ZOLL R Series

Error: PACER FAULT 115

On this page

Asset Type

Defibrillator

Manufacturer

ZOLL

Model

R Series

What This Guide Helps With

This guide helps troubleshoot PACER FAULT 115 on a ZOLL R Series defibrillator. This error indicates a pacer failure during POST, which may disable or prevent reliable transcutaneous pacing. Because pacing is a critical therapy function, the device should not be trusted for patient pacing until the fault is evaluated, external causes are ruled out, and required operational checks pass.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check Battery and AC Power

Inspect External Pacing Accessories

Check for Physical Damage or Contamination

Attempt a Controlled Power Cycle

Run the Appropriate Readiness or Operational Check

Isolate the Failure Pattern

Stop When Resolved

If the Problem Persists

If PACER FAULT 115 continues after confirming power, battery, therapy cable, pads or test load, connector condition, and physical condition, remove the defibrillator from service. External causes have been ruled out, and the repeated POST fault indicates a likely internal pacer or therapy subsystem failure.

Clinical Use Tip

Never troubleshoot PACER FAULT 115 on an active patient. Move the patient to a backup defibrillator or pacing-capable device first, then evaluate the failed unit. A defibrillator with unresolved pacing faults should not be trusted for emergency pacing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ZOLL R Series reported PACER FAULT 115 during startup/POST. Pacing function may be disabled or unavailable. Unit removed from clinical use for evaluation."

Cause

What was observed during troubleshooting.

Example:
"PACER FAULT 115 reproduced during startup. Battery, AC power source, therapy cable, pads/test load, and external connections checked. Fault persisted, indicating likely internal pacer/therapy subsystem failure."

Resolution

What action was taken.

Example:
"Removed device from service, labeled Out of Service, and sent for bench evaluation/repair. Clinical area notified to use alternate pacing-capable defibrillator."

Helpful Details to Include (If Known)

Final Thought

PACER FAULT 115 should be treated seriously because it involves a critical therapy function. Start with patient safety, verify the exact error, rule out simple external causes, and avoid chasing internal repairs in the clinical area. If the fault persists after basic isolation, escalation is the correct repair path. Accurate CCR documentation protects patients, clinical staff, and the equipment history.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide