Sysmex UF-1000i

Sample Aspiration / Probe Clog

On this page

Asset Type

Urine Analyzer

Manufacturer

Sysmex

Model

UF-1000i

What This Guide Helps With

Troubleshooting UF-1000i sample aspiration errors, probe clog warnings, low sample flow, or specimen pickup issues before repair escalation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the UF-1000i is not actively processing patient specimens before opening covers, handling racks, checking sample areas, or initiating cleaning.

Expected outcome: No patient specimen is interrupted, lost, or reported from a device with an unresolved aspiration issue.

Stop Additional Sample Processing

Ask laboratory staff not to load additional racks or specimens until the aspiration or clog issue is evaluated.

Confirm whether any specimens were in process when the error occurred and whether results were completed, flagged, rejected, or require repeat testing per laboratory policy.

Expected outcome: Incomplete or questionable urine results are not released.

Confirm the Exact Complaint

Ask staff what they observed. Determine whether the issue involves:

Expected outcome: The failure is narrowed to sample pickup, probe movement, specimen condition, or fluid path behavior.

Check the Specimen Cup and Sample Volume

Verify the urine specimen cup is properly filled, correctly positioned, and not underfilled.

Check for foam, heavy sediment, mucus, clots, crystals, or debris that may interfere with aspiration.

Expected outcome: A low-volume or obstructed specimen is identified before assuming analyzer failure.

If replacing or remixed specimen handling resolves the issue, stop troubleshooting and document the finding.

Verify Rack and Tube/Cup Positioning

Confirm the rack is the correct type, seated properly, and not damaged.

Check that specimen containers are sitting fully in the rack and are not tilted, cracked, overfilled, or placed in the wrong position.

Expected outcome: The probe can align with the specimen container without obstruction or mispositioning.

Inspect the Sample Probe Area Externally

With the device safely paused and access permitted, visually inspect the sample probe area for visible dried urine, crystals, residue, or obstruction.

Do not bend, force, scrape, or manually realign the probe.

Expected outcome: Any obvious external buildup or obstruction is identified without damaging the probe.

Check for Visible Probe Damage

Look for obvious signs of a bent, loose, cracked, or contaminated probe.

If the probe appears bent or physically damaged, remove the analyzer from service and escalate to repair.

Expected outcome: Physical probe damage is recognized as a repair issue, not a cleaning issue.

Run the User-Accessible Cleaning or Rinse Function

If available to Clinical Engineering and permitted by facility procedure, run the analyzer’s normal rinse, cleaning, or probe maintenance cycle.

Use only approved cleaning solutions and laboratory-approved maintenance materials.

Expected outcome: Minor residue, urine crystals, or partial blockage may clear through normal maintenance.

If aspiration returns to normal after cleaning, stop troubleshooting and document the cleaning action and successful test.

Check Reagent, Sheath, Diluent, and Waste Conditions

Verify required fluids are present, connected, not expired where applicable, and not empty.

Confirm waste containers are not full and that waste tubing is not kinked, pinched, or disconnected.

Expected outcome: Aspiration errors are not being caused by downstream fluid handling or pressure problems.

Inspect External Tubing and Fluid Path Connections

Check visible, user-accessible tubing for kinks, air gaps, loose fittings, leaks, crystallized residue, or pinch points.

Do not disconnect internal tubing unless authorized by service procedure and local policy.

Expected outcome: External fluid restrictions or leaks are ruled out.

Listen for Abnormal Pump or Aspiration Sounds

During a safe test or maintenance cycle, listen for unusual clicking, grinding, repeated priming, weak suction, or pump strain.

Expected outcome: Abnormal pump behavior may indicate an internal fluidics or drive issue requiring bench evaluation.

Check Error Timing

Determine when the error occurs:

Expected outcome: Repeated failure on all samples suggests an analyzer issue, while failure on certain specimens may point to container, volume, or sample condition.

Try a Known Good Control or Approved Test Specimen

Have laboratory staff run an approved control, maintenance sample, or known acceptable specimen according to laboratory policy.

Do not use patient testing as a trial if the analyzer’s reliability is questionable.

Expected outcome: A known good sample helps separate analyzer aspiration failure from poor specimen quality.

Check for Related QC or Scattergram Abnormalities

Ask staff whether QC failed, scattergrams looked abnormal, counts were unexpectedly low, or samples required frequent repeat testing before the clog error appeared.

Expected outcome: Trending quality issues may confirm a developing fluidic restriction rather than a single specimen problem.

Restart Only When Safe and Appropriate

If the analyzer is idle and results are protected, perform a normal software shutdown and restart if local procedure allows.

Do not power cycle during active sample processing.

Expected outcome: A temporary software or sensor state may clear, but repeated aspiration errors still require further evaluation.

Verify Operation After Corrective Action

After any cleaning, fluid correction, rack correction, or restart, have laboratory staff run QC or an approved verification process before returning the UF-1000i to patient testing.

Expected outcome: The analyzer is not returned to service until acceptable operation is confirmed.

If the Problem Persists

If the UF-1000i continues to show sample aspiration errors, probe clog messages, weak suction, repeated failed samples, abnormal fluid movement, or questionable results after external checks and approved cleaning, common external causes have been ruled out.

The analyzer should be:

Possible internal causes may include a clogged sample probe, internal tubing restriction, pump failure, valve issue, sensor problem, pressure/vacuum fault, or internal fluidics contamination.

Knowing when to stop is proper troubleshooting. Continued operation may risk delayed testing or unreliable urine particle results.

Clinical Use Tip

Do not troubleshoot aspiration or probe clog issues while patient specimens are actively being processed. If urine testing must continue, move testing to another validated analyzer or follow the laboratory’s approved downtime process first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory reported the Sysmex UF-1000i displayed sample aspiration errors and failed to process several urine specimens."

Cause

What was observed during troubleshooting.

Example:
"Inspection found adequate reagent levels but visible residue near the sample probe area and repeated aspiration failure during verification."

Resolution

What action was taken.

Example:
"Ran approved probe cleaning/rinse cycle, verified rack and sample cup positioning, had lab perform QC verification, and returned analyzer to service after successful operation."

Helpful Details to Include (If Known)

Final Thought

Sample aspiration and probe clog issues should be approached logically and safely. Rule out specimen condition, rack positioning, fluid levels, waste status, visible obstruction, and approved cleaning before assuming internal failure. If the analyzer cannot reliably aspirate known good specimens, remove it from service and escalate. Clear CCR documentation helps protect patient results and supports efficient repair.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide