ZOLL R Series

Error: PACER FAULT 117

On this page

Asset Type

Defibrillator

Manufacturer

ZOLL

Model

R Series

What This Guide Helps With

This guide helps troubleshoot PACER FAULT 117 on a ZOLL R Series defibrillator. This error indicates a pacer functional safety failure while the device is running, which may disable or compromise transcutaneous pacing during operation.

Because pacing is a life-support function, this fault should be treated as a serious patient safety issue. The device should not be trusted for pacing until external causes are ruled out and the unit passes required testing or receives qualified service evaluation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Remove the Device From Clinical Service

Check Power Source and Battery Condition

Check Therapy Cable and Pacing Accessories

Check for Obvious Physical or Fluid Damage

Run an Operational Check in a Bench Environment

Determine Whether the Issue Is Accessory-Related or Device-Related

If the Problem Persists

If PACER FAULT 117 remains after checking the power source, battery, therapy cable, pacing accessories, test load, and external condition, the common external causes have been ruled out.

A persistent PACER FAULT 117 should be treated as an internal pacing or therapy subsystem failure.

Clinical Use Tip

Never troubleshoot PACER FAULT 117 on an active patient. If pacing may be required, move the patient to a backup defibrillator/pacer first and continue patient care with reliable equipment.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ZOLL R Series defibrillator reported PACER FAULT 117. Staff reported a pacer fault while the unit was running, creating concern that transcutaneous pacing may be unavailable or unsafe."

Cause

What was observed during troubleshooting.

Example:
"PACER FAULT 117 was confirmed during bench evaluation. Battery, AC power, therapy cable, and external pacing accessories were checked. Fault remained with known-good external setup, indicating likely internal pacer or therapy subsystem failure."

Resolution

What action was taken.

Example:
"Device was removed from service and labeled Out of Service. Replacement defibrillator was placed in the clinical area. Unit was sent for repair or qualified bench evaluation due to persistent pacer functional safety fault."

Helpful Details to Include (If Known)

Final Thought

PACER FAULT 117 should be treated seriously because it involves the pacing function of a defibrillator. Clinical Engineering should confirm the error, rule out simple external causes, protect patient safety, and escalate the device when the fault persists. Clear CCR documentation helps show what was checked, why the unit was removed, and what action was taken.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide