Baxter SIGMA Spectrum

Preventive Maintenance Procedure

On this page

Asset Type

Infusion Pump

Manufacturer

Baxter

Model

SIGMA Spectrum

Manual Reference

Baxter Manual 41394, Revision A, July 2017, SIGMA Spectrum Infusion Pump with Master Drug Library Preventive Maintenance, Pump Operating Software Versions 6.02 and 6.05, Master Drug Library Version 6.

Manufacturer Recommendation / AEM Status

Baxter lists preventive maintenance for all pumps in service, any pump suspected of physical damage, being dropped, or fluid intrusion, and pumps requiring inspection, testing, performance evaluation, or service following a failed PM item.

Manufacturer-Listed PM Frequency

Manufacturer-Listed Annual PM Activities

Baxter instructs the facility to record the PM findings and return any pump that fails an inspection, test, or performance evaluation to Baxter for service.

Manufacturer-Listed Parts and Test Accessories

The provided manual does not list a routine battery-replacement interval, calibration interval, or electrical-safety testing requirement.

If the facility performs this PM at a different interval or changes the listed activities, document the change under the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Gravimetric Testing Equipment

Volumetric Testing Equipment

All measurement equipment should be traceably calibrated to NIST standards.

Do Not Use

Step-by-Step PM Procedure

1. Identify the Pump

Power on the pump and record:

Verify that the displayed and labeled serial numbers match.

Expected outcome: Identification information is present, legible, and consistent.

2. Verify the Active Drug Library

Record the active Drug Library BDL file name and date from the Power Status or User Options screen.

Verify the installed file against the facility’s current approved drug library through Pharmacy or the Master Drug Library Administrator.

Expected outcome: Pump and facility drug-library names and dates match.

A missing, outdated, or incorrect drug library must be corrected according to the operator manual and facility drug-library deployment process before clinical use.

3. Verify Network Connectivity

For pumps equipped with a Wireless Battery Module:

  1. Enter Biomed Options.
  2. Open Network Configuration.
  3. View Network Status.
  4. Confirm that the network configuration is installed.
  5. Confirm connectivity, indicated by the appropriate wireless-status icon.

Expected outcome: The pump shows the expected network configuration and connectivity status.

Document “Not equipped” when no Wireless Battery Module is installed.

4. Inspect the Pump

Inspect for damage, contamination, heat damage, fluid intrusion, cracks, cuts, dents, or missing components.

Inspect:

Confirm the tube channel is clean and dry.

Expected outcome: No damage, contamination, fluid intrusion, or unreadable labels are present.

5. Inspect the AC Power Adapter

Inspect:

Connect the adapter to a working receptacle and verify:

Expected outcome: The adapter is undamaged, properly labeled, and recognized by the pump.

A damaged adapter may be sent separately for service without the pump.

6. Perform the Keypad Test

Enter Basic Mode and verify keypad entry using the manufacturer procedure:

  1. Enter 90.
  2. Enter 5678.
  3. Enter 1234.
  4. Verify the display correctly registers each entry.
  5. Verify the hard-limit message appears when RUN/STOP is pressed with an entry above the allowed value.
  6. Clear the program and return to the Care Area screen.

Expected outcome: All keys, soft keys, decimal input, display fields, and prompts operate correctly.

Repeat once when a failure is suspected. Remove the pump from service if the failure repeats.

7. Perform the Memory Test

Program:

Power the pump off and back on. Select “No” when asked whether this is a new patient.

Verify that the programmed rate and VTBI were retained.

Expected outcome: Rate remains 100 mL/hr and VTBI remains 50 mL.

8. Prepare the Hydraulic Test Setup

Use a new compatible IV set. Configure the system as follows:

Avoid bright light directly over the pump keyhole because it may interfere with slide-clamp recognition.

Expected outcome: The test setup matches the manufacturer diagram and contains no air, occlusions, contamination, or incorrect tubing.

9. Test the Downstream Occlusion Sensor

Before testing:

Program:

Connect the primed pressure gauge using the three-way valve.

Maintain approximately:

Start the infusion and record the pressure at which the downstream occlusion alarm occurs.

Pass range: 48–131 kPa / 7–19 psi.

Remove the pressure restriction and verify that the pump automatically restarts.

Expected outcome: The downstream occlusion alarm occurs within range and auto-restart functions after the occlusion is removed.

Repeat a failed test using a new unused IV set before declaring the pump failed.

10. Test the Upstream Occlusion Sensor

Program:

Clear the volume-given value.

Occlude the IV set with a hemostat approximately 30.2 cm / 12 inches above the pump, then start the infusion.

Expected outcome: The pump produces an upstream occlusion alarm before the infusion-complete message.

Remove the hemostat, acknowledge the alarm, and run the pump for approximately 5–10 seconds to normalize the tubing.

Repeat a failed test using a new unused IV set.

11. Test Flow-Rate Accuracy

Use either the volumetric or gravimetric method.

Program:

Run the infusion until the infusion-complete alarm occurs. Immediately press RUN/STOP so KVO does not add fluid to the collection.

Volumetric Method

Read the collected volume at the bottom of the meniscus.

Gravimetric Method

Determine the net collected fluid weight using the manual’s tare procedure.

Expected outcome: Collected volume or weight falls within the listed range.

For a failed test:

  1. Review the complete setup.
  2. Install a new approved IV set.
  3. Repeat the test.
  4. Remove the pump from service if the failure repeats.

12. Test Air Detection

Program:

Start the infusion and introduce an air bubble at least 2.5 cm / 1 inch long upstream of the pump.

Expected outcome: The pump enters an Air In Line alarm.

Close the roller clamp before unloading the tubing. Remove the air, reload and prime the set, reopen the roller clamp, resume operation, and confirm the pump can return to normal operation.

Repeat a failed test with a new unused IV set.

13. Perform the Battery Capacity Test When Indicated

The battery-capacity test is not listed as a routine annual test. Perform it when shortened runtime or another battery problem is suspected.

Charge before testing:

Verify the initial battery voltage is 8.0–8.3 V.

Run the manufacturer Battery Life Test at:

Record start-to-dead runtime and low-to-dead runtime.

Used Standard Battery

New Standard Battery, Less Than 30 Days Old

Used Wireless Battery Module

New Wireless Battery Module, Less Than 30 Days Old

Expected outcome: Battery runtime meets the applicable criteria.

A battery below the listed threshold is reaching end of life and should be handled according to the service manual and facility battery-replacement process.

Electrical Safety Testing

The provided Baxter preventive-maintenance manual does not list an electrical-safety test, ground-resistance limit, or leakage-current limit as part of this PM.

Perform electrical-safety testing only when required by:

When facility policy includes electrical-safety testing, document:

Ground resistance measured in ohms, expected ≤ 0.5 Ω or facility/NFPA 99 criteria if different:
Result =

Touch/chassis leakage measured in microamps:
Facility limit used =
Result =
Pass / Fail =

Do not represent these limits as Baxter manufacturer-listed criteria unless supported by another applicable Baxter document.

Pass / Fail Criteria

PM Pass Criteria

The pump passes when:

Remove From Service Criteria

Remove the pump from service when:

The manufacturer directs facilities to return pumps that fail inspection or testing to Baxter for service.

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide