Werfen ACL TOP 750 LAS

Clot, Bubble, or Sample-Volume Error Repeats

On this page

Asset Type

Coagulation Analyzer

Manufacturer

Werfen

Model

ACL TOP 750 LAS

What This Guide Helps With

Troubleshoots repeated clot, bubble, or volume errors caused by specimen quality, collection, handling, positioning, aspiration conditions, or external loading problems.

Step-by-Step Troubleshooting

1. Protect Patient Testing and Confirm the Pattern

Do not report results from specimens repeatedly flagged for clot, bubble, or volume problems until specimen integrity and analyzer performance are verified.

Determine whether the problem affects:

Record the exact analyzer message.

Expected outcome: The problem is identified as specimen-specific, workflow-related, or analyzer-wide.

2. Inspect the Specimen

Visually inspect according to laboratory procedure for:

Do not manipulate a compromised patient specimen solely to make the analyzer accept it.

Expected outcome: Specimen suitability is established. If the specimen is clearly unsuitable, obtain or process an appropriate replacement according to laboratory policy.

3. Verify Collection and Handling Information

When repeated specimen-quality errors occur, confirm with laboratory staff whether:

Clinical Engineering should not redefine laboratory specimen-acceptance criteria.

Expected outcome: No obvious pre-analytical handling problem remains unexplained.

4. Verify Sample Position and Container

Confirm the specimen is:

Expected outcome: The analyzer can access the sample correctly.

5. Compare with a Known-Good Specimen or QC Material

Use appropriate known-good material according to laboratory policy.

If the known-good material processes normally while the affected sample does not, the problem is likely specimen-specific.

If multiple verified materials fail similarly, investigate the analyzer.

Expected outcome: The issue is isolated to the specimen or the analyzer.

6. Check for Related Aspiration or Fluidics Errors

Review whether the analyzer is also showing:

These may indicate that a sample-quality message is secondary to a broader liquid-handling problem.

Expected outcome: No broader analyzer condition is present.

7. Inspect Accessible Loading and Aspiration Areas

Check externally accessible areas for:

Clean only using approved procedures.

Expected outcome: No external obstruction contributes to repeated sample errors.

8. Perform Final Verification

After an external or specimen-related cause is corrected, process suitable verification material and required QC.

Expected outcome: Sample handling occurs without repeat clot, bubble, or volume errors and required QC is acceptable. Troubleshooting can stop.

If the Problem Persists

If multiple appropriate specimens or QC materials produce repeated clot, bubble, or volume errors despite correct positioning and external conditions, common specimen-related causes have been ruled out.

The remaining cause may involve liquid-level sensing, aspiration pressure or vacuum, probe operation, fluidics, sample detection, or another service-level subsystem.

The analyzer should be:

Do not bypass sample-quality detection or force acceptance of questionable specimens.

Return the analyzer to service only after reliable specimen handling and required QC are demonstrated.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Repeated clot or volume errors across multiple patient samples may indicate a broader analyzer problem and should not automatically be blamed on collection technique.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported repeated low-volume errors for one patient specimen on the ACL TOP 750 LAS."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the affected specimen did not contain enough accessible sample volume for reliable aspiration."

Resolution

What action was taken.

Example:
"Laboratory staff provided an acceptable replacement specimen, Clinical Engineering verified normal processing without repeat error, and QC remained acceptable."

Helpful Details to Include (If Known)

Final Thought

Separate specimen-quality problems from analyzer problems using careful inspection and known-good comparisons, and escalate when repeated errors occur with verified material.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide