United Imaging uMR 790

Workstation Freezes or Loses Communication

On this page

Asset Type

MRI System

Manufacturer

United Imaging

Model

uMR 790

What This Guide Helps With

Troubleshoots console freezing, application unresponsiveness, and communication loss caused by workstation, cabling, network, peripherals, startup, or infrastructure conditions.

Step-by-Step Troubleshooting

1. Protect the Patient and Stop Dependence on the Frozen Console

If a patient examination is active, do not continue using an unreliable workstation or communication path.

Safely end or transfer the clinical workflow according to local procedures before restarting or disconnecting equipment.

Expected outcome: The patient is no longer dependent on an unstable console or communication path.

2. Define the Failure

Determine whether:

Record any displayed message.

Expected outcome: The scope of the problem is clear.

3. Check Local Workstation Power and Peripherals

Verify:

Expected outcome: Basic workstation hardware and accessible connections appear normal. If restoring a peripheral connection resolves the issue, verify operation and stop.

4. Check External Communication Connections

Inspect accessible network and system communication cables for secure seating, damage, strain, or recent movement.

Do not disturb internal scanner communication hardware.

Expected outcome: External communication connections are secure and undamaged. If reseating an approved connection restores stable communication, verify and stop.

5. Determine Whether the Problem Is Local or Infrastructure-Wide

Check whether other network-dependent MRI functions are affected.

If multiple devices or hospital systems are also experiencing communication problems, coordinate with IT or network support rather than assuming scanner failure.

Expected outcome: The problem is identified as local to the MRI system or part of a wider infrastructure issue.

6. Perform an Approved Normal Restart

If the system is not actively supporting a patient and site procedures permit, perform a normal workstation or application restart.

Avoid repeated forced shutdowns.

Expected outcome: The workstation restarts normally and restores stable communication. If so, continue to verification and stop.

7. Check for Recurring Freeze Conditions

After restart, observe whether:

If the workstation freezes again, do not continue cycling power.

Expected outcome: Operation remains stable. If instability returns, remove the system from service.

8. Check Environmental Conditions

Inspect for excessive heat, blocked workstation ventilation, liquid exposure, or unusual equipment-room conditions.

Expected outcome: No environmental problem is affecting the console. If overheating or contamination is suspected, discontinue use until evaluated.

9. Perform Final Communication Verification

Confirm:

Expected outcome: Communication remains stable and the workstation is responsive. Troubleshooting can stop.

10. Escalate Persistent Freezing or Communication Loss

If power, accessible cabling, peripherals, network availability, and normal restart have been checked and the condition persists, stop troubleshooting.

Expected outcome: The system is taken out of service for qualified evaluation.

If the Problem Persists

Common external causes have been ruled out. Remaining categories may include workstation hardware, software, storage, internal communication interfaces, system controllers, operating-system problems, network configuration, or infrastructure issues requiring specialized support.

The device should be:

Coordinate with hospital IT when network infrastructure may contribute.

Knowing when to stop external troubleshooting is proper troubleshooting. Confirm stable workstation and scanner communication before return to service.

Clinical Use Tip

Do not repeatedly force-restart the MRI workstation during an unstable examination; preserve study status and patient safety before restarting system components.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"MRI staff reported that the uMR 790 console lost communication with the scanner during exam setup."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found an accessible network connection at the workstation was partially seated."

Resolution

What action was taken.

Example:
"The connection was secured, workstation-to-scanner communication remained stable, and normal exam setup functionality was verified."

Helpful Details to Include (If Known)

Final Thought

Workstation problems should be separated into local console, external connection, scanner communication, and hospital infrastructure causes before internal failure is assumed. Protect active studies, verify stable operation after correction, and document the full communication path.

That is successful troubleshooting.