On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting retained chamber water caused by blocked screens, kinked tubing, full waste containers, backed-up drains, or internal drainage faults.
Step-by-Step Troubleshooting
1. Protect Personnel and Processing Operations
Do not open plumbing connections or move the sterilizer while the chamber contains hot water or remains pressurized. Quarantine the load. Keep personnel clear of hot discharge paths. Allow the chamber to cool fully. Use another verified sterilizer for clinical processing. Treat an incomplete load as nonsterile.
Expected outcome: Personnel are protected from hot water and steam exposure.
2. Confirm the Drainage Problem
Determine whether:
- Water remains in the chamber after the cycle.
- The cycle stalls during exhaust or drain.
- Water backs up through the drain opening.
- The waste container does not receive discharge.
- The chamber drains slowly rather than not at all.
- The issue occurs with every cycle.
Expected outcome: The location and severity of the drainage restriction are identified.
3. Verify the Chamber Is Cool and Depressurized
Before inspection:
- Confirm the pressure indication has returned to a released state.
- Confirm no steam is venting.
- Allow retained water to cool.
- Electrically isolate the sterilizer when required.
Expected outcome: External drainage checks can be completed safely.
4. Inspect the External Drain Hose
Check the accessible hose for:
- Kinks.
- Pinching behind the sterilizer.
- Crushed sections.
- Sharp upward loops.
- Loose connections.
- Improper insertion into the receiving drain.
- Visible debris at the hose outlet.
- Do not disconnect a hot line.
Expected outcome: The external hose is open and correctly routed. If drainage returns normally, stop troubleshooting.
5. Check the Waste Container or Facility Drain
Confirm that:
- The collection container is not full.
- The hose outlet is not submerged.
- The facility drain accepts flow.
- No standing water is backing into the hose.
- The receiving drain is not shared with an obstructed appliance.
- Escalate building drain problems to Facilities.
Expected outcome: The receiving path is open. If the chamber drains normally after correction, stop troubleshooting.
6. Inspect the Accessible Chamber Drain Screen
After the chamber is cool:
- Remove trays.
- Locate the accessible drain screen or strainer.
- Remove loose lint, paper, packaging fragments, or visible debris.
- Clean only according to approved department practices.
- Do not insert tools into internal piping or valves.
Expected outcome: The accessible chamber drain opening is clear. If normal drainage returns, stop troubleshooting.
7. Check Loading Practices
Determine whether:
- Packaging material contacted the drain.
- Liners or wraps collapsed over the opening.
- The chamber was overloaded.
- Trays were installed incorrectly.
- Liquids spilled into the chamber.
- Excessive debris entered from instruments or containers.
Expected outcome: Loading practices do not introduce or trap material at the drain. If corrected loading prevents recurrence, stop troubleshooting.
8. Inspect for Mineral Buildup or Contamination
Look for:
- Visible scale around the chamber drain.
- Cloudy deposits in the reservoir.
- Sediment in accessible areas.
- Evidence that unapproved water was used.
- Do not perform chemical descaling unless authorized by approved manufacturer and facility procedures.
Expected outcome: Accessible contamination is documented and addressed within approved maintenance limits.
9. Perform a Controlled Drain Test
After correcting an external obstruction:
- Run an approved test cycle.
- Observe whether water enters and leaves the chamber normally.
- Confirm the exhaust phase progresses.
- Verify no leakage occurs beneath the sterilizer.
- Review the completed cycle record.
Expected outcome: The chamber drains completely and the cycle finishes normally. Troubleshooting can stop after quality-control checks pass.
10. Remove From Service When Drainage Remains Restricted
Remove the unit from service when:
- Water remains in the chamber.
- Drainage is unpredictable.
- The cycle repeatedly stalls.
- Water leaks beneath the unit.
- The facility drain is verified functional but the sterilizer does not drain.
- Internal blockage is suspected.
Expected outcome: A sterilizer with unreliable water removal is not used clinically.
If the Problem Persists
External tubing, receiving drains, loading, and accessible screens have been ruled out. Remaining categories may include internal drain plumbing, drain or exhaust valves, mineral buildup, water-level sensing, control circuitry, or improperly operating fill and exhaust systems. The sterilizer should be:
- Removed from service.
- Labeled Out of Service.
- Sent for repair or bench evaluation.
- Evaluated using appropriate Tuttnauer documentation and approved test equipment.
- Repaired or cleaned internally only by qualified personnel.
- Complete leak, drainage, cycle, and sterilization-quality testing before return to service. Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Never reach into retained chamber water until the chamber is confirmed depressurized and the water has cooled.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Sterile Processing reported standing water in the Tuttnauer 2540 chamber after completed cycles."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found packaging debris obstructing the accessible chamber drain screen."
Resolution
What action was taken.
Example:
"Removed the debris, cleaned the accessible drain area, completed an approved cycle with normal drainage, and returned the unit after quality testing passed."
Helpful Details to Include (If Known)
- Amount and location of retained water.
- Cycle phase where drainage stopped.
- External drain-hose condition.
- Waste-container level.
- Facility drain condition.
- Chamber drain-screen findings.
- Visible scale or sediment.
- Load type and packaging.
- Leakage beneath the unit.
- Test-cycle drainage result.
- Final quality-control results.
- Final device status.
Final Thought
Drainage problems should be approached from the receiving drain back toward the sterilizer. Verify external routing, containers, screens, loading, and water quality before escalating an internal restriction, and document all testing before release. That is successful troubleshooting.