Baxter PrismaFlex

Preventive Maintenance Procedure

On this page

Asset Type

Continuous Renal Replacement Therapy (CRRT)

Manufacturer

Baxter

Model

PrismaFlex

Manual Reference

Prismaflex Service Manual, Gambro Lundia AB, Order No. G5005209, for software version 7.xx.

Manual identification note: The provided service manual identifies the original manufacturer as Gambro Lundia AB and the device as Prismaflex®. This procedure uses Baxter / PrismaFlex as provided for the asset record.

Manufacturer Recommendation / AEM Status

The manual lists preventive maintenance for the PrismaFlex control unit as required every 6000 hours of operation or once per year, whichever occurs first. Upcoming and overdue PM procedures are displayed to the operator through a reminder screen at startup. The manual states that only authorized service technicians are allowed to perform preventive maintenance and that the PM checklist supplied with the PM kit is to be completed during testing.

Manufacturer-Listed PM Items

Manufacturer-Listed Frequency

Every 6000 operating hours or once per year, whichever occurs first.

Battery Replacement Interval

The manual lists an expected backup battery life of 3 to 5 years for both 12 V and 24 V backup battery types. It states that changing the backup batteries every third year under normal use and storage conditions will ensure operation according to specification, but identifies this battery change as guidance rather than a mandatory patient-safety requirement.

PM Kit Contents Listed in the Manual

Part number note: The provided service manual chapter lists PM kit contents but does not provide a PM kit part number or individual PM replacement part numbers.

Post-Repair Electrical Safety Inspection Requirements

The manual lists a full Electrical Safety Inspection after replacement of any of the following components:

The manual also lists full Electrical Safety Inspection after unexpected electrical events on the mains supply or unintentional fluid ingress.

AEM Note

If a facility performs this PM at a different interval or changes the listed activities, that should be handled under the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Tools

Test Equipment

Functional Checkout Supplies

PM Replacement Materials

Manufacturer PM kit contents listed above.

Manual caution: Do not use powered screwdrivers or drills when mounting internal components. The manual warns that damage to standoffs or plastic components is likely.

Step-by-Step PM Procedure

1. Confirm PM Eligibility and Remove the Unit From Use

Action: Confirm the unit is available for scheduled maintenance, document asset identification, software version if available, current operating hours, and the reported PM due status. Remove the unit from clinical availability for the duration of PM.

Expected outcome: Unit is identified, not connected to a patient, and available for safe service.

Why it matters: The functional checkout and internal PM work must not be performed while the device is in patient use.

2. Prepare for Internal Service

Action: Power the unit off and disconnect the power cord from the wall outlet before opening the rear panel. Apply appropriate ESD protection before accessing internal components.

Expected outcome: Unit is electrically isolated for internal inspection and protected from electrostatic damage.

Why it matters: The manual specifically requires electrostatic safety devices when working inside the PrismaFlex control unit.

3. Perform Exterior and Interior Visual Inspection and Cleaning

Action: Open the rear panel using the 8 mm hex key. Clean dust, debris, and dried fluid from the external and internal surfaces, including:

Clean spills from the control unit surface using a mild detergent. Do not use detergents with germicides where the manual prohibits them. Clean the blood leak detector tubing path as needed using a lint-free cloth and 70% isopropyl alcohol with a flossing action; dry thoroughly.

Inspect:

Expected outcome: Unit is clean, structurally intact, free of fluid contamination, and without visible conditions requiring correction.

Do not return to service if: Fluid contamination has entered a pressure sensor assembly or return pressure port and corrective replacement/testing has not been completed.

Fluid or Blood Contamination Conditions

If blood leakage is present from an access or filter pressure pod diaphragm:

If blood or infusion solution has passed through the fluid barrier and reached the return pressure port:

4. Perform Electrical Safety Inspection

Action: Complete the Electrical Safety Testing section below as part of scheduled PM.

Expected outcome: Visual inspection, protective earth, conductivity clip, earth leakage, and patient leakage tests meet manufacturer-listed limits.

Why it matters: The manual states that Electrical Safety Inspection should be performed at every maintenance service.

5. Replace Manufacturer-Listed PM Kit Components

5.1 Replace PC 104 Board Battery

Action:

Expected outcome: PC 104 board battery is replaced and installed with correct polarity.

5.2 Replace Pressure Pod Sealing Cones

Action:

Expected outcome: Four new sealing cones are installed correctly.

Important: Do not use lubricant.

5.3 Replace Blood Pump Rotor Dampers

Action:

Expected outcome: Two new blood pump rotor dampers are installed and rotor assembly is secure.

5.4 Replace Slave Pump Rotor Dampers

Action:

Expected outcome: Eight slave pump dampers are replaced and affected rotors are reinstalled.

Important: Do not adjust the rotor screws.

5.5 Empty or Exchange Technical Data Card

Action: Exchange the technical data card or download its stored data to a computer. Confirm locally whether historical data must be retained before clearing card contents and reinstalling the card.

Expected outcome: Technical data card is available for future logging and required historical data has been handled appropriately.

5.6 Replace ARPS Pump Segment and Air Filter

Action:

Expected outcome: New ARPS pump segment and filter are installed with tubing correctly routed and free of defects.

6. Perform Power Supply and Backup Battery Checks

Action: Turn on the PrismaFlex control unit. Wait for the Query or Setup screen. Measure and document the following values:

Test Location Signal Manufacturer-Listed Acceptance Range Result
PIB Board TP14 to TP3 +5VD 5.1 to 5.3 V
PIB Board TP15 to TP3 +12 V 11.9 to 12.3 V
PIB Board TP13 to TP3 -5 V -5.4 to -5.1 V
PIB Board TP16 to TP3 +24Vm 22.8 to 25.2 V
Carrier Board TP2 to TP4 +5VD 5.0 to 5.3 V
Battery connector CB, if 12 V battery installed Battery 13.0 to 14.0 V
Battery connector CB, if 24 V battery installed Battery 27.5 to 28.8 V

Expected outcome: All applicable measured voltage values are within the manufacturer-listed ranges.

7. Complete Service Diagnose Mode Checkout

Action: Enter Service – Diagnose mode and verify proper operation and/or calibration of:

Correct any item that is out of calibration or specification, then repeat the affected verification.

Expected outcome: All listed service diagnostic checks pass without unresolved calibration or function errors.

8. Perform Functional Checkout

Action: Perform the manufacturer-listed functional checkout without a patient connected.

Warning: A patient must not be connected during this test. Use a container of water in place of the patient. A Malfunction alarm during the checkout is a failure condition; do not return the device to use until corrected and the checkout passes.

Setup

Treatment Settings

After prime and prime test complete:

Loss/Gain Limit: 140 mL / 3 hr.

Set the following flow values:

Parameter Required Test Setting
Blood 180 mL/min
PBP 1100 mL/hr
Dialysate 1200 mL/hr
Replacement 1300 mL/hr
Fluid Removal Rate 200 mL/hr

Run Test

Note: The manual identifies that Return Disconnection or Cannot Detect Return alarms may occur because water is used for the test. Continue or override according to the screen instruction; these alarms do not affect the functional checkout outcome.

Access Pressure Alarm Verification

Battery Backup Verification, If Installed

Finish Functional Checkout

Expected outcome: Functional checkout completes without a Malfunction alarm; access pressure alarm behavior and battery backup behavior, when applicable, operate as expected.

9. Complete PM Timer, Sticker, and Records

Action:

Expected outcome: Device PM status is updated, visible PM labeling is applied, and service documentation is complete.

Manual-Listed PM Timer Ranges

For the manufacturer-listed PM schedule, set the PM advisory consistent with once per year or 6000 operating hours, whichever occurs first.

Electrical Safety Testing

The manual lists Electrical Safety Inspection at each maintenance service and includes visual inspection, protective earth testing, conductivity clip testing, earth leakage current testing, and patient leakage current testing. It also specifies that no insulation test shall be performed during ESI.

Electrical Safety Visual Inspection

Inspect and document:

Result: Pass / Fail =

Protective Earth Test - PET

Test condition: Unit disconnected from mains power.

Test equipment: Safety tester according to IEC 62353.

Ground resistance measured in ohms, expected ≤ 0.5 Ω or facility/NFPA 99 criteria if different. The manual lists a more restrictive protective earth resistance requirement of ≤ 300 mΩ, equivalent to ≤ 0.300 Ω, at each listed test point.

Test Point From Mains Plug Protective Earth To: Expected Result Measured Result
Earth connector at bottom plate where scales are attached ≤ 0.300 Ω
Earth connector at PCB holder at the door ≤ 0.300 Ω
Earth connector at ARPS pump ≤ 0.300 Ω
Earth connector for potential equalization ≤ 0.300 Ω

Pass / Fail =

Conductivity Clip Test

Test equipment: Multimeter capable of measuring resistance of 10 MΩ or greater with ±2% accuracy.

Action: With the unit initialized and in Setup mode, measure resistance between the conductivity clip and the connector for potential equalization.

Applied Part Classification Manufacturer-Listed Expected Result Measured Result
BF 1.1 MΩ ± 0.2 MΩ
CF 3.24 MΩ ± 0.15 MΩ

Applied part classification used =

Pass / Fail =

Earth Leakage Current Test - ELT

Test equipment: Safety tester set to measure according to IEC 60601-1.

Condition: Measure Normal Condition and Single Fault Condition, including reversed supply polarity. Record the highest reading.

Touch/chassis leakage measured in microamps:

Facility limit used =

Result =

Pass / Fail =

Supply Voltage Condition Manufacturer-Listed Limit Highest Measured Result
115 V AC Normal Condition ≤ 250 μA
115 V AC Single Fault Condition ≤ 500 μA
230 V AC Normal Condition ≤ 500 μA
230 V AC Single Fault Condition ≤ 1000 μA

Supply voltage used =

Applicable limit used =

Pass / Fail =

Patient Leakage Current Test - PLT

Action: Use the conductivity clip as the patient leakage test access point. Measure Normal Condition and Single Fault Condition, including reversed supply polarity. Record the highest DC and AC readings.

Applied Part Classification Condition Manufacturer-Listed Limit
CF Normal Condition DC ≤ 10 μA; AC ≤ 10 μA
CF Single Fault Condition DC ≤ 50 μA; AC ≤ 50 μA
BF Normal Condition DC ≤ 10 μA; AC ≤ 100 μA
BF Single Fault Condition DC ≤ 50 μA; AC ≤ 500 μA

Applied part classification used =

Normal condition DC result =

Normal condition AC result =

Single fault condition DC result =

Single fault condition AC result =

Pass / Fail =

Pass / Fail Criteria

PM Pass Criteria

The PrismaFlex passes scheduled PM when:

Remove From Service Criteria

Remove from service or do not return to service when:

Copy/Paste CMMS Work Order Procedure

Post-Repair Electrical Safety Inspection - Copy/Paste Block

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide