Verathon BladderScan i10

Wireless/cart connection or results upload failure

On this page

Asset Type

Bladder Scanner

Manufacturer

Verathon

Model

BladderScan i10

What This Guide Helps With

Troubleshooting BladderScan i10 wireless, cart, docking, or results upload issues caused by power, connection, network, or workflow problems.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Do not troubleshoot connectivity or upload issues while delaying urgent patient care.

Action:

Expected outcome: Patient care is not delayed because of a documentation or upload failure.

Why it matters: Connectivity issues usually affect documentation, but patient care decisions may still depend on the displayed scan result.

Verify the Reported Problem

Confirm whether the issue is with the wireless connection, cart connection, docking station, printer, network upload, or EMR/interface transfer.

Action:

Ask staff what failed:

Check whether the device displays an error message, connection icon, upload status, or queued result indicator.

Expected outcome: The failure type is clearly identified before parts or network issues are assumed.

Confirm the Scanner Has Adequate Battery Power

Low battery or unstable power can interrupt wireless communication and upload functions.

Action:

Expected outcome: The scanner has enough power to maintain communication and complete uploads.

If the issue resolves after charging, return the device to service and document low battery as the likely cause.

Check the Cart, Dock, and Charging Contacts

If the issue involves cart communication, docking, or upload after docking, inspect the external connection points.

Action:

Expected outcome: The scanner docks normally and any cart-based communication or charging indicator behaves as expected.

If reseating or cleaning the contacts restores connection, stop troubleshooting and document the correction.

Restart the Scanner and Cart Components

A temporary software or communication lockup may prevent upload or connection.

Action:

Expected outcome: Temporary communication or software lockup clears after restart.

If upload resumes after restart, document the reset and monitor for repeat failures.

Verify Wireless Network Availability

Wireless upload failures may be caused by local Wi-Fi coverage, network outage, or device roaming issues.

Action:

Expected outcome: The issue is narrowed to device-specific failure, location-based signal issue, or larger network problem.

If the scanner uploads successfully in another area, document possible local wireless coverage or network access issue.

Confirm the Device Is Using the Correct Network Profile

The scanner may fail to upload if it is not connected to the correct hospital wireless profile or configured upload destination.

Action:

Expected outcome: The device is confirmed to be on the intended clinical network or identified as needing IT/network support.

Check for Queued or Failed Results

The scanner may be collecting results but failing to transmit them.

Action:

Expected outcome: Upload failure is confirmed as isolated to specific results, all results, or the upload pathway.

If queued results upload successfully after reconnecting, return the device to service and document the recovery.

Confirm Patient/Operator Workflow Requirements

Some upload failures occur because required patient, operator, or encounter information is missing.

Action:

Expected outcome: Upload failure is not caused by incomplete workflow or missing required documentation fields.

If completing the required workflow allows upload, educate staff as appropriate and document the finding.

Test With a Known Good Workflow

Use a controlled test to separate scanner failure from workflow or network failure.

Action:

Expected outcome: The scanner either uploads successfully in a known-good setup or continues to fail despite external causes being ruled out.

If it uploads successfully, the original issue is likely location, workflow, or network related.

Escalate to IT or Interface Support When Appropriate

If the scanner connects to Wi-Fi but results do not appear in the destination system, the issue may be outside the scanner.

Action:

Expected outcome: The issue is routed correctly instead of replacing scanner hardware unnecessarily.

Remove From Service if Connection Failure Persists

If the scanner repeatedly fails to connect, upload, or communicate after basic checks, remove it from clinical use.

Action:

Expected outcome: A device with unreliable documentation or connectivity is removed until the fault is resolved.

If the Problem Persists

If battery, dock seating, contacts, restart, wireless coverage, workflow, and network availability have been checked, the issue may involve internal wireless hardware, software configuration, cart communication hardware, or interface configuration.

The BladderScan i10 should be:

Knowing when to stop is proper troubleshooting. Continued trial-and-error in a clinical area can delay care, create documentation gaps, and make the failure harder to track.

Clinical Use Tip

Do not allow upload failure to delay patient care. If a scan result is clinically needed, staff should view and document the result according to facility policy, then Clinical Engineering can troubleshoot the upload issue when safe.

Do not troubleshoot on an active patient. Move the patient to a backup device first if continued assessment or therapy continuity is required.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the Verathon BladderScan i10 was scanning normally but results were not uploading from the cart/wireless connection."

Cause

What was observed during troubleshooting.

Example:
"Scanner had pending results and poor wireless connection in the reported room; upload completed when tested in an area with known good wireless coverage."

Resolution

What action was taken.

Example:
"Verified scanner power, reseated scanner on cart, restarted unit, confirmed wireless connectivity, completed test upload, and returned device to service."

Helpful Details to Include (If Known)

Final Thought

Wireless and upload problems should be approached logically. Protect patient care first, confirm the scanner still produces usable results, then work outward through power, docking, wireless coverage, workflow, and interface status. If external causes are ruled out, remove the device from service and document the failure clearly so repair, IT, or vendor support can act on accurate information.

That is successful troubleshooting.