Verathon BladderScan i10

Battery Not Charging or Short Runtime

On this page

Asset Type

Bladder Scanner

Manufacturer

Verathon

Model

BladderScan i10

What This Guide Helps With

Troubleshooting charging failure, short battery runtime, docking issues, power adapter problems, or battery status concerns on the BladderScan i10.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Do not troubleshoot charging or runtime issues while the device is needed for active patient care.

Action:

Expected outcome: Patient care continues without relying on a scanner that may lose power during use.

Why it matters: A bladder scanner that powers off mid-scan can delay assessment, documentation, or catheterization decisions.

Verify the Reported Charging or Runtime Issue

Confirm what clinical staff are actually seeing.

Action:

Expected outcome: The complaint is clearly identified before replacing accessories or escalating for repair.

If the issue was due to misunderstanding normal battery status or charge indication, educate staff and return the device to service.

Inspect the Power Source

Check the wall outlet before assuming a charger or device failure.

Action:

Expected outcome: The charging setup has confirmed incoming AC power.

If the outlet is dead or loose, move the charger to a working outlet and notify Facilities if needed.

Inspect the AC Power Adapter and Cord

Look for obvious external damage.

Action:

Expected outcome: The power supply and cord appear safe and connected.

If the adapter or cord is damaged, remove it from service and replace it with an approved Verathon-compatible power supply.

Check the Charging Dock or Charging Base

If the i10 uses a dock or charging cradle, verify the device is seating correctly.

Action:

Expected outcome: The scanner sits securely and begins charging.

If reseating the device restores charging, allow it to charge and verify normal runtime before returning it to service.

Inspect Charging Contacts and External Surfaces

Dirty or obstructed contacts can prevent charging.

Action:

Expected outcome: Charging contacts are clean, dry, and unobstructed.

If cleaning restores charging, document the finding and return the unit after confirming stable charging.

Allow Adequate Charge Time

A deeply discharged battery may need time before normal operation returns.

Action:

Expected outcome: Battery level increases and the device powers on normally.

If the scanner begins charging normally, continue charging and verify runtime before returning to the department.

Check for Intermittent Charging

Charging may fail only when the cable, dock, or scanner is moved.

Action:

Expected outcome: Charging remains stable without flickering or dropping out.

If charging starts and stops with movement, suspect a loose dock, damaged cable, poor contact, or internal charging port issue.

Compare With a Known-Good Charger or Dock If Available

Use substitution to isolate the fault.

Action:

Expected outcome:

If substitution identifies a bad external accessory, replace the accessory and verify charging.

Evaluate Runtime After a Full Charge

Short runtime should be verified before assuming battery failure.

Action:

Expected outcome: Runtime is consistent with normal clinical use.

If runtime is still short after a full charge, the battery may be degraded or the device may have an internal power issue.

Check for Heat, Odor, Swelling, or Physical Damage

Battery or charging problems with physical symptoms should not be ignored.

Action:

Expected outcome: No visible safety concerns are present.

If heat, odor, swelling, liquid damage, or electrical damage is found, remove the device from service immediately and do not continue charging attempts.

Verify Final Device Status

Before returning the scanner to service, confirm the charging issue is resolved.

Action:

Expected outcome: The BladderScan i10 charges reliably and has acceptable runtime for clinical use.

If the issue is resolved, document the work order and return the device to service.

If the Problem Persists

If the BladderScan i10 still does not charge, only charges intermittently, or has short runtime after outlet, charger, dock, contacts, seating, and charge-time checks, common external causes have been ruled out.

The device should be removed from service, labeled Out of Service, and sent for repair, battery evaluation, or bench service.

Do not continue repeated charging attempts if there is heat, odor, swelling, liquid intrusion, or suspected electrical damage.

Knowing when to stop is proper troubleshooting. A persistent charging or runtime issue may involve an internal battery, charging circuit, connector, dock interface, or power management problem that requires repair-level evaluation.

Clinical Use Tip

Do not troubleshoot battery or charging issues while the scanner is needed for patient care. Move to another bladder scanner first, especially if the i10 shuts down unexpectedly or cannot complete scans reliably. Maintain therapy continuity by ensuring bladder volume assessment continues with a backup device or approved clinical workflow.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the Verathon BladderScan i10 would not hold a charge and powered off shortly after removal from the charging dock."

Cause

What was observed during troubleshooting.

Example:
"Verified outlet power and charger connection; scanner charged intermittently due to poor seating and debris on the dock contact area."

Resolution

What action was taken.

Example:
"Cleaned exterior charging contact area per approved procedure, reseated scanner in dock, confirmed stable charging indicator and normal power-on operation."

Helpful Details to Include (If Known)

Final Thought

Battery troubleshooting should stay logical and safety-focused. Confirm the simple external causes first: outlet, power adapter, dock seating, contacts, and charge time. If those checks do not restore reliable charging or runtime, remove the BladderScan i10 from service and escalate for proper repair evaluation. Clear CCR documentation helps show exactly what was ruled out and why the device was returned to service or sent for repair.

That is successful troubleshooting.