Verathon BladderScan BVI 9400

Printer or results documentation failure

On this page

Asset Type

Bladder Scanner

Manufacturer

Verathon

Model

BladderScan BVI 9400

What This Guide Helps With

Troubleshooting failed printouts, missing scan results, poor print quality, or documentation problems after bladder volume measurement.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Do not delay patient care while troubleshooting a documentation issue.

Action:

Expected outcome: Patient care continues without relying on incomplete or missing documentation.

Why it matters: A bladder scanner may still scan correctly, but missing or incorrect documentation can affect clinical decisions and charting accuracy.

If this resolves the immediate clinical need, stop and document the issue.

Verify the Reported Documentation Failure

Confirm exactly what is failing before assuming a printer problem.

Action:

Ask clinical staff what occurred:

Perform a test scan or review a recent stored result if available.

Expected outcome: The failure is narrowed to printing, result storage, user workflow, or scanner function.

Why it matters: “Printer not working” may actually be a paper issue, battery issue, user workflow issue, or failed scan documentation process.

Confirm the Scan Result Is Actually Available

Verify that the device produced a valid bladder volume result.

Action:

Expected outcome: A valid scan result is visible and available for printing or documentation.

If no valid result exists, troubleshoot the scan quality or probe issue first and stop.

Check Battery Charge and Power Status

Low battery or unstable power can interfere with printing or result handling.

Action:

Expected outcome: The unit has enough power to complete scan documentation and printer operation.

If the printer works after charging, return the device to service after confirming normal operation.

Inspect Printer Paper Installation

Check the most common external cause first.

Action:

Expected outcome: Paper is loaded correctly and can advance normally.

If correcting the paper roll restores printing, stop and document the resolution.

Check for Paper Jam or Debris

A small paper obstruction can prevent printing or cause partial printouts.

Action:

Expected outcome: Paper feeds smoothly without binding.

If the jam is cleared and printouts are normal, return the unit to service.

Verify Thermal Paper Condition

Blank or faded printouts may be caused by incorrect or degraded paper.

Action:

Expected outcome: Printout becomes readable with proper thermal paper.

If print quality improves, document the paper replacement and return the device to service.

Check Printer Door or Cover Closure

The printer may not operate if the cover is not fully latched.

Action:

Expected outcome: The printer recognizes the paper compartment as closed and prints normally.

If reseating the cover resolves the issue, stop and document the finding.

Confirm Staff Printing Workflow

Make sure the failure is not caused by a missed menu step or timing issue.

Action:

Expected outcome: The scanner is being used in the correct sequence for result documentation.

If workflow correction resolves the problem, provide user guidance and document the education given.

Check Date, Time, and Result Information

Documentation may be considered failed if printouts are missing correct identifiers or timestamps.

Action:

Expected outcome: Printed results contain usable scan information for clinical documentation.

If settings or workflow correction resolves the issue, document what was corrected.

Power Cycle and Retest

A temporary software or printer communication issue may clear with a controlled restart.

Action:

Expected outcome: The printer and result documentation function normally after restart.

If normal operation returns, monitor for recurrence and document the successful retest.

Inspect for Physical Damage

Printer or documentation failures may follow drops, fluid exposure, or rough handling.

Action:

Inspect the scanner housing, printer cover, paper door, keypad/buttons, display, charging contacts, and probe connection.

Look for:

Expected outcome: No visible damage is present that would explain printer or system malfunction.

If damage is found, remove the device from service and send it for repair evaluation.

Test With Known-Good Consumables and Accessories

Rule out simple external causes before escalating.

Action:

Expected outcome: The device prints or documents correctly with known-good external items.

If the issue remains after known-good paper and power are verified, escalation is appropriate.

If the Problem Persists

If the BladderScan BVI 9400 still will not print, produces blank or unreadable results, loses scan documentation, or fails to handle results after paper, battery, workflow, and visible damage checks have been completed, the likely cause is internal printer, keypad, software, memory, or control electronics failure.

The device should be:

Knowing when to stop is proper troubleshooting. Do not disassemble the printer or attempt internal board-level repair during clinical troubleshooting.

Clinical Use Tip

Do not troubleshoot documentation problems during active patient care if the scan result is needed for immediate decision-making. Move to another bladder scanner or approved clinical process first. A result that cannot be printed, saved, or reliably documented should be treated as incomplete from a clinical workflow standpoint.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the Verathon BladderScan BVI 9400 would scan but would not print the bladder volume result."

Cause

What was observed during troubleshooting.

Example:
"Troubleshooting found the thermal paper roll was installed backward, causing blank printouts."

Resolution

What action was taken.

Example:
"Reinstalled the paper roll correctly, performed a test scan printout, verified readable results, and returned the scanner to service."

Helpful Details to Include (If Known)

Final Thought

Printer and result documentation failures should be approached logically. Confirm patient safety first, then verify the scan result, power status, paper condition, printer cover, workflow, and visible damage before escalating. Good documentation matters because even a correct bladder volume measurement has limited value if it cannot be reliably recorded.

That is successful troubleshooting.