On this page
Asset Type
Manufacturer
Model
Manual Reference
CONMED System 2450 Electrosurgical Generator Service Manual; CONMED System 5000 Electrosurgical Generator Service Manual
Manufacturer Recommendation / AEM Status
This PM procedure covers the CONMED System 2450 and System 5000 electrosurgical generators. Both manuals follow a similar preventive maintenance structure, but some acceptance limits, power output checks, calibration steps, fault-code behavior, and output coupling capacitor limits are model-specific. When a value differs, follow the limit for the specific model being tested.
The manufacturer-listed preventive maintenance schedule includes visual inspection at least every 6 months and performance testing at least once every year.
Manufacturer-listed PM items include visual inspection, cleaning as needed, chassis ground integrity testing, display / alarm / command checks, accessory and footswitch checks, output power verification, RF leakage measurement, line-frequency leakage testing, A.R.M. return electrode monitor testing, output coupling capacitor testing, calibration verification, fault-code review, and final functional verification.
Both manuals recommend periodic inspection and annual performance testing to verify safe and effective operation. Any AEM interval should be approved by the facility’s AEM program, risk assessment, service history, clinical use, and applicable regulatory requirements. The PM steps below are based on manufacturer service manual guidance.
What This PM Procedure Covers
This procedure covers preventive maintenance checks for CONMED System 2450 and System 5000 electrosurgical generators. It is intended for trained biomedical, clinical engineering, or qualified service personnel performing scheduled PM, annual performance verification, or return-to-service testing.
This procedure includes common PM workflow items for both models and clearly identifies model-specific limits where applicable. Do not treat the System 2450 and System 5000 as identical during acceptance testing.
Required Tools and Equipment
- Electrical safety analyzer
- ESU analyzer / electrosurgical tester
- Non-inductive 500 ohm load for monopolar output testing
- Non-inductive 300 ohm load for macro bipolar testing
- Non-inductive 50 ohm load for micro bipolar testing
- Monopolar footswitch
- Bipolar footswitch
- Compatible hand-controlled accessory, if needed
- Patient plate / dispersive electrode adapter
- Decade resistance box for A.R.M. testing
- Capacitance meter capable of accurate low capacitance readings at or below 1 kHz
- Short test leads as required by the manual
Step-by-Step PM Procedure
1. Verify Device Identification
Record the manufacturer, model, serial number, control number / asset number, location, PM date, and technician.
Expected result: Device identity is confirmed before testing.
2. Ensure Safety Before Testing
Remove the ESU from clinical use before performing PM. Confirm the unit is not connected to a patient, remove accessories from patient care use, and place the unit on a safe bench or cart.
Inspect for signs of liquid intrusion, odor, heat damage, cracked casing, damaged connectors, or other unsafe conditions. Do not perform output testing near a patient or active sterile field.
Expected result: Testing is performed in a controlled and safe environment.
3. Perform Visual Inspection
Inspect the power cord and plug, accessory connectors, footswitch connectors, dispersive electrode connector, handpiece / accessory receptacles, control panel, casing, hardware, and external surfaces.
If the cover is removed as part of facility procedure, inspect for obvious internal damage, loose hardware, lint, debris buildup, or contamination in the unit or heatsink area.
Expected result: No physical damage, contamination, loose hardware, connector damage, or unsafe condition is found.
Corrective action: If damage or contamination is found, remove the unit from service and correct the issue before continuing PM.
4. Clean the Unit
Clean exterior surfaces using facility-approved cleaning practices. Do not allow liquid to enter the unit. Do not use excessive liquid around connectors, displays, or the control panel.
If internal cleaning is performed by qualified personnel, remove lint or debris by vacuuming or blowing out as appropriate.
Expected result: Unit is clean, dry, and free of debris before testing.
5. Chassis Ground Integrity Test
Connect a standard ohmmeter or safety analyzer between the earth ground prong on the power plug and the equipotential ground connection. Compensate for lead resistance and measure resistance.
Acceptance limit: Ground resistance less than 0.2 ohms.
Expected result: Ground integrity measures less than 0.2 ohms.
Corrective action: If ground resistance is high, inspect the power cord, plug, equipotential stud, ground connections, and chassis bonding. Remove from service if the issue cannot be corrected.
6. Display, Alarm, and Command Functional Check
Power on the ESU. Verify the display initializes normally, no unexpected fault code remains after startup, front panel buttons respond correctly, cut / coag / bipolar selections operate as expected, power settings adjust properly, activation tones are present, alarm tones are audible, and connected footswitches or hand controls activate only the correct function.
Expected result: Displays, indicators, alarms, tones, and controls operate normally.
Corrective action: If the display is blank, controls do not respond, tones are missing, or an error code appears repeatedly, remove the unit from service and troubleshoot before completing PM.
7. Accessory and Footswitch Check
Inspect monopolar footswitch, bipolar footswitch, and hand-control accessory inputs if used for testing. Confirm connectors seat fully and that footswitches or hand controls are not stuck or falsely activated.
