Vyaire AirSep / Mobilaire Series

Low Oxygen Purity Alarms

On this page

Asset Type

Oxygen Concentrator

Manufacturer

Vyaire

Model

AirSep / Mobilaire Series

What This Guide Helps With

Low oxygen purity alarms caused by airflow restriction, filter issues, environmental factors, or compressor performance problems.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm

Allow for Warm-Up Time (Common and Overlooked)

Check Air Intake Filters (Most Common Cause)

Inspect Air Intake Area and Placement

Check Oxygen Output Path

Verify Flow Rate Settings

Assess Environmental Conditions

Listen for Compressor Operation

Reset and Retest the Unit

If the Problem Persists

If all external causes have been ruled out and the low oxygen purity alarm remains:

Likely internal causes include sieve bed degradation, oxygen sensor failure, or compressor inefficiency. These require internal service.

Clinical Use Tip

Never troubleshoot oxygen purity alarms while the device is connected to a patient. Always transition to a verified oxygen source before testing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit alarming low oxygen purity during patient use."

Cause

What was observed during troubleshooting.

Example:
"Air intake filter heavily clogged, restricting airflow."

Resolution

What action was taken.

Example:
"Cleaned/replaced filter, verified oxygen output, and confirmed alarm cleared."

Helpful Details to Include (If Known)

Final Thought

Low oxygen purity alarms are most often caused by airflow or environmental issues rather than immediate internal failure. A structured, external-first approach ensures patient safety while preventing unnecessary repairs. Recognizing when to escalate is just as important as resolving the issue.

That is successful troubleshooting.