United Imaging uMR 790

Exposure or Scan Will Not Start or Is Aborted

On this page

Asset Type

MRI System

Manufacturer

United Imaging

Model

uMR 790

What This Guide Helps With

Troubleshoots MRI scans that will not begin or terminate unexpectedly because of setup, interlocks, coils, communication, patient motion, or system readiness.

Step-by-Step Troubleshooting

1. Protect the Patient and Stop Repeated Scan Attempts

Do not repeatedly attempt an unreliable examination while a patient depends on the system.

Assess the patient, remove them from the scanner if necessary, and arrange continuity of care if the examination cannot proceed safely.

Expected outcome: The patient is safe and troubleshooting can proceed without unnecessary repeated scan attempts.

2. Confirm When the Scan Fails

Determine whether the scan:

Record the exact displayed message or sequence of events.

Expected outcome: The failure point is identified and can guide external troubleshooting.

3. Verify Overall System Readiness

Confirm the scanner, workstation, table position, selected hardware, and acquisition system are all showing normal readiness.

Resolve any obvious startup or accessory-recognition issue before testing another scan.

Expected outcome: No general readiness fault remains. If restoring readiness allows the scan to start normally, verify operation and stop.

4. Check Patient and Table Positioning

Verify:

Expected outcome: Positioning and setup are correct. If correcting the setup allows scanning to proceed, confirm successful acquisition and stop.

5. Inspect Coil and Accessory Connections

Check the active coil and other normal external accessories for secure connection and visible damage.

If appropriate, reseat connections or compare with a known-good compatible accessory.

Expected outcome: Required acquisition accessories remain recognized and stable. If a faulty accessory is isolated, remove it from service and verify the scanner with a known-good component.

6. Review Normal Exam and Sequence Selections

Confirm the intended protocol or sequence has been selected appropriately and no obvious operator-entered parameter or workflow selection is preventing execution.

Do not alter protected or service-level configuration to force a sequence to run.

Expected outcome: The examination setup is valid. If correcting a normal workflow selection resolves the problem, complete scan verification and stop.

7. Check for Patient Motion or Communication Events

Determine whether the abort corresponds with:

Expected outcome: Any external reason for the abort is identified. If corrected and the scan subsequently completes normally, troubleshooting can stop.

8. Check Environmental and Cooling Status

Look for active temperature, cooling, ventilation, or equipment-room issues that might cause the system to inhibit or terminate scanning.

Do not continue scanning through an overtemperature or cooling-related condition.

Expected outcome: Environmental support is normal. If a facility issue is found, keep the MRI out of service until corrected.

9. Perform a Controlled Test Scan

After correcting an identified external cause, use an approved test object or appropriate nonpatient verification procedure.

Confirm:

Expected outcome: A controlled scan completes successfully. Troubleshooting can stop.

10. Escalate Repeated or Unexplained Scan Aborts

If scan setup, patient positioning, external accessories, system readiness, and environment are correct but scanning continues to fail, stop troubleshooting.

Expected outcome: The system is removed from clinical service pending service evaluation.

If the Problem Persists

External causes have been ruled out. Remaining categories may include RF, gradient, acquisition, control, cooling, timing, internal communication, or software/configuration problems requiring qualified MRI service.

The device should be:

Do not repeatedly attempt failed scans or enter unsupported service functions.

Knowing when to stop external troubleshooting is proper troubleshooting. Successful service must be followed by appropriate imaging and operational verification before return to clinical use.

Clinical Use Tip

If a scan repeatedly aborts without an obvious patient-related reason, move the patient to a safe alternative plan rather than repeatedly restarting the examination.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"MRI staff reported that the uMR 790 scan repeatedly aborted immediately after acquisition began."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found an intermittently seated external coil connection that caused the selected coil to lose readiness."

Resolution

What action was taken.

Example:
"The connection was corrected, the coil remained recognized, and a controlled scan completed without interruption before return to service."

Helpful Details to Include (If Known)

Final Thought

A scan abort should first be treated as a safety and workflow problem, then traced through positioning, accessories, readiness, and environment before internal failure is suspected. Verify a successful controlled scan and document the event clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide