Werfen ACL TOP 750 LAS

Calibration Fails or Will Not Complete

On this page

Asset Type

Coagulation Analyzer

Manufacturer

Werfen

Model

ACL TOP 750 LAS

What This Guide Helps With

Troubleshoots calibration failures caused by calibration material, reagent status, sample handling, setup, fluidics, environmental conditions, or analyzer readiness.

Step-by-Step Troubleshooting

1. Protect Patient Testing and Confirm the Calibration Failure

Do not release patient results from an assay requiring a valid calibration when calibration is failed, incomplete, or otherwise unacceptable.

Identify:

Expected outcome: The affected calibration and scope of the problem are clearly identified.

2. Verify Calibration Material

Confirm the correct calibration material is being used for the assay and that it is suitable for use according to laboratory procedures.

Inspect for:

Do not use material that appears compromised.

Expected outcome: Appropriate calibration material is available and prepared according to validated laboratory procedures.

3. Verify Reagent Status

Confirm the associated reagent:

If reagent recognition or inventory is abnormal, correct that problem before repeating calibration.

Expected outcome: Required reagent status is acceptable before calibration is attempted again.

4. Inspect the Calibration Sample and Loading Setup

Check the calibration container for:

Do not repeatedly rerun obviously compromised material.

Expected outcome: Calibration material is properly loaded and available for aspiration.

5. Check for Broader Analyzer Problems

Determine whether the analyzer also has:

A calibration failure may be secondary to a broader analyzer condition.

Expected outcome: No unresolved analyzer fault is present that would invalidate calibration.

6. Review Calibration Setup

Verify the assay and calibration selection in the normal user interface.

Confirm that staff have not selected the wrong assay, material, lot, or calibration sequence.

Do not alter validated assay parameters or restricted configuration values as a troubleshooting shortcut.

Expected outcome: The intended assay and calibration setup are selected correctly.

7. Repeat Calibration Using Verified Materials

After correcting any external issue, repeat the calibration according to the laboratory's approved workflow.

Observe whether:

Expected outcome: Calibration completes successfully and is accepted. If it does, continue with required QC and troubleshooting can stop after verification.

8. Confirm Quality Control Before Returning the Assay to Use

Following a successful calibration, run required QC according to laboratory procedure.

Do not assume a completed calibration alone is sufficient for patient testing.

Expected outcome: Required QC is acceptable and the assay is released according to laboratory policy. Troubleshooting can stop.

If the Problem Persists

If calibration material, reagent status, loading, assay selection, and obvious analyzer conditions have been verified but calibration still fails, common external causes have been ruled out.

The remaining cause may involve pipetting accuracy, optical or mechanical measurement systems, temperature regulation, fluidics, analyzer calibration functions, assay configuration, or another service-level condition.

The affected assay or analyzer should be:

Do not adjust assay parameters or internal calibration mechanisms without authorized procedures.

Return the assay to service only after successful calibration and required QC.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

A completed calibration is not the final release step; verify required QC before resuming patient testing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported that calibration for an ACL TOP 750 LAS assay repeatedly failed to complete."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found visible bubbles in the calibration material that interfered with consistent aspiration."

Resolution

What action was taken.

Example:
"Laboratory staff prepared fresh calibration material, Clinical Engineering verified normal aspiration and successful calibration, and required QC passed before patient testing resumed."

Helpful Details to Include (If Known)

Final Thought

Protect patient results by verifying materials, reagent status, sample handling, and analyzer readiness before assuming an internal calibration failure, then escalate when valid calibration cannot be established.

That is successful troubleshooting.