Stryker Secure II

Manual CPR Release Works but Powered Motion Will Not Rehome

On this page

Asset Type

Hospital Bed

Manufacturer

Stryker

Model

Secure II

What This Guide Helps With

Troubleshoot powered positioning that fails after manual CPR release due to incomplete mechanism reset, obstruction, controls, or position-related faults.

Step-by-Step Troubleshooting

1. Ensure Patient Safety and Maintain Emergency Positioning Capability

Determine whether the bed is occupied or needed for immediate patient care.

If powered positioning cannot be restored reliably, arrange another verified bed.

Do not repeatedly activate the CPR release while a patient depends on the bed.

Perform testing with the bed unoccupied and secured.

Expected outcome: Patient care is maintained without relying on an unreliable positioning mechanism.

2. Confirm the Reported CPR and Motion Behavior

Ask whether the manual CPR release lowered the head section normally.

Determine whether powered motion stopped immediately after release or whether the problem existed previously.

Identify which powered movements remain unavailable.

Expected outcome: The relationship between CPR release and powered motion failure is established.

3. Inspect the Manual CPR Release Control

Inspect the accessible release handle and cable path for visible damage, obstruction, or abnormal positioning.

Verify that the release control has returned to its normal position.

Do not force a damaged release mechanism.

Expected outcome: The manual CPR control is visibly returned to its normal operating position.

4. Check for Mechanical Obstructions

Inspect the head section, deck linkage, mattress, and nearby accessories.

Look for trapped linens, equipment cables, or objects preventing normal deck movement.

Keep hands clear of moving components and pinch points.

Expected outcome: No external obstruction prevents the head section from moving.

5. Verify AC Power and Controls

Confirm that the bed receives power from a verified outlet.

Check whether other powered functions operate.

Inspect caregiver controls and applicable lockouts.

Determine whether the failure is limited to the head section or affects multiple movements.

Expected outcome: Power and authorized controls operate normally.

6. Verify the Release Mechanism Is Properly Reset

Using the applicable Stryker instructions, confirm that the CPR release mechanism is in its required normal operating condition.

Do not manually force the deck into position or bypass any latch or safety interlock.

If the mechanism cannot be confirmed properly reset, stop testing and escalate.

Expected outcome: The release mechanism is correctly positioned for powered operation.

7. Attempt Controlled Powered Movement

With the bed unoccupied, command the affected function through a small safe range.

Observe whether movement begins normally.

Stop immediately if the deck binds, moves unexpectedly, produces unusual noise, or fails to support itself.

Expected outcome: Powered motion resumes smoothly without abnormal behavior.

8. Perform Final CPR and Positioning Verification

If the problem has been corrected, verify powered head-section movement and stopping.

Perform the applicable manufacturer-approved CPR release and reset verification.

Confirm that the deck returns to normal powered operation after the test.

If all required checks pass, troubleshooting can stop.

Expected outcome: Manual CPR release and powered positioning both operate reliably.

If the Problem Persists

External controls, power, obstructions, and the accessible release mechanism have been checked.

Remaining causes may include CPR release linkage, actuator engagement, position feedback, drive components, or internal control circuitry.

Remove the bed from service, label it Out of Service, and arrange qualified repair or bench evaluation.

Use applicable Stryker service documentation and approved test equipment. Do not force the mechanism or attempt unauthorized internal adjustments.

Complete CPR release, reset, and powered positioning verification before return to service.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

After any CPR release mechanism service, verify that powered head-section positioning has been restored before returning the bed to patient care.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the Secure II head section would not raise electrically after the manual CPR release was used."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found that the accessible CPR release handle had not fully returned to its normal position."

Resolution

What action was taken.

Example:
"Returned the handle to its normal position according to the approved instructions and verified powered head movement and CPR release reset operation."

Helpful Details to Include (If Known)

Final Thought

CPR release and powered positioning must both remain dependable. Inspect the accessible release mechanism, controls, power, and obstructions before considering internal failure. Stop when the mechanism behaves abnormally, escalate appropriately, and document the final verification.

That is successful troubleshooting.