On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Defibrillator will not charge, charge is interrupted, or shock cannot be delivered because of power, cables, electrodes, controls, or therapy-system problems.
Step-by-Step Troubleshooting
1. Protect the Patient and Maintain Immediate Defibrillation Capability
A shock charging or delivery problem makes the unit unreliable for emergency therapy.
Immediately place another verified defibrillator in service before troubleshooting.
Do not perform troubleshooting while a patient depends on the affected unit for emergency defibrillation.
Expected outcome: Effective defibrillation capability remains immediately available.
2. Confirm the Exact Therapy Complaint
Determine whether the issue is:
- Unit will not begin charging
- Charging stops unexpectedly
- Charge does not remain available
- Shock control does not respond
- Shock cannot be delivered
- Therapy electrodes are not recognized
- Problem occurs only on battery power or only on AC power
Reproduce the condition only with an approved defibrillator analyzer or test load.
Expected outcome: The specific charging or delivery failure is defined without exposing a patient or staff to unnecessary risk.
3. Verify Basic Power Condition
Check whether the Propaq MD operates normally on its available power sources.
Inspect battery seating and external power connections as applicable.
A marginal battery or unstable power source may contribute to therapy problems.
Expected outcome: The device has a stable power source for controlled defibrillator testing.
4. Inspect Therapy Electrodes or Test Interface
Check therapy pads or analyzer connections for:
- Damage
- Expiration
- Poor connection
- Contamination
- Incorrect mating
- Compromised packaging if using disposable pads
For bench testing, use the approved analyzer interface.
Expected outcome: The therapy path has appropriate external electrodes or a properly connected test load.
5. Inspect the Therapy Cable
Examine the external therapy cable for:
- Cracks
- Damaged insulation
- Connector damage
- Bent contacts
- Loose locking features
- Strain damage
- Intermittency
Expected outcome: The cable is intact and securely connected.
6. Reseat Therapy Connections
Reconnect the therapy cable and analyzer or test-load interface.
Verify full engagement at all accessible connection points.
Expected outcome: A loose connection is eliminated. If charging and discharge now function normally, complete repeated analyzer testing and troubleshooting can stop.
7. Verify Operating Controls
Confirm the device is in the intended defibrillation mode and that the relevant external controls respond normally.
Do not access unauthorized service menus or alter protected therapy configuration.
Expected outcome: Controls needed for charging and controlled discharge respond normally.
8. Test Charging With an Approved Defibrillator Analyzer
Using approved test equipment, initiate controlled charging and observe whether the unit:
- Begins charging
- Completes the charge cycle
- Maintains the charged state as expected for testing
- Responds to discharge controls
Follow facility and manufacturer safety procedures for defibrillator testing.
Expected outcome: The unit completes a consistent charge cycle without interruption.
9. Verify Delivered Energy With Approved Test Equipment
Discharge only into an approved defibrillator analyzer or test load.
Verify repeated charge-and-discharge cycles at appropriate test settings established by the manufacturer's service procedure or facility testing protocol.
Do not invent or substitute acceptance tolerances.
Expected outcome: The analyzer confirms consistent shock delivery within the applicable manufacturer-defined acceptance criteria. Troubleshooting can stop when all required testing passes.
10. Escalate Any Persistent Therapy Failure
If charging or shock delivery remains unreliable after verifying power, therapy cables, connections, controls, and analyzer setup, stop troubleshooting.
Expected outcome: The defibrillator remains out of clinical service until the therapy system is evaluated and repaired by qualified personnel.
If the Problem Persists
After external power, batteries, therapy electrodes, cables, connectors, controls, and approved analyzer setup have been ruled out, the remaining cause may involve the high-energy therapy system, charging circuitry, discharge control, internal sensing, configuration, or another service-level condition.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate manufacturer documentation and approved test equipment
- Repaired or configured only by qualified personnel
Do not perform internal high-voltage troubleshooting beyond authorized service procedures.
Following repair, complete all applicable charge-time, delivered-energy, pacing, ECG, alarm, battery, and electrical safety return-to-service testing before clinical release.
Knowing when to stop troubleshooting an unreliable defibrillation system is proper troubleshooting.
Clinical Use Tip
Any uncertainty about shock charging or delivery warrants immediate replacement with another verified defibrillator.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the Propaq MD intermittently failed to charge for defibrillation."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found an intermittent external therapy cable connection and reproduced the charging problem during analyzer testing."
Resolution
What action was taken.
Example:
"The defective therapy cable was replaced and repeated charging and shock-delivery tests passed using an approved defibrillator analyzer."
Helpful Details to Include (If Known)
- Charge failure behavior
- Shock delivery behavior
- AC or battery operation
- Battery condition
- Therapy pad or analyzer connection
- Therapy cable condition
- Control response
- Known-good accessory substitution
- Analyzer results
- Repeated charge/discharge results
- Final device status
Final Thought
Treat any charge or shock-delivery complaint as a critical safety issue, establish alternate defibrillation capability, verify external power and therapy connections first, test only with approved equipment, escalate unresolved failures, and document complete return-to-service testing.
That is successful troubleshooting.