What This Guide Helps With

Displayed temperature does not match expected warming output or fluctuates, often due to setup, airflow, or accessory-related issues rather than internal failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify Reported Issue

Power Cycle the Unit

Check Set Temperature and Mode

Inspect Hose and Blanket / Pad Connection

Verify Proper Airflow

Inspect Air Intake and Filter Area

Confirm Correct Blanket / Pad Type

Check for Environmental Factors

Observe for Consistency

If the Problem Persists

If all external causes have been ruled out, the issue is likely related to an internal temperature sensor or control system.

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Do not troubleshoot while the device is in use on a patient. Always transfer the patient to another warming method before evaluation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit reported inaccurate temperature readings and inconsistent warming."

Cause

What was observed during troubleshooting.

Example:
"Airflow restriction from kinked hose causing unstable temperature sensing."

Resolution

What action was taken.

Example:
"Corrected hose routing, verified stable temperature operation."

Helpful Details to Include (If Known)

Final Thought

Temperature accuracy is critical for safe patient warming. Always rule out airflow, setup, and environmental causes first before suspecting internal failure. Proper escalation and clear documentation ensure patient safety and efficient repair.

That is successful troubleshooting.