Stryker Secure II

Siderail Controls, Nurse Call, or Bed Exit Interface Fails

On this page

Asset Type

Hospital Bed

Manufacturer

Stryker

Model

Secure II

What This Guide Helps With

Troubleshoot unresponsive siderail controls, nurse-call signaling, or bed-exit interfaces caused by lockouts, damaged connections, configuration, or external communication problems.

Step-by-Step Troubleshooting

1. Protect Patient Safety and Maintain Communication

Determine whether the failure affects patient positioning, nurse-call communication, or bed-exit monitoring.

If nurse call is unreliable, provide an alternate verified communication method.

If bed-exit monitoring is unavailable, notify nursing so an appropriate alternate safety plan can be established.

Do not troubleshoot these functions while a patient depends on them.

Expected outcome: Patient communication and fall-prevention needs are maintained independently of the affected bed.

2. Identify the Failed Function

Confirm whether the complaint involves:

Determine whether the problem occurs on one siderail or multiple control locations.

Expected outcome: The failed function and affected control location are identified.

3. Verify Bed Power and Basic Operation

Inspect the power cord and connect the bed to a verified appropriate outlet.

Check whether normal powered positioning functions operate.

Observe available status indicators for abnormal conditions.

Expected outcome: The bed has normal power and basic control operation.

4. Inspect Siderail Controls and Lockouts

Check buttons for visible damage, contamination, sticking, or physical obstruction.

Verify relevant caregiver lockouts.

Compare the affected function using another authorized control location.

Do not force damaged buttons or open sealed control assemblies.

Expected outcome: Controls respond normally when enabled and undamaged.

5. Inspect Accessible Communication Connections

If the bed is connected to a facility nurse-call system, inspect the external interface cable and connectors.

Look for:

Reconnect approved external connections when safe.

Expected outcome: Communication cables are correctly installed and undamaged.

6. Isolate the Bed From the Facility Interface

Use an approved known-good interface cable or designated test connection when available.

Determine whether the failure follows the bed, the cable, or the room connection.

Coordinate with the nurse-call or facilities support team before changing infrastructure connections.

Expected outcome: The likely failure location is identified without unauthorized infrastructure changes.

7. Verify Bed-Exit Interface Configuration

Confirm whether the installed bed configuration supports the reported bed-exit interface.

Inspect available user-accessible settings and alarm status.

Do not assume that every Secure II configuration provides identical communication features.

Do not modify restricted configuration settings.

Expected outcome: The installed interface and authorized settings match the intended clinical workflow.

8. Perform End-to-End Functional Verification

With the bed unoccupied and using approved test methods, verify siderail control response.

Test nurse-call activation and confirm receipt at the appropriate staff communication location.

Verify bed-exit alarm operation and remote signaling if equipped.

If all affected functions operate correctly, troubleshooting can stop.

Expected outcome: Local controls and the complete communication path operate reliably.

If the Problem Persists

Common external power, control, lockout, cable, and interface causes have been ruled out.

Remaining problems may involve internal siderail controls, communication electronics, wiring, bed-exit circuitry, or facility infrastructure.

Remove the bed from service, label it Out of Service, and arrange qualified bench evaluation or repair.

Coordinate infrastructure faults with the responsible facility team.

Evaluate the bed using applicable Stryker documentation and approved test equipment. Verify local and remote functions before return to service.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

A nurse-call button lighting up does not confirm successful communication; verify that the call reaches the intended receiving station.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the Secure II siderail nurse-call button did not activate the room nurse-call system."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the external nurse-call interface cable disconnected from the bed."

Resolution

What action was taken.

Example:
"Reconnected the cable and verified that siderail nurse-call activation was received at the nursing station."

Helpful Details to Include (If Known)

Final Thought

Patient communication and fall-prevention systems require complete functional verification. Check accessible controls, cables, settings, and facility interfaces before assuming internal failure. Escalate unresolved faults and document the verified communication path.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide