ZOLL AED Plus

CPR Feedback or Compression Detection Failure

On this page

Asset Type

Defibrillator

Manufacturer

ZOLL

Model

AED Plus

What This Guide Helps With

Troubleshooting missing, intermittent, or inaccurate CPR feedback caused by incompatible pads, CPR sensor position, pad connections, surface movement, or accessory damage.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not interrupt clinically indicated CPR to investigate the feedback system during an active resuscitation.

If CPR feedback fails during patient care:

Expected outcome: CPR and defibrillation continue without depending on unreliable compression feedback.

Continue Clinical Engineering troubleshooting only after the AED Plus has been removed from active patient use.

2. Confirm the Reported Failure

Determine whether the complaint involves:

Record the exact prompts, indicator behavior, electrode type, and point in the rescue sequence when the failure occurred.

Expected outcome: The complaint is clearly distinguished from an audio, electrode-recognition, or general device failure.

3. Verify That CPR-D-padz Are Installed

Inspect the connected electrodes and package labeling.

The AED Plus CPR monitoring function requires compatible CPR-D-padz with the compression sensor. Standard electrodes or pediatric Pedi-padz II do not provide adult CPR depth feedback. ZOLL states that the CPR monitoring function is intended for adult patients and is not intended for patients under eight years old.

Confirm that:

Expected outcome: A sealed, compatible CPR-D-padz set is available.

If the wrong electrode type caused the complaint, replace it with the correct accessory, verify readiness, document the correction, and stop.

4. Inspect the CPR-D-padz Package and Sensor

Without opening a service-ready pad package unnecessarily, inspect it for:

A damaged or expired pad set should not be used for patient care. ZOLL instructs users to apply freshly opened, undamaged electrodes that are within their expiration date.

Expected outcome: The electrode package and CPR sensor appear intact.

If damaged, opened, used, or expired, replace the CPR-D-padz, verify the green readiness indicator, document the accessory failure, and stop.

5. Check the Electrode Connector

Verify that the electrode cable connector is:

Inspect the device-side connector externally for bent contacts, damaged plastic, looseness, or foreign material. Do not insert tools into the connector.

ZOLL recommends keeping the electrode cable connected to the AED Plus and lists cable attachment as a check when the device displays a readiness fault.

Expected outcome: The electrode connection is secure and mechanically stable.

If reseating the connector resolves the problem, complete the functional check, return the AED to service only when readiness is normal, document the repair, and stop.

6. Confirm Correct Sensor Placement During Controlled Testing

Use an approved AED simulator, CPR manikin, and facility-authorized test procedure. Do not test compression feedback on a person.

Confirm that the CPR sensor would be positioned:

The CPR-D-padz sensor reports chest-compression motion to the AED Plus. Incorrect sensor placement or movement outside the compression point can prevent accurate feedback.

Expected outcome: The sensor remains centered and detects controlled compressions.

If repositioning restores consistent feedback, document improper sensor placement as the cause and stop.

7. Check the Test Surface and Manikin

Place the CPR manikin on a firm, stable surface.

Avoid testing on:

Confirm that the manikin allows measurable chest displacement and returns fully after each compression.

Expected outcome: Compression movement occurs through the sensor rather than being absorbed by the support surface.

If changing the surface or manikin resolves the issue, repeat the test for consistency, document the test-condition problem, and stop.

8. Observe the CPR Sequence and Prompts

Using the simulator and test manikin, advance the AED through its normal rescue sequence until it instructs the rescuer to begin CPR.

Verify:

The AED Plus CPR monitoring function provides a metronome and voice or visual feedback intended to encourage appropriate adult compression rate and depth.

Expected outcome: The metronome and corrective prompts operate during the CPR phase.

If the device has not entered a CPR interval, absent feedback may be normal rather than a failure.

9. Compare With a Known-Good CPR-D-padz Test Accessory

When permitted by the facility and ZOLL service procedures, repeat the controlled test using a known-good, compatible CPR-D-padz test accessory or approved training interface.

Do not open a patient-use electrode package merely for troubleshooting unless it will be removed from inventory afterward.

Interpret the results:

Expected outcome: The failure is isolated to either the electrode assembly or the AED.

If replacing the accessory resolves the issue, complete the readiness check, document the result, and stop.

10. Check Battery and Readiness Status

Inspect the AED Plus status indicator.

Do not return the AED to service if:

ZOLL instructs that a unit displaying a red “X” or emitting a warning beep should not be placed in service.

If battery replacement is indicated, replace all ten batteries as a complete matched set using the manufacturer-specified type and follow the approved battery-reset procedure.

Expected outcome: The AED completes its self-test and shows the green check readiness symbol.

If normal readiness and CPR feedback return, document the battery or readiness condition and stop.

11. Perform a Final Functional Verification

After correcting any accessory, connection, or battery issue:

ZOLL’s checkout procedure requires successful startup, correct pad recognition, and a green readiness indicator before the AED is placed into service.

Expected outcome: CPR feedback functions consistently and the AED returns to normal ready status.

If the Problem Persists

If compatible CPR-D-padz, connector seating, sensor placement, test conditions, battery condition, and readiness checks are satisfactory but compression feedback remains absent or inaccurate, common external causes have been ruled out.

The problem may involve the electrode connector, compression-detection circuitry, audio system, or internal processing.

The device should be:

Do not open or internally disassemble the AED Plus. ZOLL warns that internal disassembly presents a shock hazard and directs servicing to qualified personnel.

Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Do not troubleshoot CPR feedback on an active patient. Continue CPR using trained clinical judgment, obtain another verified defibrillator when needed, and evaluate the affected AED only after patient care has been transferred safely.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the ZOLL AED Plus did not provide “Push Harder” or “Good Compressions” feedback during CPR."

Cause

What was observed during troubleshooting.

Example:
"Inspection found that standard adult defibrillation pads without the CPR compression sensor were connected instead of CPR-D-padz."

Resolution

What action was taken.

Example:
"Installed compatible, in-date CPR-D-padz, completed a simulated rescue and compression-feedback test, and confirmed a green readiness indicator before returning the AED to service."

Helpful Details to Include (If Known)

Final Thought

Reliable CPR feedback depends on the correct electrode assembly, proper sensor placement, stable test conditions, secure connections, and normal AED readiness. Patient safety comes first, followed by logical external checks, appropriate escalation, and complete CCR documentation.

That is successful troubleshooting.