Verathon BladderScan i10

Poor bladder detection or inaccurate volume reading

On this page

Asset Type

Bladder Scanner

Manufacturer

Verathon

Model

BladderScan i10

What This Guide Helps With

Troubleshooting poor bladder detection, low-confidence scans, or bladder volume readings that appear inconsistent with patient condition.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Do not troubleshoot scan accuracy on an active clinical concern without involving the care team.

Action:

Expected outcome: Patient care decisions are not based on a suspected inaccurate scan.

Why it matters: A poor bladder scan can lead to delayed catheterization, unnecessary catheterization, or incorrect assessment of urinary retention.

Verify the Reported Problem

Confirm whether the issue is poor bladder detection, inconsistent readings, or clearly inaccurate volume values.

Action:

Ask staff what they observed:

Expected outcome: The failure pattern is identified before assuming a device problem.

If the problem only occurred once or on a difficult patient scan, the issue may be technique, anatomy, positioning, or patient condition rather than device failure.

Remove the Device From Active Use for Testing

Take the BladderScan i10 out of patient care workflow before evaluating it.

Action:

Expected outcome: The device can be checked without delaying care or contaminating equipment.

Inspect the Probe and Scanner Exterior

Check for obvious external damage.

Action:

Inspect the probe face for:

Inspect the scanner body, touchscreen, charging contacts, and any connected accessories.

Expected outcome: The scanner and probe should be clean, intact, and free of visible damage.

If the probe face is cracked, damaged, contaminated, or physically compromised, remove the device from service and send it for repair evaluation.

Confirm Proper Cleaning of the Probe Face

A dirty or damaged probe surface can interfere with ultrasound transmission.

Action:

Expected outcome: The probe surface is clean and smooth before retesting.

If cleaning resolves the poor detection or inconsistent readings, return the device to service and document the cause.

Check Ultrasound Gel Use

Poor gel contact is one of the most common external causes of poor bladder detection.

Action:

Expected outcome: The probe should have full acoustic contact with the patient during scanning.

If proper gel use restores consistent readings, educate staff and stop troubleshooting.

Verify Scan Technique and Probe Placement

Incorrect probe angle or placement can produce poor detection or inaccurate volume readings.

Action:

Expected outcome: The scanner should guide the user toward an acceptable bladder image and produce repeatable results when used correctly.

If readings improve with corrected technique, the issue is likely user technique, not device failure.

Consider Patient-Related Factors

Some patients are harder to scan accurately.

Action:

Ask whether the issue occurred on a patient with:

Expected outcome: Patient-specific factors are identified as possible causes of poor detection.

If the device works normally on other patients or test conditions, document that no device fault was confirmed.

Compare Repeatability

Check whether the reported readings are repeatable.

Action:

Expected outcome: A properly functioning scanner should provide reasonably consistent results under similar scan conditions.

If readings vary wildly despite good technique, clean probe surface, proper gel, and appropriate patient conditions, continue troubleshooting.

Check Battery Level and Power Stability

Low battery or unstable power may contribute to poor device behavior, freezing, or incomplete scans.

Action:

Expected outcome: The scanner should operate normally with sufficient battery power.

If poor scanning only occurs at low battery levels, document the battery behavior and evaluate battery/runtime separately.

Restart the Scanner

Software or temporary scan processing issues may clear after a restart.

Action:

Expected outcome: The touchscreen and scan workflow should respond normally after restart.

If restart resolves the issue and readings are consistent, return the device to service after documenting the reset and successful test.

Check for Screen Prompts, Error Messages, or Software Warnings

The scanner may provide guidance that helps identify the cause.

Action:

Look for messages related to:

Expected outcome: Any displayed message should be documented exactly as shown.

If a persistent system, probe, calibration, or service message appears, remove the device from service and escalate.

Confirm the Correct Exam Workflow Is Being Used

Incorrect patient or scan workflow can create documentation confusion or perceived inaccurate results.

Action:

Expected outcome: The displayed scan result and saved result should match the intended patient/workflow.

If the scanner reads correctly but results are being assigned, saved, or uploaded incorrectly, troubleshoot the connectivity/documentation workflow instead.

Compare Against Another Scanner if Available

A comparison can help separate device failure from patient or technique issues.

Action:

Expected outcome: If another scanner consistently detects the bladder while the BladderScan i10 does not, the i10 may require service evaluation.

If both devices struggle, the issue may be patient condition, technique, or scan environment.

Review Recent History

Recent drops, cleaning events, software issues, or charging problems may explain the failure.

Action:

Ask staff whether the scanner was recently:

Check work order history for repeat complaints.

Expected outcome: Repeated or event-related failures are identified.

If the issue started after physical damage or fluid exposure, remove the device from service.

If the Problem Persists

If poor bladder detection or inaccurate volume readings continue after external checks, cleaning, gel use, technique, patient factors, battery status, restart, and workflow have been ruled out, the issue may involve the probe, ultrasound subsystem, software, calibration, or internal electronics.

The device should be:

Knowing when to stop is proper troubleshooting. Clinical Engineering should not continue relying on a bladder scanner that cannot provide repeatable and clinically reasonable scan results.

Clinical Use Tip

Do not troubleshoot questionable bladder volume readings while staff are actively making a patient care decision. Move the patient to another verified bladder scanner or alternate approved assessment method first. Troubleshooting should only continue when patient care is protected.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the Verathon BladderScan i10 was giving inconsistent bladder volume readings and frequently failed to detect the bladder."

Cause

What was observed during troubleshooting.

Example:
"Probe face was found with dried gel residue, and repeat testing showed improved scan consistency after cleaning and proper gel application."

Resolution

What action was taken.

Example:
"Cleaned probe surface, verified adequate gel use, restarted the unit, performed functional scan workflow check, and returned the BladderScan i10 to service."

Helpful Details to Include (If Known)

Final Thought

Poor bladder detection should be approached logically before assuming device failure. Clinical Engineering should first rule out safety concerns, gel contact, probe condition, patient factors, technique, battery status, and workflow problems. If the device still cannot produce repeatable, reasonable results, escalation is the correct action. Clear CCR documentation protects patient safety and supports future repair decisions.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide