Drager Babyleo TN500 IncuWarmer

Preventive Maintenance Procedure

On this page

Asset Type

Infant Incubator

Manufacturer

Drager

Model

Babyleo TN500 IncuWarmer

Manual Reference

Dräger Babyleo TN500 Instructions for Use, software 1.0n, document 9054863, and applicable supplement 9056310. Later software and regional configurations may differ. Confirm the installed software, oxygen, humidity, scale, heated mattress, gas-supply, battery, and accessory configuration against the current controlled documentation before service.

Manufacturer Recommendation / AEM Status

Safety checks do not replace manufacturer-listed maintenance or preventive replacement of wear parts. If a facility performs this PM at a different interval or changes the listed activities, that should be handled under the facility’s approved AEM program, risk assessment, maintenance history, current Dräger documentation, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Safety: Perform service only with no patient in the device. Provide another safe thermal environment, clean and disinfect the unit, isolate gas and electrical energy before internal work, allow heated parts to cool, and use qualified service personnel for restricted procedures.

Step-by-Step PM Procedure

1. Remove the IncuWarmer From Clinical Use

Action: Under clinical direction, transfer the patient, remove patient-use items, label the unit unavailable, allow heated components to cool, and complete current reprocessing instructions.

Expected outcome: No patient depends on the device and the system is safe and clean for maintenance.

2. Verify Identity, Configuration, and Due Work

Action: Record asset tag, serial number, software revision, mains rating, commissioning date, operating hours, and installed oxygen, humidity, scale, mattress, gas, battery, and accessory options. Review history, alarms, repairs, recalls/field actions, maintenance dates, and due parts.

Expected outcome: The asset matches the CMMS record and the correct manual revision, test path, and intervals are identified.

3. Inspect Documentation, Labels, and Completeness

Action: Confirm the applicable IFU is available; verify labels are complete and legible; and check that all components and accessories required for use are present.

Expected outcome: Documentation, identification, and configured equipment are complete and correct.

4. Inspect Trolley, Column, Hood, and Patient Barriers

Action: Inspect the trolley, housing, castors, steering function, locking brakes, lifting column, display, sensor unit, hood and hood arm, mattress tray, X-ray tray, access panels, hand ports, seals, grommets, strain reliefs, mounts, and fasteners. Move the lifting column through its complete range and confirm the hood is secured and moves safely.

Expected outcome: The system is stable and intact; steering and brakes work; the column and hood move smoothly; and no cracked, loose, deformed, or missing barrier is present.

5. Verify Panel and Hand-Port Closure

Action: Check side and front access panels and all hand ports for secure latching. Where performing the manufacturer safety check, verify the specified 20 N closing force in accordance with the applicable IEC 60601-2-19/-21 procedure.

Expected outcome: Every barrier closes, latches, and remains secure without an unsafe gap or excessive force.

6. Inspect Mattress, Air Path, Hoses, and Scheduled Parts

Action: Inspect the SoftBed or heated mattress, tray, fan impeller, sensor-wall filter cloth, gas inlets, internal/external hoses visible under the authorized procedure, cables, regulators, cylinder holder, seals, and hose grommets. Replace the filter cloth at six months and condition-based parts when due. For applicable configurations, perform the six-year internal compressed-gas-hose inspection and subsequent 24-month checks.

Expected outcome: The air and gas paths are clean and intact, the fan is undamaged, the mattress is serviceable, and all scheduled or damaged items are replaced and documented.

7. Run Startup, Controls, and Mode-Transition Checks

Action: Restore approved utilities, start the unit, and verify self-test, touchscreen, rotary knob, display, indicators, lights, date/time, menus, and configured interfaces. Confirm the device changes between incubator and radiant-warmer operation according to hood position.

Expected outcome: Startup completes without unresolved messages, controls respond, indications are clear, and hood-dependent mode transition works correctly.

