On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots oxygen source, tubing, connection, supply, concentration monitoring, configuration, and integrated oxygen-related performance problems.
Step-by-Step Troubleshooting
1. Maintain Safe Oxygenation and Ventilation
If oxygen delivery or monitored oxygen concentration is questionable while the VOCSN is supporting a patient, transfer the patient to an appropriate verified oxygen and ventilation source according to clinical procedure.
Do not troubleshoot unreliable oxygen delivery while a patient depends on it.
Expected outcome: The patient receives stable ventilation and oxygen independently of the suspect VOCSN.
2. Identify the Oxygen Configuration and Complaint
Determine whether the VOCSN was using:
- External high-pressure oxygen
- External low-pressure oxygen
- An integrated oxygen capability, if equipped
- Another approved configuration
Record the oxygen-related message, monitored concentration behavior, and whether the problem is continuous or intermittent.
Expected outcome: The affected oxygen source and symptom are clearly identified.
3. Verify the External Oxygen Supply
When external oxygen is used, verify:
- Source is connected
- Source is turned on
- Supply is not depleted
- Regulator or facility source is functioning
- Oxygen hose is fully connected
- Hose is not kinked, damaged, or contaminated
If a wall source is involved, compare with another verified source or have the medical gas infrastructure checked when appropriate.
Expected outcome: A known-good oxygen source reaches the VOCSN through intact external components.
If correcting the external source restores stable oxygen performance, proceed to final verification.
4. Inspect Oxygen Connections and Accessories
Inspect accessible oxygen fittings, hoses, adapters, and circuit components for looseness, incorrect connection, physical damage, or leaks.
Verify the external oxygen connection being used is appropriate for the configured therapy.
Expected outcome: Oxygen accessories are intact, correctly connected, and appropriate for the configuration.
If replacing damaged external oxygen tubing resolves the issue, troubleshooting can stop after verification.
5. Verify Device and Circuit Configuration
Confirm the installed circuit and oxygen configuration correspond to the settings in use.
Do not change prescribed therapy settings merely to make an alarm disappear. Configuration changes should be limited to authorized setup or controlled Clinical Engineering testing.
Expected outcome: The VOCSN configuration matches the actual oxygen source and circuit arrangement.
6. Compare Monitored Oxygen With an Independent Analyzer
Under controlled bench conditions, use an appropriate calibrated oxygen analyzer to compare delivered oxygen concentration with the VOCSN's monitored value according to manufacturer procedures.
Allow the system to stabilize before interpreting results.
Expected outcome: Independent and device-reported oxygen measurements demonstrate acceptable agreement under the applicable manufacturer checkout procedure.
If the system performs correctly after correcting an external source or configuration problem, the issue is resolved.
7. Verify Oxygen Alarms and Final Function
Verify oxygen delivery remains stable through an appropriate test period and confirm relevant alarms respond when the oxygen source is intentionally removed or changed as allowed by the manufacturer test procedure.
Expected outcome: Oxygen delivery, monitoring, and alarms operate correctly with no unresolved messages.
If these requirements are met, troubleshooting can stop.
If the Problem Persists
If source supply, hoses, fittings, configuration, circuit components, and independent oxygen measurement have been verified, the fault may involve internal oxygen sensing, blending, concentrator performance on equipped units, pneumatic controls, or calibration.
The VOCSN should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated with current Ventec Life Systems documentation and calibrated oxygen test equipment
- Repaired or calibrated only by qualified personnel
Do not make unauthorized oxygen calibration or internal pneumatic adjustments. After service, verify oxygen concentration, ventilation performance, and associated alarms before clinical use.
Knowing when to stop external troubleshooting and escalate is proper troubleshooting.
Clinical Use Tip
When oxygen delivery is uncertain, verify the complete path from source to patient rather than relying only on the setting displayed by the ventilator.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Respiratory Therapy reported low monitored oxygen concentration while the VOCSN was connected to an external oxygen source."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the external oxygen hose incompletely seated at the device connection."
Resolution
What action was taken.
Example:
"Reconnected the oxygen hose securely and verified stable oxygen concentration with an independent analyzer and successful alarm checks before return to service."
Helpful Details to Include (If Known)
- Oxygen source type
- Exact oxygen message or alarm
- Source pressure or availability status
- Hose and fitting condition
- Circuit configuration
- Device oxygen configuration
- Independent analyzer result
- Whether problem occurred with another oxygen source
- Alarm verification results
- Final device status
Final Thought
Oxygen complaints require verification of both the supply and the device. Protect the patient, confirm the source, connections, configuration, and measured output before assuming an internal oxygen-system failure, and escalate whenever accurate oxygen delivery cannot be verified.
That is successful troubleshooting.