brake application as part of establishing the Secure Connect connection. Current guidance also directs technicians to confirm the Secure Connect icon on the bed display before testing nurse call.

Expected outcome: The bed displays a valid connection to the correct room’s Secure Connect device.

If the connection icon appears and nurse call functions correctly after repositioning or applying the brakes, perform repeated tests and stop.

10. Rule Out a Room Mapping or Configuration Problem

Determine whether:

Stryker states that Secure Connect or an iBed Locator must be mapped to its physical location and remapped when moved. Room-interface configuration is performed through authorized setup procedures and should not be changed without the approved configuration information and service resources.

Expected outcome: Incorrect mapping or configuration is identified as an infrastructure or authorized-service issue.

Escalate mapping, server, gateway, or configuration problems to the appropriate Stryker-trained, nurse call, or clinical systems support team.

11. Perform a Final Functional Verification

After any corrective action:

Stryker’s setup instructions require activating the nurse call button to verify communication between the bed and the hospital nurse call system.

Expected outcome: Nurse call communication is consistent, correctly identified, and reliable from all available controls.

If all tests pass, return the bed to service according to facility policy.

If the Problem Persists

If the bed fails with a known-good cable in multiple verified room connections, or if Secure Connect remains unavailable after external positioning, brake, power, and mapping checks, common external causes have been ruled out.

The fault may involve the room interface board, control wiring, siderail control assembly, gateway, bed software, or another internal communication component. Stryker identifies both wired and wireless bedside room interface board configurations within the ProCuity platform.

The bed should be:

Knowing when to stop external troubleshooting is proper troubleshooting, especially when nurse call reliability directly affects patient safety.

Clinical Use Tip

Never leave a patient dependent on a nurse call system that has not been verified at the nurse station. Provide an alternate call method before troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the ProCuity nurse call button responded at the bed, but no call appeared at the nurse station."

Cause

What was observed during troubleshooting.

Example:
"The DB-37 nurse call interface cable had a damaged connector and failed when tested against a known-good cable."

Resolution

What action was taken.

Example:
"Replaced the interface cable, verified nurse call from all patient and caregiver controls, and confirmed correct room identification at the nurse station."

Helpful Details to Include (If Known)

Final Thought

Reliable nurse call operation must be confirmed through the entire communication path—not only at the bed. Logical isolation of the controls, cable, room connection, and wireless interface protects patients and prevents unnecessary internal repair. Clear CCR documentation ensures that any recurring bed-side or room-side failure can be recognized quickly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide