Werfen ACL TOP 750 LAS

Wash, Waste, or Fluidics System Fault Occurs

On this page

Asset Type

Coagulation Analyzer

Manufacturer

Werfen

Model

ACL TOP 750 LAS

What This Guide Helps With

Troubleshoots wash, waste, and fluidics faults caused by empty supplies, full waste, tubing connections, external leaks, consumables, or accessible obstructions.

Step-by-Step Troubleshooting

1. Protect Patient Testing and Confirm the Fault

Stop patient testing if the analyzer cannot reliably wash, aspirate, dispense, or manage waste.

Determine whether the message involves:

Record the exact message.

Expected outcome: The fluidics fault is clearly identified and patient testing is moved to another verified analyzer if necessary.

2. Inspect for Leaks or Spills

Check around and beneath the analyzer and accessible fluid containers for:

Use appropriate PPE and follow laboratory spill procedures.

Expected outcome: No uncontrolled leak is present. If leakage cannot be safely contained or its source is internal, remove the analyzer from service immediately.

3. Verify Wash or Supply Fluids

Check externally accessible fluid supplies required for analyzer operation.

Confirm:

Expected outcome: Required supply fluids are available and connected correctly. If correcting the supply restores normal operation, proceed to verification.

4. Check Waste Capacity and Connections

Inspect the waste container or waste connection according to the installed configuration.

Confirm:

Handle waste using appropriate laboratory precautions.

Expected outcome: Waste can be safely collected and the analyzer no longer detects an external waste condition.

5. Inspect External Tubing and Connections

Without opening protected service areas, inspect accessible tubing for:

Do not disconnect internal tubing or open fluidic assemblies without authorized service procedures.

Expected outcome: Accessible fluid connections are intact.

6. Check Related Consumables and Covers

Verify any associated wash, rinse, waste, or fluid consumables are installed correctly.

Confirm covers or compartments associated with external consumables are fully closed and nothing is interfering with normal operation.

Expected outcome: All externally serviceable consumables are correctly installed.

7. Resume the Normal Fluidics Sequence

After correcting an external condition, use normal analyzer controls to allow the system to perform its standard recovery or readiness process.

Do not repeatedly restart a leaking or mechanically abnormal system.

Expected outcome: The fluidics fault clears and the analyzer returns to normal readiness.

8. Perform Functional Verification

Before returning to patient use, verify:

Expected outcome: Fluid handling is stable and required QC is acceptable. Troubleshooting can stop.

If the Problem Persists

If fluids, waste capacity, consumables, accessible tubing, and external connections are correct but the fluidics fault continues, common external causes have been ruled out.

The remaining cause may involve pumps, valves, pressure or vacuum systems, internal tubing, sensors, wash mechanisms, drainage components, or other service-level fluidics assemblies.

The analyzer should be:

Do not operate an analyzer with unresolved leakage, waste-control problems, or unreliable fluid movement.

Return the analyzer to service only after leak-free operation, stable fluidics, and required QC have been demonstrated.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Treat any uncontrolled analyzer-fluid or waste leak as both an equipment problem and a potential laboratory exposure hazard.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported a wash-system fault and the ACL TOP 750 LAS would not return to ready status."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the external wash-fluid connection was not fully seated."

Resolution

What action was taken.

Example:
"Clinical Engineering secured the connection, verified the fluidics fault cleared without leakage, and confirmed acceptable QC before the analyzer returned to service."

Helpful Details to Include (If Known)

Final Thought

Start with fluid supply, waste, connections, and visible leaks before assuming an internal fluidics failure, and remove the analyzer from service whenever safe and reliable fluid handling cannot be confirmed.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide