ZOLL Propaq MD

EtCO2 Sampling Failure or No Capnogram

On this page

Asset Type

Defibrillator

Manufacturer

ZOLL

Model

Propaq MD

What This Guide Helps With

No EtCO2 value or capnogram appears because of sampling-line blockage, loose connections, moisture, consumables, positioning, or sampling-system problems.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Ventilation Monitoring

If capnography is clinically required, provide another verified method of EtCO2 or ventilation monitoring before troubleshooting the affected channel.

Do not troubleshoot an unreliable capnography system while a patient depends solely on it.

Expected outcome: Required ventilation monitoring continues throughout evaluation.

2. Confirm the Exact Capnography Complaint

Determine whether the reported issue is:

Use an appropriate capnography simulator or gas source when available.

Expected outcome: The specific sampling or display failure is identified.

3. Inspect the Sampling Line and Consumables

Check the external sampling path for:

Replace blocked, contaminated, or damaged single-use components rather than attempting to clear them for reuse when not intended.

Expected outcome: The sampling path is clean, open, intact, and properly connected.

4. Verify Airway Adapter or Sampling Connection

Inspect the patient-end connection for secure placement and proper orientation.

Check whether the sampling inlet is positioned where exhaled gas can reliably reach it.

Expected outcome: Exhaled gas can enter the sampling system without a loose connection or obvious leak.

5. Reseat the Sampling-Line Connection

Disconnect and reconnect the line at the Propaq MD sampling port and patient-side connection as appropriate.

Inspect accessible connectors for contamination or damage.

Expected outcome: The sampling connection is secure and the device begins displaying a consistent capnogram. If normal operation returns, complete verification and troubleshooting can stop.

6. Substitute a Known-Good Sampling Line

Install a new or known-good compatible sampling line and associated consumables.

Do not assume an internal pump failure until the disposable sampling path has been eliminated.

Expected outcome: If the waveform and EtCO2 value return with the replacement line, the original consumable was the cause and troubleshooting can stop after verification.

7. Check for Moisture or Secretion Recurrence

If lines repeatedly obstruct, evaluate whether clinical setup, excessive moisture, secretion burden, or placement is contributing.

Do not bypass moisture-management components or modify the sampling system.

Expected outcome: The cause of repeated external blockage is corrected without altering the approved sampling configuration.

8. Test With Approved Capnography Test Equipment

Use an appropriate simulator or controlled CO2 source to verify:

Expected outcome: Stable simulated capnography confirms correct system operation.

9. Perform Final Functional Verification

After correction:

Expected outcome: Capnography operates consistently without sampling interruption. Troubleshooting can stop.

10. Escalate Persistent Sampling Failure

If the unit does not sample correctly with a verified external setup and known-good line, remove it from service.

Expected outcome: An internal pump, flow path, sensor, interface, configuration, or other service-level condition is evaluated by qualified personnel.

If the Problem Persists

Once sampling lines, airway adapters, moisture, secretions, connections, positioning, compatible consumables, and external test conditions have been ruled out, the remaining cause may involve the internal sampling system, sensing system, flow control, configuration, or another service-level condition.

The device should be:

Following service, complete applicable EtCO2, alarm, monitoring, defibrillator, and electrical safety testing before release.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Maintain alternate ventilation monitoring whenever the capnogram or EtCO2 value is unreliable or absent.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported no EtCO2 waveform despite the sampling line being connected to the Propaq MD."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the sampling line obstructed by moisture and confirmed normal operation with a known-good replacement line."

Resolution

What action was taken.

Example:
"The blocked sampling line was replaced and stable capnogram, EtCO2 values, and alarm operation were verified with approved test equipment."

Helpful Details to Include (If Known)

Final Thought

Maintain ventilation monitoring, inspect and substitute the complete external sampling path before suspecting the device, verify capnography with controlled test equipment, escalate unresolved failures, and document both the external cause and final performance.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide