Vyaire AirSep / Mobilaire Series

Filter Blockage or Airflow Restriction

On this page

Asset Type

Oxygen Concentrator

Manufacturer

Vyaire

Model

AirSep / Mobilaire Series

What This Guide Helps With

Reduced airflow, low output, overheating, or alarms caused by clogged filters or restricted air intake/exhaust.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Patient continues receiving prescribed oxygen
Why it matters: Airflow restriction can reduce oxygen delivery and compromise therapy

Verify the Reported Issue

Expected: Symptoms are consistent and reproducible
Why it matters: Confirms airflow restriction vs unrelated issue

Check External Air Intake (Most Common Cause)

Expected: Air can freely enter the concentrator
Why it matters: Restricted intake reduces compressor efficiency and oxygen production

Inspect and Clean External Filters

Expected: Filter is clean and allows airflow
Why it matters: Clogged filters are the most common cause of airflow restriction

Check Internal/Secondary Filters (If Accessible Without Disassembly)

Expected: Filters are clean or replaced
Why it matters: Secondary filters can silently restrict airflow over time

Inspect Exhaust Venting

Expected: Warm air is expelled freely
Why it matters: Blocked exhaust traps heat and reduces system performance

Check for Environmental Factors

Expected: Unit is in a clean, well-ventilated location
Why it matters: Poor environment accelerates filter blockage and airflow issues

Re-test Unit Performance

Expected: Airflow returns to normal and alarms clear
Why it matters: Confirms issue was caused by airflow restriction

If the Problem Persists

If airflow issues continue after cleaning filters and ensuring proper ventilation, internal components such as the compressor or internal airflow pathways may be restricted or failing. Remove the device from service, label it Out of Service, and send for bench evaluation or repair. Knowing when to stop prevents unnecessary risk and improper repair attempts.

Clinical Use Tip

Never troubleshoot airflow or filter issues while the device is actively supplying oxygen to a patient. Always transition the patient first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Oxygen concentrator alarming for low flow and reduced output."

Cause

What was observed during troubleshooting.

Example:
"Intake filter heavily clogged with dust restricting airflow."

Resolution

What action was taken.

Example:
"Cleaned filter, verified proper airflow and normal operation."

Helpful Details to Include (If Known)

Final Thought

Airflow issues are one of the most common and preventable causes of oxygen concentrator problems. Starting with simple external checks like filters and ventilation ensures safe, efficient troubleshooting. Proper escalation protects both the patient and the equipment, while clear documentation supports long-term reliability and maintenance tracking.

That is successful troubleshooting.