What This Guide Helps With

Addresses Tango+ failing self-test or calibration due to setup issues, accessories, or external factors affecting accuracy or system readiness.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Failure

Check Power Source and Stability

Inspect and Reseat Cuff and Tubing

Check for Environmental or Setup Issues

Perform Manual Calibration Procedure

Restart the Device

Check for Repeated Failure Patterns

If the Problem Persists

If all external causes have been ruled out and the Tango+ continues to fail self-test or calibration, the issue is likely internal (pressure sensor, pump, or control board).

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop prevents unnecessary troubleshooting and ensures patient safety.

Clinical Use Tip

Always perform calibration checks before starting a stress test. Never troubleshoot calibration issues with a patient connected.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Tango+ failing self-test and unable to complete calibration prior to stress test."

Cause

What was observed during troubleshooting.

Example:
"Calibration failure persisted after verifying power, tubing integrity, and performing manual calibration—likely internal pressure sensing issue."

Resolution

What action was taken.

Example:
"Removed device from service and sent to Clinical Engineering shop for further evaluation and repair."

Helpful Details to Include (If Known)

Final Thought

Calibration and self-test failures should always be approached by ruling out external factors first. Once those are eliminated, escalation is the correct and safe decision. Proper documentation ensures traceability and supports efficient repair.

That is successful troubleshooting.