What This Guide Helps With

Imaging detector or acquisition hardware does not become ready, is unavailable, or prevents required treatment imaging from proceeding.

Step-by-Step Troubleshooting

1. Protect the Patient and Suspend the Imaging Workflow

Do not proceed with treatment when required imaging or acquisition hardware cannot be relied upon. Move the patient to a safe condition and coordinate with Radiation Oncology before continuing.

Expected outcome: No treatment proceeds without the imaging capability required by the clinical workflow.

2. Identify the Affected Imaging Component

Determine which imaging or acquisition function is reported unavailable and whether the problem occurs during initialization, positioning, image acquisition, or readiness confirmation. Record all displayed messages.

Expected outcome: The affected imaging function and failure point are clearly identified.

3. Check External Positioning and Clearance

Inspect externally visible imaging components for correct deployed or parked position as applicable. Verify that accessories, immobilization equipment, cables, or other objects are not blocking movement or positioning.

Expected outcome: Imaging hardware has appropriate clearance and no external obstruction is present.

4. Inspect Accessible Connections

Check accessible external cables, workstation connections, and supporting communication connections for looseness or visible damage. Do not disconnect internal or restricted-service connections.

Expected outcome: Accessible imaging-system connections are secure and intact.

5. Verify Supporting Workstations

Confirm that the imaging-related workstation, display, and associated software interface have completed startup and are responsive.

Expected outcome: Supporting acquisition systems are powered and operating normally.

6. Check System Readiness and Workflow State

Verify that the overall TrueBeam system has completed initialization and is in an appropriate state for imaging. Confirm that no unresolved safety condition, incomplete startup process, or positioning state is preventing readiness.

Expected outcome: The system is correctly prepared for acquisition.

7. Perform an Approved Restart if Appropriate

If external checks are normal and facility procedures allow, perform the approved restart of the affected user-accessible subsystem or system. Avoid repeated restart attempts if the same failure returns.

Expected outcome: Acquisition hardware initializes and reports ready. If so, continue to functional verification.

8. Verify Imaging Function

With no patient depending on an unreliable system, perform the appropriate facility-approved imaging verification or quality-control check.

Expected outcome: Imaging hardware becomes ready and successfully completes the required verification. Troubleshooting can stop if results are acceptable.

9. Remove From Service if Readiness Cannot Be Confirmed

If the detector or acquisition system remains unavailable, intermittently drops out, or fails verification, stop external troubleshooting.

Expected outcome: The system is withheld from treatment workflows requiring the affected imaging capability.

If the Problem Persists

External positioning, connections, system state, supporting workstation operation, and basic workflow conditions have been ruled out. The remaining problem may involve imaging electronics, detector hardware, motion/position feedback, communications, calibration data, software, or another internal service-level subsystem.

The device should be:

Any imaging subsystem involved in patient positioning or treatment verification must pass required functional and quality checks before return to service.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Do not bypass required image-guidance steps simply because the treatment-delivery portion of the system appears operational.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Radiation Oncology reported that the TrueBeam imaging system remained not ready and imaging acquisition could not begin."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found an accessible acquisition workstation cable connection loose at the external interface."

Resolution

What action was taken.

Example:
"Clinical Engineering secured the connection, restarted the affected workflow, and verified imaging-system readiness and successful functional imaging verification."

Helpful Details to Include (If Known)

Final Thought

Required imaging is part of safe radiation-treatment delivery. Rule out positioning, external connections, workstation status, and workflow conditions first, then escalate persistent readiness failures rather than assuming a specific internal component has failed.

That is successful troubleshooting.