On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting delayed pressure release caused by blocked drain paths, restricted tubing, improper routing, loading, or internal exhaust-control problems.
Step-by-Step Troubleshooting
1. Protect Personnel and Maintain Processing Continuity
Do not force the door open or disconnect tubing while the chamber is hot or pressurized. Keep personnel clear of the door and exhaust path. Quarantine the load. Allow the chamber to cool and depressurize naturally. Use another verified sterilizer for urgent processing. Treat the load as nonsterile if the cycle did not complete acceptably.
Expected outcome: Personnel and clinical operations are protected while the affected sterilizer is evaluated.
2. Confirm the Venting Failure
Determine whether:
- Pressure decreases slowly.
- Pressure remains above zero after the expected exhaust phase.
- The cycle stalls at exhaust.
- The display indicates completion while pressure remains.
- Steam discharges continuously or not at all.
- The issue occurs only with certain loads or cycles.
- Record the pressure trend and cycle phase.
Expected outcome: The reported condition is verified without opening the chamber under pressure.
3. Allow Complete Cooling Before Inspection
Confirm that:
- The pressure indication has returned to a safe released state.
- No steam is escaping.
- The door is cool enough to approach.
- The sterilizer is electrically isolated when inspection requires it.
- Do not rely solely on elapsed time.
Expected outcome: The sterilizer can be inspected without steam or burn exposure.
4. Inspect the External Drain and Exhaust Tubing
Check accessible tubing for:
- Kinks.
- Pinching.
- Sharp bends.
- Improper elevation.
- Loose connections.
- Crushed sections.
- Obstruction at the facility drain or collection container.
- Correct only accessible routing or connection problems.
Expected outcome: Tubing is open, secure, and properly routed. If normal venting returns, stop troubleshooting.
5. Check the Receiving Drain or Waste Container
Verify that:
- The facility drain is not backed up.
- A waste container is not full.
- The hose end is not submerged.
- The exhaust path is not blocked by debris.
- The receiving location can safely accept hot discharge.
Expected outcome: The external receiving path allows unrestricted flow. If the exhaust phase completes normally, stop troubleshooting.
6. Inspect Accessible Chamber Drain Areas
After cooling:
- Remove trays as needed.
- Inspect the accessible chamber drain screen or strainer.
- Remove loose lint, packaging material, or visible debris.
- Check for scale buildup around accessible openings.
- Do not probe or force internal valves.
Expected outcome: The accessible chamber drain path is clear. If the issue is resolved during testing, stop troubleshooting.
7. Evaluate Load Arrangement
Confirm that:
- The chamber was not overloaded.
- Packs were not placed over drain or vent openings.
- Trays were positioned correctly.
- Liquids, containers, or wrapped items matched the selected cycle.
- Packaging did not collapse into the chamber drain area.
Expected outcome: Load arrangement does not obstruct steam movement or exhaust. If proper loading restores normal venting, stop troubleshooting.
8. Observe an Approved Test Cycle
Run an unloaded or approved test cycle when permitted. Observe:
- Pressure decrease during exhaust.
- Sound of normal venting.
- Drain discharge.
- Display progression.
- Steam leakage from unintended areas.
- Whether the door releases only after pressure falls.
Expected outcome: Pressure returns normally to a released condition and the cycle advances. Troubleshooting can stop after required tests pass.
9. Check for Environmental Drainage Problems
When the sterilizer’s external tubing is clear but drainage remains slow:
- Check for facility drain restrictions.
- Compare operation using an approved alternate receiving arrangement, when permitted.
- Confirm nearby equipment is not sharing or pressurizing the same drain path improperly.
- Escalate building drain concerns to Facilities.
Expected outcome: The facility drainage path is verified as functional or routed for correction.
10. Remove From Service When Exhaust Remains Unreliable
Remove the sterilizer from service when:
- Pressure does not release predictably.
- The door remains interlocked for an extended period.
- Steam or hot water backs up.
- The drain path repeatedly blocks.
- Abnormal valve noise or continuous venting occurs.
- A controlled test cannot be completed safely.
Expected outcome: An unreliable pressure-release system is not used for clinical loads.
If the Problem Persists
External tubing, drain conditions, loading, and accessible obstructions have been ruled out. Remaining categories may include the exhaust valve, internal drain plumbing, pressure sensing, temperature sensing, control circuitry, or mineral buildup requiring service-level cleaning. The sterilizer should be:
- Removed from service.
- Labeled Out of Service.
- Sent for repair or bench evaluation.
- Evaluated using appropriate Tuttnauer documentation and approved test equipment.
- Repaired or configured only by qualified personnel.
- Complete required exhaust, pressure, cycle, and sterilization-quality testing before return to service. Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Keep personnel away from the door until chamber pressure is confirmed released, not merely until the cycle timer ends.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Sterile Processing reported that the Tuttnauer 2540 remained pressurized and exhausted very slowly after the sterilization phase."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the external drain hose sharply kinked behind the sterilizer."
Resolution
What action was taken.
Example:
"Corrected the hose routing, completed an approved cycle with normal pressure release, and returned the unit after quality-control testing passed."
Helpful Details to Include (If Known)
- Cycle selected.
- Pressure remaining at the end of exhaust.
- Time or phase where the cycle stalled.
- External hose condition.
- Receiving drain or container condition.
- Chamber drain-screen condition.
- Load arrangement.
- Steam or water discharge behavior.
- Door-release behavior.
- Test-cycle results.
- Final quality-control results.
- Final device status.
Final Thought
Slow exhaust must be treated as a pressure-safety concern. Verify external tubing, drains, loading, and accessible chamber obstructions before escalating possible internal valve or sensing problems, and document the final cycle verification clearly. That is successful troubleshooting.