Varian TrueBeam

Study Will Not Save, Reconstruct, or Export

On this page

Asset Type

Radiation Therapy System

Manufacturer

Varian

Model

TrueBeam

What This Guide Helps With

Imaging or study data will not save, reconstruct, complete processing, or export to the required clinical destination.

Step-by-Step Troubleshooting

1. Protect Patient Data and Clinical Continuity

Do not delete, overwrite, or repeatedly recreate clinically important data while troubleshooting. Confirm the patient is safe and notify Radiation Oncology if the failed study affects treatment decisions or workflow.

Expected outcome: Patient care continues safely and potentially recoverable clinical data is preserved.

2. Identify Which Function Failed

Determine whether the problem is saving, reconstruction, local processing, export, or transfer to another system. Record the exact message and the last successfully completed step.

Expected outcome: The failed portion of the data workflow is clearly identified.

3. Confirm the Workstation Is Responsive

Check whether the associated workstation and application respond normally or whether the system is frozen, unusually slow, or reporting other faults.

Expected outcome: A workstation-performance issue is distinguished from a study-specific failure.

4. Check Whether the Problem Is Study-Specific

Determine whether the issue affects only one study or all new studies. Do not alter or delete patient records simply to test the system.

Use approved test data when available.

Expected outcome: The problem is narrowed to one dataset or the overall processing workflow.

5. Verify Required Network Connectivity for Export

If export is affected, inspect accessible network connections and confirm that the intended destination is online and reachable through approved methods.

Expected outcome: A basic network or destination outage is either identified or ruled out.

6. Verify the Intended Export Destination

Confirm that the correct configured clinical destination is selected and available. Do not modify restricted DICOM, storage, or network settings without authorization.

Expected outcome: The expected destination is correctly selected and active.

7. Check for Obvious Storage or System Messages

Review user-accessible notifications for storage, processing, communication, or application conditions. Do not manually remove protected system files or patient data to free space.

Expected outcome: Any obvious user-visible storage or processing condition is identified without compromising data integrity.

8. Perform an Approved Restart if Needed

If the workstation or application appears stalled and the patient is no longer dependent on it, follow the approved restart process. Preserve data according to facility and manufacturer procedures.

Expected outcome: Processing resumes normally after restart or the failure remains reproducible.

9. Perform a Controlled Save, Reconstruction, or Export Test

Use approved test data to verify the affected function and confirm completion at the intended endpoint when applicable.

Expected outcome: The study processes successfully through the full required workflow. If it does, troubleshooting can stop.

10. Escalate Persistent Data-Handling Failures

If studies repeatedly fail to save, reconstruct, or export, if data integrity is uncertain, or if storage cannot be verified, discontinue the affected workflow.

Expected outcome: Unreliable data handling is prevented from affecting patient treatment.

If the Problem Persists

External network connectivity, destination availability, workstation responsiveness, and basic workflow selection have been ruled out. Remaining causes may involve application software, storage systems, reconstruction services, databases, internal communications, DICOM services, or other service-level infrastructure.

The device or affected workflow should be:

Return to service requires successful saving, processing, reconstruction, and transfer testing for the functions affected, with data integrity confirmed.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Never delete a failed or incomplete clinical study simply to clear the workflow until the care team confirms that the data is no longer required.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Radiation Oncology reported that TrueBeam imaging studies completed locally but would not export to the configured destination."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the destination clinical system unavailable due to a network-side service interruption."

Resolution

What action was taken.

Example:
"Clinical Engineering coordinated restoration with IT and verified successful export and receipt of an approved test study before returning the workflow to service."

Helpful Details to Include (If Known)

Final Thought

Data-processing failures require protection of both the patient and the clinical record. Preserve existing data, verify the workstation, network, and destination before assuming an internal failure, and escalate whenever study integrity or reliable completion cannot be confirmed.

That is successful troubleshooting.