On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting absent electrical or mechanical capture, pacing-electrode connection problems, poor pad contact, incorrect settings, or suspected pacer-output failure.
Step-by-Step Troubleshooting
1. Ensure Patient Safety First
Do not perform extended pacing troubleshooting while the X Series is the only device supporting an unstable patient.
If external pacing is unsuccessful during clinical use:
- Notify the responsible clinical provider immediately.
- Continue ECG, pulse, blood-pressure, oxygen-saturation, and patient assessment.
- Prepare another verified defibrillator with external pacing capability.
- Follow the facility’s emergency bradycardia and equipment-failure procedures.
- Transfer pacing support to another device when capture cannot be established promptly.
Expected outcome: Patient care does not remain dependent on equipment with uncertain pacing performance.
Continue Clinical Engineering troubleshooting only after the device is safely removed from patient use.
2. Confirm the Exact Reported Failure
Determine whether the complaint involves:
- Pacer mode will not activate.
- The pacing rate or current cannot be adjusted.
- No pacing markers appear on the ECG display.
- Pacing markers appear, but no ECG capture occurs.
- ECG capture appears without a corresponding pulse.
- A pads-off, poor-contact, or electrode-related message appears.
- The pacing current stops unexpectedly.
Record the selected pacing rate, selected current in milliamperes, ECG lead, electrode type, displayed messages, and whether the problem occurred continuously or intermittently.
Expected outcome: The complaint is separated into an operating, connection, electrical-capture, mechanical-capture, or output-performance problem.
3. Verify the Device Is in Pacer Mode
Confirm that the X Series is operating in manual mode and that pacing has been selected according to the facility-approved operating process.
Verify that:
- The pacer controls are displayed.
- A pacing rate is selected.
- Pacing current is above zero.
- The pacer is actively running rather than paused.
Expected outcome: Pacing markers and the selected pacing settings appear on the display.
If the controls now function normally, complete an operational check and stop troubleshooting.
4. Check the Therapy Cable Connection
Inspect the multifunction therapy cable and confirm that it is fully seated in the X Series therapy connector.
Check for:
- A partially inserted connector.
- A loose locking connection.
- Bent, recessed, contaminated, or damaged contacts.
- Cracked connector housings.
- Cable cuts, crushed sections, severe kinks, or exposed conductors.
Disconnect and reconnect the cable only when the device is not connected to a patient.
Expected outcome: The therapy cable connects securely without movement, intermittent messages, or visible damage.
If reseating the connection restores stable pacing operation, verify the device with an approved analyzer and stop.
5. Verify the Correct Pacing Electrodes Are Installed
Confirm that compatible ZOLL multifunction electrodes approved for pacing are being used and that they are within their expiration date.
Inspect the package and electrodes for:
- Open or damaged packaging.
- Dried or uneven conductive gel.
- Folded, contaminated, or damaged adhesive surfaces.
- Incorrect adult or pediatric electrode selection.
- Damage to the electrode lead or connector.
ZOLL multifunction electrodes provide the current path used for external pacing, so defective, expired, or poorly connected electrodes can prevent effective current delivery.
Expected outcome: A new, compatible, undamaged electrode set is connected securely.
If replacing the electrodes resolves the issue, remove the defective set from use and document the lot and expiration information when available.
6. Check Electrode Placement and Contact
When evaluating a clinical report, confirm that the patient’s skin was exposed, dry, and prepared according to facility protocol.
Verify that:
- Electrodes were positioned as directed on their packaging.
- The entire conductive surface adhered to the skin.
- Excessive hair, moisture, oils, medication patches, clothing, or trapped air did not interfere with contact.
- The pads were not overlapping or touching.
- The placement created an appropriate current path through the heart.
Poor adhesion, moisture, hair, or improper positioning can increase transthoracic resistance and reduce current delivery.
Expected outcome: Electrodes have complete skin contact and the pads-off or poor-contact indication clears.
Clinical Engineering should not reproduce pacing on a person. Use an approved defibrillator/pacer analyzer for bench verification.
7. Confirm the ECG Signal Is Suitable for Pacing
Inspect the displayed ECG waveform.
Verify that:
- The ECG waveform is visible and reasonably free of artifact.
- The selected ECG lead produces clear QRS complexes.
- Lead wires and electrodes are securely connected when separate ECG leads are used.
- Pacing spikes are not being confused with motion artifact or unrelated ECG noise.
- The display gain allows the technician or clinician to distinguish intrinsic complexes from paced complexes.
Expected outcome: The underlying rhythm and pacing markers can be evaluated reliably.
If changing a loose ECG cable, electrode, or lead selection restores a stable display, repeat the functional test and stop.
8. Distinguish Electrical Capture From Mechanical Capture
Electrical capture is suggested when each pacing marker is followed by a consistent widened QRS complex or other expected depolarization pattern.
Mechanical capture requires evidence of effective cardiac output, such as:
- A corresponding palpable pulse.
- A pulse waveform matching the paced rate.
- A blood-pressure response.
- Other approved clinical confirmation.
Do not assume capture based only on the presence of pacing markers. Pacing markers indicate a pacing command, not necessarily effective myocardial capture.
Expected outcome: The reported problem is correctly identified as failure of electrical capture, failure of mechanical capture, or a display-interpretation issue.
Clinical Engineering should document the reported clinical behavior but should not determine patient treatment settings.
9. Review the Reported Pacing Current
Confirm whether the clinician increased pacing current gradually according to clinical protocol until capture was achieved or the condition was determined to be nonresponsive.
A low selected current may produce visible pacing markers without sufficient delivered current to achieve capture. However, the required current varies by patient and is a clinical determination.
Expected outcome: The problem is not caused simply by the pacing current remaining at zero or an unintentionally low setting.
Do not direct Clinical Engineering staff to establish patient-specific pacing thresholds.
10. Check for Patient-Related High-Impedance Conditions
Review whether the failed capture occurred under conditions that could increase resistance or make transcutaneous capture difficult, including:
- Incomplete electrode adhesion.
- Large body habitus.
- Excessive chest hair.
- Very dry or damaged electrodes.
- Incorrect electrode positioning.
- Patient movement.
- Severe physiologic or metabolic conditions reported by the clinical team.
Expected outcome: A patient-related capture difficulty is not incorrectly classified as equipment failure.
The device must still be bench tested before being returned to service.
11. Test With an Approved Defibrillator/Pacer Analyzer
Remove the X Series from clinical use and connect it to a calibrated analyzer that supports transcutaneous pacing measurements.
Following the analyzer manufacturer’s instructions and the facility’s approved test procedure:
- Connect the multifunction therapy cable to the analyzer.
- Select pacer mode.
- Test multiple pacing rates.
- Test multiple output-current settings.
- Confirm that the analyzer detects pacing pulses.
- Compare measured rate and current with displayed settings.
- Check for stable output without unexpected interruption.
- Gently manipulate the external cable near strain-relief points while observing for intermittent output.
Expected outcome: The measured pacing rate and output remain stable and within the facility’s approved acceptance criteria.
Do not use improvised resistive loads, handheld meters, or direct contact methods to measure pacing output.
12. Substitute Known-Good External Accessories
When the output is absent or intermittent, repeat the analyzer test using:
- A known-good compatible therapy cable.
- A known-good electrode adapter or test connection, when applicable.
- A second approved analyzer channel or cable when available.
Change one component at a time.
Expected outcome: The failed component is isolated without assuming an internal defibrillator failure.
If pacing operates normally with the known-good cable, remove the original cable from service and replace it.
13. Inspect Power and Battery Status
Confirm that:
- A charged, approved battery is installed.
- The battery is fully seated.
- No power or battery warning is present.
- The device remains powered during pacing-output testing.
- External power, when used, is connected securely.
Expected outcome: Pacing is not being interrupted by unstable device power.
If the unit restarts, shuts down, or changes operating state during testing, discontinue use and evaluate the power system separately.
14. Perform the Approved Operational or Readiness Check
Run the facility-approved X Series operational check and review the device for active service indicators or failed readiness results.
Verify that defibrillation, monitoring, and pacing-related accessories are correctly connected for the test being performed.
Expected outcome: The X Series completes the approved checks without errors.
A successful automated check does not override a failed measured pacing-output test or a reproducible intermittent complaint.
15. Compare With the Original Complaint
After testing, determine whether the failure:
- Was corrected by replacing electrodes.
- Was corrected by replacing the therapy cable.
- Was caused by poor electrode contact or incorrect setup.
- Could not be reproduced and all analyzer measurements passed.
- Remains reproducible with known-good accessories.
- Includes unstable, absent, or inaccurate pacing output.
Expected outcome: The final device disposition is supported by repeatable evidence.
If the device passes all testing but the original complaint involved intermittent behavior, document the exact accessories tested and consult the reporting department before returning the unit to service.
If the Problem Persists
If the X Series produces no pacing output, inaccurate output, intermittent pulses, unresponsive controls, repeated therapy-cable errors, or failed analyzer measurements with known-good accessories, common external causes have been ruled out.
The problem may involve the therapy connector, pacer-output circuitry, control interface, power system, or another internal component.
The device should be:
- Removed from service.
- Labeled Out of Service.
- Kept with the suspect cable or accessories when they may be involved.
- Sent for qualified bench evaluation or ZOLL-authorized repair.
- Tested for pacing, defibrillation, monitoring, alarms, and electrical safety before return to clinical use.
Do not perform unapproved internal board-level repairs. Knowing when to stop and escalate is proper troubleshooting.
Clinical Use Tip
Never troubleshoot failed capture on an active patient beyond immediate external checks needed to maintain care. Move pacing support to another verified device before investigating cables, electrodes, settings, or output performance.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the ZOLL X Series displayed pacing markers but external pacing did not produce consistent capture."
Cause
What was observed during troubleshooting.
Example:
"Bench testing found intermittent pacing output when the multifunction therapy cable was moved near its connector strain relief."
Resolution
What action was taken.
Example:
"Removed the damaged therapy cable from service, installed a verified replacement, and confirmed stable pacing rate and current at multiple settings with a calibrated defibrillator/pacer analyzer."
Helpful Details to Include (If Known)
- Complete reported alarm or message
- Pacing rate selected
- Pacing current selected
- ECG lead used
- Presence of pacing markers
- Electrical capture observed
- Mechanical capture reported
- Electrode type, lot, and expiration date
- Electrode placement used
- Therapy cable condition
- Known-good cable substitution result
- Analyzer model and asset number
- Measured pacing rate
- Measured pacing current
- Battery and external-power status
- Operational-check result
- Intermittent behavior during cable movement
- Final device status
Final Thought
External pacing complaints require a clear distinction between equipment output, electrode contact, electrical capture, and mechanical capture. Patient safety comes first, followed by systematic accessory checks and calibrated analyzer testing. Escalate reproducible internal failures and document the evidence clearly.
That is successful troubleshooting.