Werfen ACL TOP 750 LAS

Quality-Control Results Are Out of Range

On this page

Asset Type

Coagulation Analyzer

Manufacturer

Werfen

Model

ACL TOP 750 LAS

What This Guide Helps With

Troubleshoots out-of-range QC caused by control material, reagent condition, calibration status, handling, loading, aspiration, or broader analyzer performance problems.

Step-by-Step Troubleshooting

1. Protect Patient Testing and Confirm the QC Failure

Do not release affected patient results while required QC is unacceptable.

Identify:

Follow laboratory policy for evaluating patient results generated since the last acceptable QC.

Expected outcome: The scope of the QC problem is known and affected patient testing is contained.

2. Verify QC Material

Confirm the intended QC material and level were used.

Inspect for:

Do not continue rerunning suspect QC material.

Expected outcome: Appropriate QC material is available and properly prepared.

3. Verify Reagent Condition and Status

Confirm the reagent associated with the failed QC:

Expected outcome: Reagent status is acceptable.

4. Verify Calibration Status

Check whether the affected assay has an acceptable calibration status.

Consider whether anything recently changed, such as:

Do not recalibrate automatically without first understanding the sequence of events and following laboratory policy.

Expected outcome: Calibration status and recent changes are understood.

5. Inspect QC Loading and Aspiration Conditions

Verify:

If the analyzer shows aspiration, probe, or fluidics errors, address those conditions first.

Expected outcome: QC material can be aspirated normally.

6. Compare the Pattern Across Assays and QC Levels

Determine whether:

A broader pattern may indicate an analyzer-wide condition rather than a single reagent or control problem.

Expected outcome: The failure pattern points toward material-specific, assay-specific, or analyzer-wide troubleshooting.

7. Use Fresh Verified QC Material When Appropriate

Following laboratory policy, test fresh QC material prepared correctly.

Avoid repeated testing solely to obtain an acceptable result.

Expected outcome: Fresh verified QC produces acceptable results. If so, document the material-related cause and troubleshooting can stop after the laboratory completes required release steps.

8. Perform Final Verification

If QC returns to acceptable range, verify:

Expected outcome: QC performance is stable and the affected assay can return to clinical use.

If the Problem Persists

If QC material, reagent condition, calibration status, loading, aspiration, and handling have been verified but QC remains unacceptable, common external causes have been ruled out.

The remaining cause may involve assay performance, fluid delivery, pipetting accuracy, temperature control, optical or mechanical measurement systems, configuration, or another service-level issue.

The affected assay or analyzer should be:

Do not repeatedly rerun QC to achieve an acceptable value without identifying the cause.

Return the assay to service only after required QC is acceptable and any necessary laboratory review is complete.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Out-of-range QC is a patient-result integrity issue; contain affected testing before attempting to restore analyzer availability.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported that one QC level for an ACL TOP 750 LAS assay was repeatedly out of range."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering and laboratory staff found the QC material had visible bubbles and had been inadequately prepared before loading."

Resolution

What action was taken.

Example:
"Fresh QC material was prepared according to laboratory procedure, acceptable QC results were obtained, and the assay was released for patient testing."

Helpful Details to Include (If Known)

Final Thought

Treat QC failure as a test-quality problem first, verify controls, reagents, calibration, and aspiration logically, and escalate when acceptable performance cannot be demonstrated.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide