United Imaging uMR 790

Study Will Not Save, Reconstruct, or Export

On this page

Asset Type

MRI System

Manufacturer

United Imaging

Model

uMR 790

What This Guide Helps With

Troubleshoots failed study saving, reconstruction, or export caused by workflow, storage, workstation, network, study-specific, or communication problems.

Step-by-Step Troubleshooting

1. Protect Patient Data Before Making Changes

Do not delete studies, clear storage, restart unnecessarily, or recreate patient records until the status of acquired data is understood.

If the patient is still present, confirm whether additional imaging is clinically necessary before removing them from the department.

Expected outcome: Existing patient data is preserved and unnecessary repeat imaging is avoided.

2. Identify Which Part of the Workflow Failed

Determine whether the study:

Record the exact displayed message.

Expected outcome: The failed stage of the imaging workflow is identified.

3. Confirm Workstation Responsiveness

Verify the workstation remains responsive and the MRI application is functioning normally.

Check for:

Expected outcome: The workstation is stable enough to evaluate the study. If an approved normal restart is required, protect data first.

4. Verify the Study Exists Locally

Using normal clinical functions, confirm whether acquired images or series are present in local study storage.

Do not delete partial or failed data during troubleshooting.

Expected outcome: The available local data is identified and preserved.

5. Determine Whether the Problem Is Study-Specific

Check whether:

Use nonpatient testing when practical.

Expected outcome: The issue is narrowed to one study or a broader system condition.

6. Check Export and Network Availability

If saving and reconstruction work but export fails, inspect the external network connection and determine whether PACS or the selected destination is available.

Coordinate with IT or PACS support when appropriate.

Expected outcome: The export pathway is confirmed available. If network restoration allows successful export, verify destination receipt and stop.

7. Check Normal Destination and Export Selection

Confirm that the intended destination and normal export workflow are selected.

Do not modify protected DICOM or network configuration without authorization.

Expected outcome: The correct destination is being used. If correcting workflow selection resolves export, verify receipt and stop.

8. Check for Obvious Storage or System Warnings

Review normal user-visible system messages for indications related to local study storage or processing availability.

Do not manually purge storage or modify file systems without approved procedures and appropriate data safeguards.

Expected outcome: No unresolved user-visible storage condition remains. A service-level storage issue is escalated rather than altered experimentally.

9. Perform a Controlled Workflow Verification

After correcting an identified external cause, perform an approved nonpatient test and confirm:

Expected outcome: The complete workflow operates successfully. Troubleshooting can stop.

10. Escalate Persistent Save, Reconstruction, or Export Failure

If workstation operation, local data presence, network connectivity, destination availability, and normal workflow have been checked but processing still fails, stop troubleshooting.

Expected outcome: The system is removed from clinical use as appropriate and escalated without risking patient data.

If the Problem Persists

Common external and workflow causes have been ruled out. Remaining categories may include storage hardware, reconstruction services, application software, databases, internal communication, system configuration, or other service-level computing functions.

The device should be:

Preserve affected studies and logs according to local procedures before service actions that could alter stored data.

Knowing when to stop external troubleshooting is proper troubleshooting. Return the system to clinical service only after the full acquisition-to-save-to-reconstruction-to-export workflow has been verified.

Clinical Use Tip

Before releasing a patient after an unusual save or reconstruction failure, confirm that the clinically required images are actually stored and usable so avoidable repeat imaging is not needed.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"MRI staff reported that a uMR 790 study reconstructed normally but would not export to PACS."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the workstation's external network connection was loose, interrupting communication with the configured destination."

Resolution

What action was taken.

Example:
"The connection was secured, the affected study exported successfully, receipt at PACS was confirmed, and the complete workflow was verified."

Helpful Details to Include (If Known)

Final Thought

Study-processing failures require protecting patient data before troubleshooting anything else. Identify exactly where the workflow stops, verify workstation and network conditions before assuming internal failure, and confirm the complete imaging-data path after correction.

That is successful troubleshooting.