Expected result: Accessories connect securely and activate only when commanded.
Corrective action: Replace suspect accessories before condemning the generator.
8. Output Power Verification
Connect the ESU analyzer using the correct non-inductive load. Use 500 ohms for monopolar output testing, 300 ohms for macro bipolar testing, and 50 ohms for micro bipolar testing. Test the required output modes and power settings from the applicable service manual table.
Record measured output in watts or amps. It is not necessary to record both watts and amps if the analyzer supports one acceptable method and the reading meets the applicable table.
System 2450 Output Notes
Monopolar Cut modes include Pure and Blend. Monopolar Coag modes include Standard and Spray. Bipolar testing uses the model-specific bipolar table. Use the System 2450 service manual output power tables for exact acceptance ranges.
System 5000 Output Notes
Monopolar Cut modes include Pure, Standard, Blend 1, Blend 2, and Blend 3. Monopolar Coag modes include Spray, Standard, Pinpoint, Standard Pulse, and Spray Pulse. Bipolar modes include Macro Bipolar and Micro Bipolar. Use the System 5000 service manual output power tables for exact acceptance ranges.
Expected result: Measured output power is within the applicable manufacturer table for the model and mode being tested.
Corrective action: If output is out of tolerance, verify load resistance, ESU analyzer setup, test leads, accessory connection, selected mode, and power setting. If still out of specification, remove from service for calibration, repair, or further bench evaluation.
9. RF Leakage Measurement
Use an ESU tester with RF leakage capability or the manual-approved RF leakage test method. Ensure the unit is fully assembled and fasteners are tight. Place test equipment away from other conductive surfaces as described in the manual.
Test RF leakage to ground using the applicable model table. Test RF leakage from inactive monopolar outputs using the applicable model table. Test RF leakage between inactive bipolar outputs where required. Use short test leads where specified. Use caution with hand-control activation jumpers to prevent RF burns.
System 2450 RF Leakage Notes
- RF leakage to ground: generally less than 100 mA for monopolar paths listed
- Bipolar right / left to ground: less than 67 mA
- Inactive monopolar output limits vary by test path, including less than 50 mA, less than 20 mA, and less than 48 mA depending on the test condition
- Inactive bipolar right-to-left during hand-controlled Standard Coag: less than 48 mA
System 5000 RF Leakage Notes
- RF leakage to ground: generally less than 100 mA for monopolar paths listed
- Bipolar right / left to ground: less than 67 mA
- Inactive monopolar output limits vary by test path, including less than 50 mA, less than 20 mA, less than 40 mA, and less than 48 mA depending on the test condition
- Inactive bipolar right-to-left during right hand-controlled Standard Coag: less than 48 mA
Expected result: RF leakage readings do not exceed the applicable manufacturer limit for the model and test condition.
Corrective action: If RF leakage exceeds the limit, stop testing, verify test setup, leads, loads, and activation method. If confirmed, remove the unit from service for repair.
10. A.R.M. Return Electrode Monitor Check
Connect a decade resistance box to the dispersive electrode receptacle using the proper adapter. Reset A.R.M. by disconnecting the dispersive electrode connector or setting resistance above 10K ohms until the Single and Dual Dispersive Electrode indicators flash red in alternating fashion.
Allow approximately 2 seconds after resistance changes before judging the result. Confirm only the expected indicators are active.
A.R.M. Test Points
Dual Electrode Alarm Limit: Set DRB to 158 ohms. Expected result: Single and Dual Dispersive Electrode Status / Alarm Indicators flash red in alternating fashion.
Dual Electrode Upper Limit: Set DRB to 140 ohms. System 2450 expected result: Dual Dispersive Electrode Status / Alarm Indicator is green. System 5000 expected result: Dual Dispersive Electrode Status / Alarm Indicator flashes green and a single bar in the bargraph is illuminated.
Dual Electrode Lower Limit: Set DRB to 15 ohms. System 2450 expected result: Dual Dispersive Electrode Status / Alarm Indicator is green. System 5000 expected result: Dual Dispersive Electrode Status / Alarm Indicator flashes green and eight bars in the bargraph are illuminated.
Single Electrode Upper Limit: Set DRB to 7 ohms and reset A.R.M. System 2450 expected result: Single Dispersive Electrode Status / Alarm Indicator is green. System 5000 expected result: Single Dispersive Electrode Status / Alarm Indicator is green and not flashing.
Expected result: A.R.M. responds correctly at each resistance point for the model being tested.
Corrective action: If A.R.M. fails, verify adapter, DRB setting, connector seating, and test setup. If failure remains, remove from service for repair or calibration.
11. Output Coupling Capacitor Check
Safety warning: Set all power settings to 0 watts before performing this test. Use short leads and a capacitance meter that can accurately measure at or below 1 kHz.
Connect the shorting plug / banana adapter to the two-pin dispersive electrode receptacle. Use 6 inch or shorter test leads between the adapter and the footswitched ReadiPlug / Universal Accessory Receptacle. Measure inactive monopolar capacitance.
