On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
External pacing does not start, output appears ineffective, or capture cannot be established because of pads, connections, controls, positioning, or device problems.
Step-by-Step Troubleshooting
1. Protect the Patient and Maintain Emergency Therapy
External pacing is a critical therapeutic function. Do not troubleshoot unreliable pacing while a patient depends on the suspect device.
Immediately transition to another verified pacing-capable defibrillator or appropriate clinical backup when pacing is required.
Expected outcome: Patient therapy remains uninterrupted using reliable equipment.
2. Confirm the Exact Pacing Problem
Determine whether the pacer will not activate, pacing markers are absent, output cannot be adjusted, pads are not recognized, the device appears to deliver pacing but electrical capture is absent, or the pacing function stops unexpectedly.
Differentiate device output concerns from clinical non-capture.
Expected outcome: The problem is defined without assuming the device is at fault for a clinical capture issue.
3. Inspect Therapy Pads
Verify compatible multifunction therapy electrodes are connected, properly positioned, within usable condition, and making adequate patient contact.
Inspect packaging and electrode condition when relevant and replace questionable pads.
Expected outcome: Therapy pads are correctly applied and recognized. If correcting pad condition or placement resolves the problem, continue to final verification.
4. Inspect Therapy Cable and Connections
Trace the therapy connection from the pads through the cable to the defibrillator.
Inspect connectors and cable insulation for contamination, looseness, bent contacts, physical damage, or strain-related intermittency.
Reconnect securely.
Expected outcome: The complete therapy connection is intact and stable.
5. Verify Pacing Controls and Mode
Confirm the device is placed in the intended pacing mode and that required controls are responsive.
Verify the selected pacing rate and output settings are clinically appropriate under qualified clinical direction.
Do not use unauthorized service settings.
Expected outcome: The pacer activates and responds normally to approved front-panel controls.
6. Evaluate ECG Sensing When Relevant
If the pacing mode depends on ECG sensing, verify a stable ECG signal and correct lead connections.
Excessive artifact or lead-off conditions may interfere with pacing behavior.
Expected outcome: A stable ECG signal is present and pacing behavior is consistent with the selected mode.
7. Test With Known-Good Therapy Accessories
When clinically removed from use, test the device using approved defibrillator/pacer test equipment and known-good compatible therapy accessories.
Do not test therapeutic output directly on a patient as part of equipment troubleshooting.
Expected outcome: Pacing output is detected and responds appropriately during bench testing. If the original accessories were defective, replace them and proceed to final verification.
8. Perform Final Pacing Verification
Using approved test equipment and current manufacturer procedures, verify pacing activation, output response, rate control, applicable indicators, and related alarm behavior.
Expected outcome: Pacing function operates consistently through the tested range required by the approved procedure. If all checks pass, troubleshooting is complete.
If the Problem Persists
Common external causes involving therapy pads, therapy cable, ECG sensing, controls, and connections have been ruled out. The remaining cause may involve the therapy-output circuit, control interface, internal sensing, configuration, or another service-level fault.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate manufacturer documentation and approved test equipment
- Repaired or configured only by qualified personnel
Do not continue into internal high-voltage or board-level therapeutic-output repairs as routine troubleshooting. Complete full pacing and defibrillation performance testing before return to service.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
When pacing is clinically required, move therapy to another verified device before investigating whether the problem is accessory-related, sensing-related, or internal.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported external pacing would not start and the connected therapy pads were intermittently not recognized."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found damage at the therapy cable connector causing intermittent pad recognition."
Resolution
What action was taken.
Example:
"Replaced the defective therapy cable and verified pacing output, controls, and therapy-pad recognition using approved test equipment."
Helpful Details to Include (If Known)
- Exact pacing symptom
- Pads recognized or not recognized
- Therapy-pad condition
- Therapy-cable condition
- ECG sensing status
- Pacing mode observed
- Control response
- Known-good accessory results
- Pacer analyzer results
- Final device status
Final Thought
External pacing is a critical therapy, so protect the patient first, verify pads, cables, ECG sensing, and controls before assuming internal failure, use approved test equipment, and escalate immediately when reliable output cannot be confirmed.
That is successful troubleshooting.