Welch Allyn Connex 6000 Series

Error: #000000015 Software Watchdog Timer Reset

On this page

Asset Type

Vital Signs Monitor

Manufacturer

Welch Allyn

Model

Connex 6000 Series

What This Guide Helps With

Error #000000015 indicating watchdog reset causing unexpected reboot, freezing, or system instability during operation.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Error and Behavior

3. Perform a Full Power Cycle

4. Check Power Source Stability

5. Inspect Battery Condition and Connection

6. Disconnect External Accessories

7. Check Network Connectivity (If Connected)

8. Observe for Overheating or Environmental Factors

If the Problem Persists

If the watchdog reset continues after all external checks:

Remove the device from service, label as Out of Service, and send for repair or bench evaluation.

Common likely causes include corrupted firmware/software, main board instability, or internal communication failures.

Clinical Use Tip

Never rely on a monitor that is rebooting or freezing during patient care. Always transfer monitoring before troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Device displaying Error #000000015 with intermittent resets during operation."

Cause

What was observed during troubleshooting.

Example:
"External causes ruled out; watchdog timer resets indicate likely internal system instability or main board/software fault."

Resolution

What action was taken.

Example:
"Performed power cycle, verified power source, battery, and accessories. Issue persisted. Device removed from service and sent for repair."

Helpful Details to Include (If Known)

Final Thought

Watchdog timer errors are critical indicators of system instability. Focus on ruling out power, accessory, and environmental causes, then escalate appropriately to protect patient safety and equipment reliability.

That is successful troubleshooting.