Confirm cut power is set to 0 watts. Activate as required and measure active cut capacitance. Do not activate for the bipolar capacitance test. Move test leads to the bipolar output accessory receptacles and measure bipolar capacitance.
System 2450 Output Coupling Capacitor Limits
- Inactive monopolar capacitance: less than 0.2 nF
- Active cut capacitance: 0.6 to 0.9 nF
- Bipolar output capacitance: 2.2 to 2.5 nF
System 5000 Output Coupling Capacitor Limits
- Inactive monopolar capacitance: less than 0.5 nF
- Active cut capacitance: 4.3 to 6.3 nF
- Bipolar output capacitance: 9.6 to 14 nF
Expected result: Capacitance readings fall within the model-specific limits.
Corrective action: If readings are out of range, verify meter accuracy, lead length, test frequency, adapter connection, and selected output. If confirmed out of range, remove from service for repair.
12. Calibration Check / Calibration Guidance
Check calibration during annual PM by verifying performance in normal operating mode. Do not recalibrate just because PM is being performed. Recalibrate only if output is out of specification, calibration-related fault codes are present, calibration memory errors occur, required boards are replaced, or the manufacturer manual requires it after repair.
System 2450 Calibration Is Required When
- Err 138, Err 139, or Err 140 occurs
- Err 143 or Err 321 occurs
- Err 135 occurs for stored A.R.M. calibration values
- Control / Display Board assembly is replaced
- Power Board assembly is replaced
- Calibration differences are found during PM
System 5000 Calibration Is Required When
- Err 140 occurs
- Err 143 occurs
- Controller board assembly is replaced
- Transformer board assembly is replaced
- Output board assembly is replaced
- Calibration differences are found during PM
Expected result: Calibration is verified through performance testing and only performed when indicated.
Corrective action: If calibration is required, follow the applicable model service manual procedure. If the unit will not calibrate or fails after calibration, remove from service for repair.
13. Fault Code / Error Check
Observe startup behavior and record any displayed error codes. If an Err code appears, cycle power once to determine if it clears. If the error repeats, remove the unit from service. Retrieve last fault codes if needed per the applicable service manual.
Expected result: No active or repeating fault codes remain after PM.
Corrective action: Do not return the ESU to service with a repeating fault code, failed POST condition, A.R.M. failure, missing activation tone, output error, or failed leakage / power test.
14. Final Functional Verification
Verify the unit powers on normally, no unexpected error codes are present, displays are readable, settings return to the appropriate default or facility-approved configuration, cut / coag / bipolar controls respond, activation tones are audible, all accessories used for testing are removed, and all covers, screws, and fasteners are secure.
Apply PM sticker or update electronic PM status as required.
Expected result: The ESU passes all required PM checks and is safe to return to service.
Electrical Safety Testing if applicable
Electrical safety testing applies to this PM procedure because the equipment is line-powered medical equipment and the provided service manual guidance includes chassis ground integrity and line-frequency leakage testing.
Ground resistance measured in ohms, expected ≤ 0.5 Ω or facility/NFPA 99 criteria if different. For this specific procedure, both manuals list a chassis ground integrity limit of less than 0.2 ohms between the earth ground prong and the equipotential ground connection.
Touch/chassis leakage measured in microamps:
Facility limit used = __________
Result = __________
Pass / Fail = __________
Line-Frequency Leakage Testing
Set all ESU power settings to 0 watts before line-frequency leakage testing. Use a biomedical electrical safety analyzer. Measure leakage using the applicable service manual table. Measure inactive leakage conditions and active-to-neutral leakage while activated where required by the manual.
Use caution when opening ground and follow facility electrical safety policy.
General Acceptance Limits
Inactive line-frequency leakage: equipotential ground, normal / reversed, ground closed: 30 µA maximum. Equipotential ground, normal / reversed, ground open: 270 µA maximum. Dispersive electrode: 15 µA maximum. Bipolar output with bipolar connections shorted together: 15 µA maximum.
Active line-frequency leakage: active output to neutral during cut activation: 15 µA maximum.
Expected result: Line-frequency leakage readings are within the applicable manufacturer limits.
Corrective action: If leakage is high, verify test setup, power cord condition, ground condition, and analyzer configuration. Remove from service if the unit fails confirmed leakage testing.
Pass / Fail Criteria
The CONMED System 2450 / 5000 Series ESU passes PM only when all required inspection, functional, electrical safety, output power, RF leakage, line-frequency leakage, A.R.M., output coupling capacitor, calibration verification, and final functional checks pass using the applicable model-specific service manual limits.
Remove the unit from service if any of the following are found: damaged power cord or plug, unsafe physical damage, liquid intrusion, failed ground integrity, failed leakage test, failed RF leakage test, output power outside the applicable table, failed A.R.M. check, failed output coupling capacitor check, calibration failure, missing activation or alarm tone, repeated fault code, failed POST condition, damaged connector, unsafe accessory behavior, loose hardware, or any condition that could affect safe clinical use.
Copy/Paste CMMS Work Order Procedure
Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.