Verathon Prime Plus

Calibration, Self-Test, or Startup Check Failure

On this page

Asset Type

Bladder Scanner

Manufacturer

Verathon

Model

Prime Plus

What This Guide Helps With

Helps isolate power, probe, accessory, connection, environmental, startup, and authorized calibration-related causes when required checks do not complete successfully.

Step-by-Step Troubleshooting

1. Protect Patient Care and Remove an Unverified Unit From Use

Do not use the Prime Plus for patient assessment when a required startup, self-test, or calibration-related check has failed and measurement reliability cannot be confirmed.

Provide another verified bladder scanner.

Record the exact displayed message or observed behavior.

Expected outcome: The affected unit is removed from clinical dependency and the exact reported failure is documented.

2. Restart the Scanner Under Normal Conditions

Power the device down using the normal procedure. Allow the shutdown to complete, then restart it without repeatedly cycling power.

Observe the complete startup sequence.

Expected outcome: The scanner completes startup normally. If the failure was transient and a representative functional test is also successful, troubleshooting can stop after verification.

3. Verify Power Condition

Confirm the scanner has adequate power for testing. If battery status is questionable, connect the approved external power or charging arrangement.

Inspect the power supply and accessible connections for damage or looseness.

Expected outcome: Stable power is available and the startup check is not being interrupted by low or intermittent power. If stable power resolves the failure, verify normal operation before return to service.

4. Inspect the Probe and Accessories

Inspect the probe, cable, accessible connectors, dock, and other required external accessories for:

Do not use a damaged probe or connector.

Expected outcome: Required external components are intact and correctly connected. If reseating an accessory allows the self-test or startup check to pass, proceed to final verification.

5. Remove Unnecessary External Connections

Disconnect nonessential external accessories or communication connections when permitted, then restart the scanner in its basic approved configuration.

This can help identify whether an external accessory is interfering with startup.

Expected outcome: The scanner completes startup in its basic configuration. If an optional external accessory causes the failure, remove or replace that accessory and verify the scanner again.

6. Check the Environment

Confirm the device is not being tested immediately after:

Allow the device to reach an appropriate normal-use environment when needed before retesting.

Expected outcome: Environmental conditions are suitable for normal operation. If the startup check then completes successfully, proceed to final verification.

7. Use Only Authorized Calibration or Verification Functions

If the reported condition specifically involves calibration, follow the applicable Verathon service documentation and use only authorized test equipment and procedures available to qualified Clinical Engineering personnel.

Do not guess at calibration values, enter undocumented service menus, or modify calibration constants simply to clear a failure.

Expected outcome: The authorized calibration or verification process either completes successfully or confirms that service-level evaluation is required.

8. Check for Repeatability

Restart and test the device again under the same verified conditions.

A single successful startup is not sufficient if the original failure was intermittent.

Expected outcome: Startup and required checks pass consistently. If failures recur, remove the unit from service and escalate.

9. Perform Final Functional Verification

Before return to service, confirm:

Use any additional return-to-service testing required by manufacturer documentation and facility policy.

Expected outcome: The scanner demonstrates consistent normal operation and required verification has passed. The device may be returned to service.

If the Problem Persists

External power, probe, accessories, connections, environment, and permitted startup or calibration checks have been ruled out. The remaining issue may involve internal electronics, software, stored calibration data, probe performance, or another service-level condition.

The device should be:

Do not return the scanner to clinical use until required startup, self-test, calibration, and functional verification requirements are satisfied.

Recognizing that a failed verification condition requires escalation rather than trial-and-error adjustment is proper troubleshooting.

Clinical Use Tip

A bladder scanner that cannot complete required verification should not be trusted for patient measurement until its performance has been confirmed.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported the Prime Plus stopped during startup and would not complete its normal startup check."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the probe connection was partially seated, preventing stable probe recognition during startup."

Resolution

What action was taken.

Example:
"Clinical Engineering reseated and inspected the probe connection, verified repeated successful startups and representative scans, and returned the unit to service."

Helpful Details to Include (If Known)

Final Thought

A failed startup, self-test, or calibration check should be treated as a verification problem until proven otherwise. Protect the patient, eliminate external causes methodically, use only authorized calibration procedures, escalate unresolved failures, and document the final verified condition clearly.

That is successful troubleshooting.