ZOLL AED Plus

Unexpected Shutdown or Intermittent Power Loss

On this page

Asset Type

Defibrillator

Manufacturer

ZOLL

Model

AED Plus

What This Guide Helps With

Troubleshooting unexpected shutdowns, restarts, or intermittent operation caused by depleted batteries, improper installation, battery contacts, environmental exposure, or internal power faults.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended troubleshooting while the AED Plus is required for an active cardiac-arrest response.

If the device shuts down or repeatedly loses power during patient care:

Expected outcome: Patient care continues without relying on an AED with uncertain power stability.

Continue Clinical Engineering troubleshooting only after the affected AED has been removed from active use.

2. Confirm the Reported Power Failure

In a controlled, nonclinical area, document whether the AED:

Do not repeatedly cycle a device that produces unusual heat, odor, smoke, or battery-compartment damage.

Expected outcome: The exact shutdown pattern and associated indicators are identified.

3. Inspect the Exterior and Battery Compartment

Turn the AED off and inspect for:

Do not disassemble the AED or attempt internal repair. ZOLL directs internal servicing to qualified repair personnel because of electrical hazards.

Expected outcome: No external condition is found that could interrupt battery contact or make further testing unsafe.

If significant damage, corrosion, heat, odor, or liquid intrusion is present, stop and remove the AED from service.

4. Verify the Battery Type and Installation

The AED Plus uses 10 Duracell Coppertop CR123 batteries made in the USA. Confirm that:

ZOLL specifies replacing all 10 batteries together rather than replacing individual cells.

Expected outcome: A complete, correctly installed, matched battery set is present.

If installation was incorrect, correct it and proceed to the checkout test. If the issue resolves and the AED passes all readiness checks, stop.

5. Replace the Complete Battery Set When Battery Condition Is Uncertain

Replace all 10 batteries when:

Remove all existing batteries before installing the new set. Insert 10 new specified batteries, following the compartment polarity markings. Do not mix the removed batteries with the new set.

Press the battery-compartment reset button only after all 10 new batteries have been installed, as directed by ZOLL.

Expected outcome: The AED has a complete, known-good battery source and initializes normally.

6. Confirm the Electrode Cable Connection

Inspect the electrode connector and confirm that the compatible electrode cable is fully seated and undamaged.

ZOLL instructs that the electrode cable be connected after battery installation and remain connected during normal readiness storage. A disconnected or improperly attached cable may produce a failed-readiness indication that could be mistaken for a power problem.

Expected outcome: The electrode cable is securely connected, and no connector damage is visible.

If reseating the connector corrects the warning and the AED passes checkout, stop.

7. Perform the Power-On Checkout

Turn the AED on and observe the entire startup sequence.

Confirm that:

Do not use a patient or simulate therapy delivery during this check.

Expected outcome: The AED completes startup without shutting down, restarting, or displaying a fault.

If the unit shuts down again with a complete new battery set, remove it from service.

8. Perform the Manual Self-Test

With the AED safely disconnected from patient use, press and hold the ON/OFF button for more than five seconds to initiate the manual test.

ZOLL recommends this test when a self-test has failed. If the unit fails again, it should be removed from service.

Observe for:

Expected outcome: The AED completes the manual self-test without interruption.

If it fails, shuts down, or restarts, stop troubleshooting and remove it from service.

9. Verify the Final Readiness Indicator

Turn the AED off and wait approximately two minutes.

Confirm:

ZOLL’s checkout procedure requires the green check to appear and no beeping to occur before the AED is placed into service. A red “X” or off-state beeping requires additional action and, if unresolved, removal from service.

Expected outcome: The AED maintains a normal green readiness indication.

If the green check remains stable after successful testing, return the unit to service according to facility policy.

If the Problem Persists

If the AED Plus continues to shut down, restart, lose power when moved, fail its manual self-test, display a red “X,” or beep after installation of a complete approved battery set, common external causes have been ruled out.

The failure may involve internal battery contacts, power-management circuitry, wiring, controls, or another internal component.

The device should be:

Do not open the housing, bypass battery contacts, or attempt board-level repair. Knowing when to stop testing an intermittently powered defibrillator is proper troubleshooting.

Clinical Use Tip

Never troubleshoot intermittent AED power while the device is supporting an active rescue. Continue CPR, move immediately to another verified defibrillator, and evaluate the affected AED only after emergency coverage has been restored.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the ZOLL AED Plus powered on normally but shut down intermittently during the startup voice prompts."

Cause

What was observed during troubleshooting.

Example:
"Inspection found a mixed and partially depleted set of CR123 batteries; the AED failed to maintain power until all 10 batteries were replaced together."

Resolution

What action was taken.

Example:
"Installed 10 new approved Duracell Coppertop CR123 batteries, completed the manual self-test and startup checkout, verified the green readiness check with no beeping, and returned the AED to service."

Helpful Details to Include (If Known)

Final Thought

Intermittent power loss in an AED must be treated as a serious readiness failure. Start with patient protection, verify the complete battery system and external connections, perform the approved checkout, and escalate immediately when stable operation cannot be demonstrated. Accurate CCR documentation preserves the observed failure, testing performed, and final disposition.

That is successful troubleshooting.