Verathon BladderScan i10

Touchscreen, Boot, or Software Freeze Issue

On this page

Asset Type

Bladder Scanner

Manufacturer

Verathon

Model

BladderScan i10

What This Guide Helps With

Troubleshooting touchscreen lockups, startup freezes, delayed booting, or software response issues on the BladderScan i10.

Step-by-Step Troubleshooting

1. Ensure Patient Safety

Do not troubleshoot the bladder scanner while it is actively being used for a patient scan.

Action:

Expected outcome: Patient care continues without relying on equipment that may be unresponsive or unreliable.

Why it matters: A frozen or partially booted bladder scanner may display old data, fail to save results, or provide unreliable workflow information.

2. Verify the Reported Issue

Confirm exactly what the user experienced.

Action:

Expected outcome: The failure mode is confirmed before assuming a software or hardware issue.

If the device boots normally and the touchscreen responds, document that the issue was not duplicated and continue functional checks.

3. Check Battery Charge and Power Status

Low battery or unstable power can cause startup or software behavior that looks like a freeze.

Action:

Expected outcome: The device boots normally and the touchscreen responds.

If the issue resolves after charging, return the device to service after confirming normal operation and documenting the low battery condition.

4. Inspect the Charging Base and Power Supply

A poor charging connection may leave the battery too depleted for normal startup.

Action:

Expected outcome: The unit charges consistently and powers on normally.

If the scanner does not charge or repeatedly loses power, troubleshoot as a power or charging issue rather than a software issue.

5. Perform a Normal Restart

A temporary software lockup may clear with a controlled restart.

Action:

Expected outcome: The BladderScan i10 completes startup and responds normally.

If the issue resolves, perform a basic scan workflow check before returning the device to service.

6. Check for Touchscreen Contamination or Interference

Fluids, gel, gloves, cleaning residue, or screen protectors can interfere with touchscreen response.

Action:

Expected outcome: Touch input responds accurately after the screen is clean and dry.

If touchscreen response returns to normal, document contamination or residue as the likely cause.

7. Check for Physical Damage

Screen or housing damage may cause intermittent touchscreen failure or boot problems.

Action:

Expected outcome: No visible damage is found.

If damage, liquid exposure, heat, or odor is present, remove the device from service and send it for repair evaluation.

8. Verify the Unit Is Not Stuck During Startup

Determine whether the issue occurs before or after the operating screen loads.

Action:

Expected outcome: The point of failure is identified for repair documentation.

If the unit repeatedly freezes at the same point during startup, suspect a software, memory, battery, or internal hardware issue.

9. Check for Accessory or Docking-Related Problems

Some lockups may occur when the device is docked, charging, printing, exporting, or communicating with another system.

Action:

Expected outcome: The failure is narrowed to the scanner, charging base, or communication/documentation workflow.

If removing the device from the dock restores normal touchscreen use, inspect the dock and power setup further.

10. Confirm Date, Time, User, and Workflow Screens Respond

Software lockups may appear only during certain workflow steps.

Action:

Expected outcome: The device responds throughout the normal clinical workflow.

If freezing occurs only at one workflow point, document the exact screen and action that triggers the freeze.

11. Perform a Basic Functional Check

Do not return the device based only on a successful restart.

Action:

Expected outcome: The BladderScan i10 operates normally through the expected workflow.

If all checks pass, return the device to service.

12. Do Not Attempt Deep Internal Repair During Floor Troubleshooting

Avoid opening the unit or attempting board-level troubleshooting during normal clinical response.

Action:

Expected outcome: Troubleshooting remains safe, controlled, and appropriate for Clinical Engineering response.

If the Problem Persists

If the BladderScan i10 still freezes, will not complete startup, has an unresponsive touchscreen after cleaning and restart, or repeatedly locks up during normal use, common external causes have been ruled out.

The device should be:

Knowing when to stop is proper troubleshooting. A bladder scanner with unreliable boot or touchscreen behavior should not be returned to clinical use until normal operation is verified.

Clinical Use Tip

Do not troubleshoot touchscreen or startup failures during active patient care. Move the patient workflow to another available bladder scanner first. A scanner that freezes may delay care, display incomplete information, or prevent proper documentation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported the Verathon BladderScan i10 touchscreen froze during startup and would not allow a bladder scan to be started."

Cause

What was observed during troubleshooting.

Example:
"Verified the unit had a very low battery and would not fully boot until placed on a known-good charging base."

Resolution

What action was taken.

Example:
"Charged the unit, confirmed normal startup, verified touchscreen response, completed a functional scan workflow check, and returned the device to service."

Helpful Details to Include (If Known)

Final Thought

Touchscreen and boot issues should be approached logically. Start with patient safety, then rule out battery charge, charging setup, screen contamination, docking issues, and obvious damage. If the BladderScan i10 continues to freeze after external checks, remove it from service and escalate it for repair. Clear CCR documentation helps show what was tested, what was found, and why the device was or was not returned to service.

That is successful troubleshooting.