8. Verify Incubator Temperature Performance

Action: Using the applicable controlled setup, set 36.0 °C and measure stabilized incubator temperature with a reference thermometer accurate to ±0.05 °C. The published supplemented criterion is 36.0 °C ±0.8 °C.

Expected outcome: Air-temperature regulation is stable and remains within ±0.8 °C of the 36.0 °C set value.

9. Verify Skin-Temperature Measurement and Control

Action: Inspect sensor connectors and channels. For the annual safety check, apply 1412 Ω in place of a sensor and verify 36.0 °C ±0.2 °C under the supplemented procedure. When the 24-month metrological check is due, also test 1586 Ω = 33.2 °C and 1312 Ω = 37.8 °C; each displayed value must be within ±0.2 °C. Verify sensor-fault and temperature-alarm responses using the current procedure.

Expected outcome: Both applicable channels display the simulated temperatures within tolerance and protective alarms/control responses function correctly.

10. Verify Heating, Oxygen, and Humidity Options

Action: Verify radiant-warmer operation, heated-mattress control when installed, and applicable overtemperature/protective responses. When oxygen control is installed and active, verify the controlled test point of 50 Vol% within ±3 Vol%. When humidity is installed, verify the controlled test point of 70% within ±10% using the current service setup.

Expected outcome: Heating systems regulate without damage or an unresolved alarm; oxygen and humidity performance meet the applicable published criteria.

11. Verify Audible, Visual, and Mains-Failure Alarms

Action: Test the acoustic alarm generator, optical alarm display, secondary alarm path where applicable, and mains-supply-failure alarm using the manufacturer procedure. Restore power and confirm normal recovery.

Expected outcome: Alarm audio and visual indications are unmistakable and the device recovers normally after power restoration.

12. Check the Optional Scale

Action: Confirm the scale belongs to the device, the cable and seals are intact, and regulatory markings remain valid. Verify zero, weighing, stability, units, and annual measurement accuracy with calibrated weights under the current scale procedure. Arrange official verification when due or invalidated by adjustment, repair, altered basic settings, damaged seals/marks, or exceeded service-error limits.

Expected outcome: The scale passes its annual accuracy check and its verification status is valid for the facility and jurisdiction.

13. Perform IEC 62353 Electrical Safety Testing

Action: Test the complete installed configuration according to IEC 62353, the current Dräger procedure, equipment class, applied-part classification, and regional requirements. Include the powered trolley, heated mattress, scale, and electrical accessories as applicable.

Expected outcome: Every required electrical-safety measurement passes its governing criterion with traceable results.

14. Reassemble and Complete the Final Check

Action: Remove test equipment, reinstall covers and approved accessories, restore authorized settings, clean surfaces, and repeat a final startup, mode, brake, panel, temperature, alarm, and option check. Record parts, measurements, corrective work, and disposition.

Expected outcome: The Babyleo is complete, clean, current on service, and either formally returned to use or clearly quarantined.

Electrical Safety Testing

Dräger includes electrical safety according to IEC 62353 in the safety check. The public IFU does not supply one universal set of numeric limits for every regional method and configuration. Use the current controlled service procedure, installed options, analyzer method, equipment class, applied parts, and regional/facility requirements.

Analyzer / serial / calibration due:
Result =

Equipment class / applied parts / configuration / method:
Result =

Protective-earth resistance, where applicable:
Test points =
Governing limit =
Highest result =
Pass / Fail / N/A =

Equipment or touch leakage:
Facility / IEC 62353 limit used =
Highest result =
Pass / Fail =

Applied-part / heated-mattress / scale / accessory tests:
Method / limits / results / N/A =

Pass / Fail Criteria

Pass

Remove From Service

Copy/Paste CMMS Work Order Procedure

Do not return the Babyleo TN500 IncuWarmer to service with a failed barrier, temperature-control, heating, oxygen, humidity, alarm, scale, mobility, or electrical-safety check; an overdue safety inspection or critical maintenance item; or any unresolved condition that could place an infant or user at risk.

Sources

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide