Stryker T/Pump Series

Pad / Hose Not Detected or No Circulation

On this page

Asset Type

Patient Warming System

Manufacturer

Stryker

Model

T/Pump Series

What This Guide Helps With

No warming, flow errors, or unit running without heat due to hose disconnection, pad issues, or restricted water circulation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify Reported Issue

Inspect Hose Connections

Check for Kinked or Obstructed Tubing

Inspect the Pad / Blanket

Verify Water Level in Reservoir

Check for Air in the System

Power Cycle the Unit

Test with a Known-Good Pad and Hose Set

If the Problem Persists

If circulation issues continue after verifying connections, tubing, water level, and accessories, the problem is likely internal (pump, flow sensor, or control system).

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot while connected to a patient. Always move the patient to another warming device before evaluating circulation or heating issues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit running but not warming patient; staff reports no heat output."

Cause

What was observed during troubleshooting.

Example:
"Kinked hose preventing water circulation between unit and pad."

Resolution

What action was taken.

Example:
"Straightened tubing, verified flow restored, and confirmed normal heating operation."

Helpful Details to Include (If Known)

Final Thought

Most circulation-related issues come from simple external causes like connections, tubing, or accessories. Focus on flow first, protect the patient, and escalate appropriately when internal faults are suspected.

That is successful troubleshooting.